background image

GE Healthcare

Solar

 8000M/i patient monitor

Service Manual

Software Version 5

Solar™ 8000M/i patient monitor
English
2026266-004 (CD)
2026264-042C (paper)
© 2007 General Electric Company
All rights reserved.

Содержание Solar 8000M/i

Страница 1: ...GE Healthcare Solar 8000M i patient monitor Service Manual Software Version 5 Solar 8000M i patient monitor English 2026266 004 CD 2026264 042C paper 2007 General Electric Company All rights reserved ...

Страница 2: ...the abbreviation GE is used for the legal entity name GE Medical Systems Information Technologies Listed below are GE Medical Systems Information Technologies trademarks All other trademarks contained herein are the property of their respective owners APEX DASH MUSE RSVP SOLAR TRAM TRIM KNOB and UNITY NETWORK are trademarks of GE Medical Systems Information Technologies registered in the United St...

Страница 3: ...2 1 System components 2 2 Solar 8000M i patient monitoring system 2 2 Solar 8000M i patient monitor 2 3 iPanel computer 2 4 UnityView remote display controller 2 4 Tram rac housing 2 5 Patient Data Module PDM 2 6 Connectivity devices 2 6 PRN 50 PRN 50 M digital writer 2 7 Laser printer 2 7 Remote control or keypad 2 8 Remote displays 2 8 Device compatibility 2 9 Acquisition modules 2 9 Peripheral ...

Страница 4: ...ork terms 3 13 4 Maintenance and checkout 4 1 Maintenance schedule 4 2 Manufacturer recommendations 4 2 Manufacturer responsibility 4 2 Preventive maintenance 4 2 Visual inspection 4 3 Cleaning 4 4 Cleaning precautions 4 4 Exterior cleaning 4 4 Cleaning disinfecting and storing GE ECG cables and leadwires 4 5 Electrical safety tests 4 8 Recommendations 4 8 Set up 4 9 Power outlet test 4 9 Ground e...

Страница 5: ...ed 5 2 Abort main code 5 2 Boot loader or boot code 5 2 Cold start 5 2 Continue main code 5 2 Monitor memory 5 3 Protected memory 5 3 Power cycle or reboot 5 3 Service mode main code 5 3 Service menu boot code 5 3 Warm start boot code 5 3 Country selection 5 3 Set language 5 3 Service menus 5 4 Boot code Service Menu 5 4 Main code SERVICE MODE menu 5 6 Fault analysis 5 10 Overview 5 10 Calibration...

Страница 6: ...et graph locations 6 5 Admit Menu 6 6 Set line frequency 6 7 Set defib sync voltage and pulse width 6 7 Set country selection 6 8 Set language 6 8 Calibrate touchscreen 6 8 Completion 6 8 Advanced user procedures 6 9 Procedures 6 9 Set time and date 6 9 Change software level 6 10 Enable options 6 10 Transfer monitor defaults 6 10 Change Ethernet address 6 14 Set internet address 6 15 Power cycle 6...

Страница 7: ... A 22 Purchaser s responsibility A 22 Medical grade displays A 22 Non medical grade displays A 23 Required specifications for non medical grade CRT displays A 24 Recommended specifications for non medical grade CRT displays A 25 Required specifications for non medical grade digital flat panel displays A 26 Recommended specifications for non medical grade digital flat panel displays A 27 Appendix B...

Страница 8: ...vi Solar 8000M i patient monitor 2026265 075C ...

Страница 9: ...2026265 075C Solar 8000M i patient monitor 1 1 1 Introduction ...

Страница 10: ...field repairable Where necessary the manual identifies additional sources of relevant information and technical assistance See the operator s manual for the instructions necessary to operate the equipment safely in accordance with its function and intended use Intended audience This manual is intended for service representatives and technical personnel who maintain troubleshoot or repair this equi...

Страница 11: ...fore connecting any devices to the equipment that are not recommended in this manual Parts and accessories used must meet the requirements of the applicable EN 60601 series safety standards and or the system configuration must meet the requirements of the EN 60601 1 1 medical electrical systems standard Periodically and whenever the integrity of the device is in doubt test all functions The use of...

Страница 12: ... defined as a source of potential injury to a person DANGER indicates an imminent hazard which if not avoided will result in death or serious injury WARNING indicates a potential hazard or unsafe practice which if not avoided could result in death or serious injury CAUTION indicates a potential hazard or unsafe practice which if not avoided could result in minor personal injury or product property...

