Chapter 5 - Components and Functions (Theory)
5-71
3.) Recovering will task some time.
Please wait until the final window (see figure below) appears and then press
Enter
.
4.) In the displayed menus select [Exit] -> [Exit/Reboot].
5.) Remove the System DVD from the DVD/CD±R/RW Drive drive.
6.) Please wait until all processes are finished and the final “DOS” window (
below) appear.
7.) Press
A
key to activate the Ultrasound Application Software [UIS].
Figure 5-42 Restore C; done
Figure 5-43 Exit -> Exit/Reboot
NOTICE
The system boots with windows. An automated process was developed to check disk, assign an unique
computer name and matching drive letters.
The system will reboot at least 2 times!
Figure 5-44 press “A” key to start UISApp