The unit is now calibrated and the new calibration constants are stored in the EEPROM.
For calibration accessories from General Monitors, refer to Sections 9.5.4 and 9.7.
Use Table 16 to determine 50% of full scale for the different gases that are monitored.
Sensor Type
50% of Full-Scale (ppm)
Recommended Flow
Rates (mL/min)
Ammonia 25,
50 1000
Carbon Monoxide
50, 250
500
Chlorine 5
1000
Chlorine Dioxide
1.5
1000
Hydrogen 250 500
Hydrogen Chloride
10
1000
Hydrogen Sulfide
10, 25, 50
500
Nitric Oxide
50
500
Nitrogen Dioxide
10
1000
Sulfur Dioxide
10, 50
500
Table 16: Sensor Flow Rates