Страница 13: ...d part suitable for intentional external and internal application to the patient excluding direct cardiac application Paddles outside the box indicate the applied part is defibrillator proof Medical Standard Definition F type applied part floating isolated complying with the specified requirements of IEC EN UL 60601 1 Medical Standards to provide a higher degree of protection against electric shoc...

Страница 14: ...n Limits for Certain Hazardous Substances in Electronic Information Products The number in the symbol is the Environment friendly User Period EFUP which indicates the period during which the toxic or hazardous substances or elements contained in electronic information products will not leak or mutate under normal operating conditions so that the use of such electronic information products will not...

Страница 15: ... of the responsible individual hospital or institution using this equipment to implement a satisfactory maintenance schedule may cause undue equipment failure and possible health hazards Regular maintenance irrespective of usage is essential to ensure that the equipment will always be functional when required Equipment identification Every GE device has a unique serial number for identification A ...

Страница 16: ...1 8 Solar 8000M i patient monitor 2026265 075C Introduction Service information ...

Страница 17: ...2026265 075C Solar 8000M i patient monitor 2 1 2 Equipment overview ...

Страница 18: ...M i processing unit Display Keypad and or remote control Two possible acquisition devices Tram rac housing with acquisition module s Patient Data Module also referred to as PDM Optional components include iPanel computer Clinical Information Center central station Remote display VGA and DFP NOTE Available on Solar 8000M and Solar 8000i patient monitors with dual display capability Printer PRN 50 P...

Страница 19: ...ng unit is the center of the Solar 8000M i patient monitoring system It provides the user controls the processors to communicate with various patient monitoring modules and it analyzes patient data It can display up to eight different waveforms at one time System software may be updated using a laptop computer connected to the Solar 8000M i processing unit or the Unity Network or from a Clinical I...

Страница 20: ...ication is compatible with the following patient data web portals Centricity CIV Centricity CV Cardiology Web MUSE CV Web Centricity Enterprise Web UnityView remote display controller The UnityView remote display controller consists of a remote display controller with a compatible display purchased from GE or another vendor The controller connects to the Unity Network and may be configured to disp...

Страница 21: ...r the patient monitor There are two Tram rac housings available for the monitor Tram rac 2 housing holds one Tram module Tram rac 4A housing holds one Tram module and two additional single high modules See the Tram rac Housing Service Manual for additional information Shown below is a Tram rac 4A housing with a Tram module and two single parameter modules inserted 005B ...

Страница 22: ...l for additional information Connectivity devices The Unity Network ID connectivity device acquires digital data from eight individually isolated serial ports The data is collected from up to eight peripheral devices not necessarily manufactured by GE then the device transmits the formatted data to the Solar 8000M i patient monitor See the appropriate connectivity device service manual for additio...

Страница 23: ...t have a serial port and an interface adapter is required for the cable between the laser printer and the monitor Refer to the instructions provided in Laser Printer Support Kit pn 2013421 001 for details on the interface adapter and installing a serial card in a laser printer WARNINGS SHOCK HAZARD Laser printers are UL 60950 EN 60950 certified equipment which may not meet the leakage current requ...

Страница 24: ...allows the user to operate the patient monitor from across a room Eighteen hard keys are configured for adult neonatal or operating room applications The keypad can be mounted on the display or on a separate holster that has various mounting configurations Remote displays Depending on your Solar 8000M i configuration there are up to two VGA CRT analog flat panel ports and two DFP digital flat pane...

Страница 25: ...eries 400 Dual BP Module BP Module BP Dual Temp Module GE SpO2 Module Masimo SET SpO2 Module Capnostat Mainstream EtCO2 Module Capnostat Mainstream Module Capnostat Dual CO2 Module Pryon Mainstream Module Pryon Sidestream Module SAM Module SAM80 Module Tram Module w ECG Resp CO 2 BP NIBP SpO2 Tram Module w ECG Resp CO 3 BP NIBP SpO2 Tram Module w ECG Resp CO 4 BP SpO2 Tram Module w ECG Resp CO NIB...

Страница 26: ...spiratory Mechanics Module Impedance Cardiograph Module Device Interface Solar 8000M i Remote M Port Solar 8000M i Keypad M Port PRN 50 M Port with M Port compatible PRN50 RAC 4A Comm TRAM NET RAC 4A DAS TRAM NET RAC 2 TRAM NET Unity Network ID M Port RM Module M Port with M Port compatible RM module Serial download RS 232 1 Elo Touchscreen RS 232 2 Remote Alarm M Port Laser printer M Port iPanel ...

