GE
DRAFT
V
OLUSON
™ P8/V
OLUSON
™P6
DIRECTION 5775469, R
EVISION
3
DRAFT (J
ULY
19, 2018)
B
ASIC
S
ERVICE
M
ANUAL
Chapter 7 - Diagnostics/Troubleshooting
7-27
7-6-2
Diagnostics
The Diagnostic page uses a web-controlled user interface to perform diagnostics and troubleshooting.
NOTE:
Remote access to the Voluson S&P- series system requires permission and customer input before a
GE service technician or OLC can access the customer’s system remotely.
Disruptive Mode can be selected by the customer directly on the Voluson S&P - series (See,
), or requested remotely by the service technician or OLD (see
).
NOTICE
Plea
Whenever any hardware diagnostic tests have to be executed on site, the CSD must be accessed via
the Windows Start Menu - All Programs - Voluson - GE Field Engineer - Common Service Desktop
Figure 7-28 Common Service Desktop -Diagnostics
Section 7-6-2-1 Disruptive Mode
Section 7-6-2-2 S.M.A.R.T. Tests
Section 7-6-2-4 GPU (Graphics Processing Unit) Test
Section 7-6-2-5 Show all HW Test Results
Please observe that the displayed screens may slightly differ from images shown in this manual. The