Chapter 10 - Care & Maintenance
10-33
ELECTRICAL SAFETY
Outlet and Ground Continuity test
Earth leakage current test
Enclosure leakage current test
Outlet/Ground Continuity
Max Value
Allowed
Value
Measured
OK?
Comments
Outlet (correct ground &wiring
config.)
N/A
N/A
System Ground Continuity
0.2 Ohm
Tester Neutral
Tester
Polarity
Switch
Max Value Allowed
Value
Measured
OK?
Comments
Enter Name of tested peripheral here:
Closed
Normal
0.3mA(USA)/0.5mA(Others)
Closed
Reversed
0.3mA(USA)/0.5mA(Others)
Open
Normal
1mA
Open
Reversed
1mA
Tester
Neutral
Tester
Polarity
Switch
Protective
Earth
Switch
Max Value Allowed
Test 1
Wheel
Test 2
LCD
Optional
Test 3
Optional
Test 4
OK?
Comments
Enter Name of tested peripheral here:
Closed
Normal
Closed
0.1mA(USA)/
0.1mA(Others)
Closed
Normal
Open
0.3mA(USA)/
0.5mA(Others)
Closed
Reversed
Closed
0.1mA(USA)/
0.1mA(Others)
Closed
Reversed
Open
0.3mA(USA)/
0.5mA(Others)
Open
Normal
Closed
0.5mA(USA)/
0.5mA(Others)
Open
Reversed
Closed
0.5mA(USA)/
0.5mA(Others)