Страница 27: ... Pager LAN MARS 5000 8000 ApexPro MUSE Aware Gateway Octacomm CDT LAN QS CIC RSVP CIC Pro C Solar 7000 8000 Dash 2000 Solar 8000M Dash 3000 4000 5000 Solar 9000 9500 Eagle 3000 ST Guard Eagle 4000 Tramscope 12 HL7 Outbound Auto View ICMMS Service Web CO2 Module StatView Transcutaneous Module iPanel Unity Network ID interface Impact Pager Unity Network IS Patient Viewer Managed Care Unity Network P...

Страница 28: ...bility Interfacing with other peripheral devices The Solar 8000M i patient monitor can interface with other peripheral bedside monitoring devices through an Unity Network ID connectivity device For a list of supported devices see the Unity Network Interface Device ID Service Manual ...

Страница 29: ...l functions System software may be updated using a laptop computer connected to the Solar 8000M i processing unit or the Unity Network or from a Clinical Information Center CIC on the Unity Network The following theory of operation provides an overview of the various functional circuit boards in the monitor Processor Board Ethernet TRAM NET 1 TRAM NET 2 ePort Serial Port 1 Serial Port 2 DFP 1 DFP ...

Страница 30: ...2 14 Solar 8000M i patient monitor 2026265 075C Equipment overview Theory of operation Block diagram of internal connections Power Supply Assembly Speaker Processor Board 036B ...

Страница 31: ...t overview Theory of operation Processor board The processor board processes acquired data for the generation of displayed information audible alarms and supports communication channels for the acquisition system serial peripherals and the Unity Network 035B ...

Страница 32: ...t State Generator via Eight 8 General Purpose Chip Select Machines GPCM and two 2 Universal Programmable Machines UPM Development Port Background Debug Monitor System Configuration and Protections such as the Bus Monitor Software Watchdog Timer and Periodic Interrupt Timer PLL Clock synthesizer Communication Processor Module CPM including One 1 Fast Ethernet Channel Media Independent Interface Fou...

Страница 33: ...on reset Battery switchover Battery charge level indication battery OK Power going away indication Low Line SRAM chip select gating Manual reset input Watchdog facility Real time clock The Epson RTC 8593AA is the Real Time Clock RTC used It is an I2C RTC with an internal crystal and can operate with a supply voltage of 3 3V The storage of the year will have to be done in another location since thi...

Страница 34: ...d two digital flat panel displays can be used A 65 00 MHz clock oscillator is used to drive the video system The MPC860 accesses the video systems over the PCI Bus using the Host Bridge implementation within the FPGA Video system components The video system has a video graphics chip some discretes and connectors for VGA RGB and DFP Digital Flat Panel video displays as well as a RS 232 serial port ...

Страница 35: ...b facility The TRAM NET 1 port provides TRAM NET communication only for connection to the Tram rac 4A or Tram rac 2 The TRAM NET 2 ePort port provides TRAM NET communication or ePort communication for connection to the PDM with an ePort cable When the TRAM NET 2 ePort connector is connected to PDM the connection uses ethernet through the Broadcom ethernet switch Processor SCC channel The MPC860 fo...

Страница 36: ...transfer additional bytes of data is synchronized by the MPC860 waiting for a received character interrupt response from the DS2480 before it can load the next character The ID signal has its own Return line The ID Signal and its Return share the RJ 45 connector pins that the M Port Ethernet 10 Base T Transmit differential pair uses The functionality of the shared connector pins is determined by a...

Страница 37: ...the ferrite bead and the copper traces form an impedance divider with the planar capacitor formed by the dielectric material in the PCB layer between the copper of the isolated return plane and the copper of the common earthed plane The expectation is that the physical impedance divider formed by the bead trace connector inductance and the planar capacitor will accomplish the following two results...

Страница 38: ...upply generates DC voltages necessary to power the processor board and the communication channels M ports and devices connected to TRAM NET It consists of a mains AC line PWM converter that creates a 16 75 output voltage bus from which two outputs are developed The main 16 75V output also provides external power to the RAC data acquisition modules and plug in modules Speaker The speaker generates ...

