GE
RAFT
V
OLUSON
™ P8/V
OLUSON
™P6
DIRECTION 5775469, R
EVISION
3
DRAFT (J
ULY
19, 2018)
B
ASIC
S
ERVICE
M
ANUAL
10-36
Section 10-9 - When There's Too Much Leakage Current...
Total Patient Leakage Current - Type CF probe to Ground
Total Patient Leakage Current - Mains to Type CF probes
Final Check. All system covers are in place. System scans with all probes as expected.
Accepted by: ______________________________________________________________________
Tester
Neutral
Tester Polarity
Switch
Ground Switch
Max Value
Allowed
Value
Measured
OK?
Comments
Closed
Normal
Closed
0.05mA
Closed
Normal
Open
0.1mA
Closed
Reversed
Closed
0.05mA
Closed
Reversed
Open
0.1mA
Open
Normal
Closed
0.1mA
Open
Reversed
Closed
0.1mA
Tester
Neutral
Tester Polarity
Switch
Ground Switch
Max Value
Allowed
Value
Measured
OK?
Comments
Closed
Normal
Closed
0.1mA
Closed
Reversed
Closed
0.1mA