Страница 39: ...2026265 075C Solar 8000M i patient monitor 3 1 3 Installation ...

Страница 40: ...wer supply connect the power cord as shown NOTE The number of video connectors varies by configuration TRAM NET connectors Video connectors Power cord Ethernet connector RS 232 connectors Power switch Digital flat panel connectors Voltage select 007A 008C 010C Solar 8000M Solar 8000i Power switch Power cord Voltage select TRAM NET ePort connectors Ethernet connector RS 232 connectors Video connect...

Страница 41: ...acquisition modules NOTE If two Tram racs are connected in any configuration one must have a power supply The following devices connect to either of the two TRAM NET ports on the Solar 8000M i monitor The connector is a 9 pin D type Tram rac 4a housing with or without power supply Tram rac 2 housing Tram rac 4 housing with or without power supply Tram rac 3 housing with or without power supply Pat...

Страница 42: ...lugging the power cord into an AC outlet Connecting these devices to a powered Solar 8000M i patient monitor could damage connectors Patient Data Module The Patient Data Module PDM is an acquisition device that connects to the Solar 8000i patient monitor via the ePort cable Connect one end of the ePort cable to the TRAM NET2 ePort connector and the other end to the Patient Data Module 013C One Tra...

Страница 43: ...e connector is an 8 pin RJ 45 type VGA VID 1 and 2 NOTE The number of video connectors varies by configuration The two VGA connectors provide an interface to analog VGA displays The connector is a 15 pin high density D type WARNING Do not connect a monochrome display to the Solar 8000M i patient monitor Visual alarm indictors may not appear properly resulting in a hazard to the patient 043B One Tr...

Страница 44: ...32 1 The RS 232 1 serial connector provides an interface to the iPanel computer If not using an iPanel it interfaces to a PC for software upgrades or polled parameter service The connector is a 9 pin D type RS 232 2 The RS 232 2 serial connector provides a touchscreen interface The connector is a 9 pin D type NOTE Use cable 2006733 00X for touchscreen connection The cable supplied with the display...

Страница 45: ...to M Port adapter PN 2001973 001 is required The adapter must connect to the AutoPort device not the M Port host the Solar 8000M i patient monitor The following devices connect directly to the M Ports The connector is an 8 pin RJ 45 type Solar 8000M i remote control Solar 8000M i keypad PRN 50 M digital writer RM M respiratory mechanics module Remote Alarm Terminal Nurse Call and Alarm Light Syste...

Страница 46: ... for specific care areas adult neonatal or operating room A keypad holster mount is available for mounting under the display NOTE The Solar 8000M i requires an interface remote control or keypad with all display types NOTE The error message WARNING REMOTE MISMATCHED WITH MONITORING MODE displays if a mismatched keypad remote control is connected to the Solar 8000M i patient monitor PRN 50 digital ...

Страница 47: ...ails on interconnection using the interface adapter and installing a serial card in a laser printer WARNING SHOCK HAZARD Laser printers are UL 60950 EN 60950 certified equipment which may not meet the leakage current requirements of patient care equipment This equipment must not be located in the patient vicinity unless the medical system standard EN 60601 1 1 is followed Do not connect a laser pr...

Страница 48: ... but there is no communication Quick flashing yellow indicates that too many identical devices are connected or the device cannot be identified Refer to Chapter 5 Troubleshooting if an LED is not green Power up NOTE Check power voltage at your location and set power to either 120 V or 240V After making all connections plug the power cord into an AC wall outlet turn the power switch to 1 on and tur...

Страница 49: ...akes a TRAM NET small area network available for the peripheral devices The TRAM NET controller resides within the main processor which provides efficient data transfer by sharing main memory Internal hub TRAM NET is a small network that offers ample flexibility a high rate of communication and relatively inexpensive cabling Data is transmitted at the rate of 921 6K bits per second It uses a star ...

Страница 50: ...used The Solar 8000M i is designed to be used with twisted pair cabling Consult GE when trying to interface with either thick net or thin net cabling The real time GE Unity Network operates at 10 Mbps half duplex Twisted pair Twisted pair is the most popular cabling because it is easy to install and flexible to work with It uses the star topology with a switch as the hub of the segment A maximum o...

Страница 51: ...12 to 48 network devices and may be linked to other switches to form larger networks Segment A network segment is comprised of all devices connected to 1 or many switches which are in turn connected together to form a larger network The boundaries of the segment are defined by networking equipment that regulate the flow of packets into and out of the segment e g routers and switches IP address A 3...

Страница 52: ...3 14 Solar 8000M i patient monitor 2026265 075C Installation Ethernet communication ...

Страница 53: ...2026265 075C Solar 8000M i patient monitor 4 1 4 Maintenance and checkout ...

Страница 54: ...y Manufacturer responsibility WARNING Failure on the part of all responsible individuals hospitals or institutions employing the use of this device to implement the recommended maintenance schedule may cause equipment failure and possible health hazards The manufacturer does not in any manner assume the responsibility for performing the recommended maintenance schedule unless an Equipment Maintena...

Страница 55: ...ent for physical damage to the case the display screen and the keypad Do not use the monitor if damage is determined Refer damaged equipment to qualified service personnel Inspect all external connections for loose connectors or frayed cables Have any damaged connectors or cables replaced by qualified service personnel Inspect the display face for marks scratches or other damage Physical damage to...

Страница 56: ...alcohol based cleaning agents Betadine solution a wax containing a cleaning substance or abrasive cleaning agents Exterior cleaning Clean the exterior surfaces with a clean lint free cloth and one of the cleaning solutions listed in the table above Wring the excess solution from the cloth Do not drip any liquid into open vents switches plugs or connectors Dry the surfaces with a clean cloth or pap...

Страница 57: ...s to prevent pulling the long wires from the connector ends Metal connections can be pulled away from the connectors 3 For general cleaning of cables and leadwires wipe using a lightly moistened cloth with a mild soap and water solution Then wipe and air dry 4 For disinfecting the cables and leadwires wipe exterior with a soft lint free cloth using the following solution as recommended in the APIC...

Страница 58: ... follow the recommendations from the sterilizer manufacturer for required aeration Cautions Never immerse the handheld device cables or leadwires in any liquid Do not pour or spray any liquid directly on cables or leadwires or permit fluid to seep into connections or openings Never use conductive solutions solutions that contain chlorides wax or wax compounds to clean handheld devices cables or le...

Страница 59: ...leaning products known to cause the types of problems listed above include but are not limited to Sani Cloth Wipes Ascepti Wipes HB Quat Clorox Wipes they do not contain bleach Over the counter detergents e g Fantastic Tilex etc Products that contain active ingredients and solutions similar to these products should also be avoided NOTE For additional information refer to the How to Reach Us page o...

Страница 60: ...every component in the system If the system includes a Tram rac housing without its own power supply it should remain connected to the monitor throughout the safety tests WARNING Failure to implement a satisfactory maintenance schedule may cause undue equipment failure and possible health hazards Unless you have an Equipment Maintenance Contract GE does not in any manner assume the responsibility ...

Страница 61: ...electrical safety tests The results of the following tests will be inaccurate unless a properly wired power outlet is used Ground earth integrity Listed below are two methods for checking the ground earth integrity Ground Continuity Test and Impedance of Protective Earth Connection These tests determine whether the device s exposed metal and power inlet s earth ground connection have a power groun...

Страница 62: ...0751 standards Consult your country local safety agency if in question Check compliance as follows 1 A current of 25A from a current source with a frequency of 50 or 60 Hz with a no load voltage not exceeding 6 V is passed for at least 5 seconds through the protective earth terminal or the protective earth pin in the mains plug and the equipotential stud which could become live in case of failure ...

Страница 63: ... Read the current leakage indicated on DMM NOTE If either reading is greater than the appropriate specification below the device under test fails Contact GE Technical Support 300 μA 0 3 volts on the DMM and the device under test is powered from 100 120 V 50 60 Hz 300 μA 0 3 volts on the DMM and the device under test is powered from a centered tapped 200 240 V 50 60 Hz single phase circuit 500 μA 0...

Страница 64: ...age currents and the UL and EN limits are different 5 Set the polarity switch to RVS 6 Read the current leakage indicated on DMM NOTE If either reading is greater than the appropriate specification below the device under test fails Contact GE Technical Support 300 μA 0 3 volts on the DMM and the device under test is powered from 100 120 V 50 60 Hz 300 μA 0 3 volts on the DMM and the device under t...

Страница 65: ...nd measures the leakage current from the ECG RESP connector then the SpO2 connector of the device to ground 1 Configure the leakage tester like the circuit shown below with GND switch OPEN and polarity switch NORM 2 Connect an ECG test body to the ECG RESP connector of the device under test 3 Set the power switch of the device to ON 4 Read the leakage current indicated on the DMM 5 Change the leak...

Страница 66: ...is test pertains to the Solar 8000M i part of the monitoring system This procedure only applies to Class I grounded earthed equipment and measures the leakage current from a mains voltage source into the ECG RESP connector then the SpO2 connector 1 Configure the leakage tester like the circuit shown below with GND switch CLOSED and polarity switch NORM WARNING Shock hazard The following step cause...

Страница 67: ...MM at 120 240 VAC using the ECG cable NOTE The 10 and 20 μA limits are based on internal design standards The 50 μA limit is common to all standards AAMI ES 1 standard requires using the patient cable 6 Set the power switch on the device to OFF 7 Repeat the steps in this procedure using the appropriate SpO2 Test Body Connect the SpO2 Test Body to the blue SpO2 connector of the device under test Te...

Страница 68: ...chapter Required tools and equipment NOTE Be sure test equipment is maintained according to manufacturer s maintenance schedule to ensure accuracy The following table lists the test equipment adapters and cables necessary to complete the checkout procedures Equivalent equipment may be substituted Acquisition device Tram 100 851 module Any GE Tram 100 851 module Patient Data Module Any GE Patient D...

Страница 69: ...tor operator s manual if necessary 4 The waveforms should look clean no noise Temperature test 700 400 series dual temperature adapter 402015 004 Temperature simulator cable for use with Marq III Cardiac output test Cardiac output simulator II 900028 001 no longer available for ordering Cardiac output cable adapter 700092 001 Sp02 test Masimo SpO2 Test Kit includes Masimo Tester and SpO2 Sensor Ad...

Страница 70: ...r LED is ON at the Tram rac housing 2 Disconnect and reconnect the Tram rac housing communication cable Verify the recovery of the waveforms 3 If the Tram rac housing has additional slots for input modules insert a BP module Connect simulator and verify communication to the monitor Repeat for each slot 4 If the Tram rac housing has an optional power supply check the following on the connector that...

Страница 71: ...c 4A bezel number for BP output Pin 1 Pin 8 Signal GND for Tram Waveforms Pin 2 Pin 2 Trace I ECG II1 Tram1 Pin3 Pin 6 Tram BP3 or SPO2 Value Tram1 Pin 4 Reserved for Future Use Pin 5 Pin 4 Tram ART 1 or BP1 Tram Pin 6 Pin 9 Slot 3 Series 7000 Waveform A Right Side or Module Parameter 6 Pin 7 Pin 11 Slot 4 Series 7000 Waveform A Right Side or Module Parameter 8 Pin 8 Pin 8 Signal GND for Series 70...

Страница 72: ... Connect the Patient Data Module between the ePort connector and the TRAM NET 2 ePort connector on the Solar 2 Observe correct type of device identified at the monitor 3 Simulate and observe waveform on monitor Communication 1 Connect the Tram rac housing cable into each of the two TRAM NET connectors Or Connect the Patient Data Module between the ePort connector and the TRAM NET 2 ePort connector...

Страница 73: ...e information appears on the display 5 Select Tools Menu 6 Use password mei to open Service and Diagnostic Tools Menu 7 Select M Port Tools Show M Port X Serial Incoming Data Where X the M Port number 8 If the message No incoming data read appears nothing is connected to the M Port Connect the keypad or remote control or choose the correct M Port 9 As the columns of numbers scroll press the keys a...

Страница 74: ...ameters except ECG b In MONITOR SETUP select WAVEFORMS ON OFF and set leads in the following order Waveform 1 II Waveform 2 I Waveform 3 III Waveform 4 V1 Waveform 5 AVL Waveform 6 AVF c From the ECG Menu select MORE ECG QRS VOLUME and set to the desired volume percentage 4 Admit a patient into patient monitor MORE MENUS ADMIT MENU ADMIT PATIENT 5 Check that the following conditions are true The m...

Страница 75: ...t the waveform pulses are 1mV in amplitude both on the printout and patient monitor display 12SL ECG analysis program functions 1 Connect a 12 lead patient cable to the patient monitor and simulator 2 At the patient monitor select the ECG parameter box 3 At the ECG menu select 12 LEAD ECG ANALYSIS 4 Check that all 12 ECG waveforms display clearly are noise free and display an amplitude equal to th...

Страница 76: ...eading of approximately 120 80 93 on the patient monitor 10 Repeat these tests with each IP BP connector selecting BP2 BP3 etc NOTE Use the Dual IP Adapter Cable to test the third and fourth IP functionality for each IP BP connection Temperature functions 1 Go to TEMP functions on the simulator and set a 37 C temperature output 2 Connect a temperature sensor adaptor to the TEMP CO connector of the...

Страница 77: ...ad approximately 5 of simulated value 7 Repeat this test with the adapter and simulator set to 24 and the COMPUTATIONAL CONSTANT set to 0 595 The computed output should be 5 of simulated value SpO2 functions Masimo 1 Connect the Masimo SpO2 tester cable to the SpO2 sensor adapter cable 2 Connect the SpO2 adapter cable to the SpO2 connector on the acquisition device 3 Turn the patient monitor on 4 ...

Страница 78: ...MOD LOW 7 Verify the following SpO2 readings for saturation and pulse rate Saturation 75 2 Rate bpm 200 3 194 to 206 8 Press and release the light button on the SRC MAX Verify default indicators are lit as follows Heart rate 200 bpm Light HIGH SpO2 75 MOD LOW 9 Verify the following SpO2 readings for saturation and pulse rate Saturation 75 2 Rate bpm 200 3 194 to 206 10 Press and release the SpO2 b...

Страница 79: ...and MODULATION switches as follows Set the LIGHT switch to HIGH 1 Set the MODULATION switch to HIGH Set the RATE switch to 112 Allow the Tram a few seconds to obtain a steady reading 5 Verify the following SpO2 readings for saturation and pulse rate Saturation 81 2 Rate bpm 112 2 110 to 114 6 Set the LIGHT and MODULATION switches as follows Set the LIGHT switch to LOW Set the MODULATION switch to ...

Страница 80: ...heck 1 Connect the acquisition device to the patient monitor 2 Disconnect all parameter cables 3 Apply power to the patient monitor 4 Make sure the power indicator on the acquisition device s front panel is ON green WARNING Do not put the NBP cuff around a human limb during the checkout procedure due to the potential for injury When the NBP cuff is used in this procedure it must be tightly wrapped...

Страница 81: ...res calibration Complete this section then continue with NBP calibration below 15 Select CHECK CAL IN PROGRESS 16 Select STOP The module then releases pressure in the cuff 17 Disconnect the NBP cuff and manometer from the acquisition device NBP calibration 1 From the monitor s main menu select MONITOR SETUP 2 Select SERVICE MODE 3 Enter the password to get into the service mode The first two digit...

Страница 82: ... to the MAIN MENU button 13 Select CHECK CAL OFF 14 Select START 15 The text on the menu item changes from CHECK CAL OFF to CHECK CAL IN PROGRESS Make sure that the pressure readings shown as CUFF in the NBP parameter box on the monitor display and manometer are equal 1 mmHg for at least one full minute 16 Select CHECK CAL IN PROGRESS 17 Select STOP The module then releases pressure in the bulb or...

Страница 83: ...Wire Color Signal Name 1 Brown ECG_ANALOG_OUT 2 Red BP_ANALOG_OUT 3 Orange NO_CONNECTION 4 Yellow MARKER_RETURN 5 Shield DRAIN WIRE 6 Green ANALOG_RETURN 7 Blue NO_CONNECTION 8 Violet DEFIB_SYNC_MARKER_IN 9 Gray DEFIB_SYNC_MARKER_OUT ECG 023A Signal pin 1 Brown wire Ground pin 6 Green wire Time division 0 2s Volts division 0 5v ...

Страница 84: ...d 10ms respectively from the factory Arterial BP 024A Signal pin 2 Red wire Ground pin 6 Green wire Time division 0 2s Volts division 0 2v Marker Out Frequency 025A Signal pin 9 Gray wire Ground pin 4 Yellow wire Time division 10ms Volts division 0 5v Marker Out Pulse Width 026A Signal pin 9 Gray wire Ground pin 4 Yellow wire Time division 5ms Volts division 1v ...

Страница 85: ... two Marker Out traces shown The top trace shows the frequency of the pulses the bottom trace shows the pulse width 039A Observe normal R waves before shorting Marker Out and Marker In Observe negative spikes in the R waves after shorting Marker Out and Marker In Note that the spikes are small and can be diffucult to see at times 432B Defibrillator Synchronization connector J2 Pin Name Description...

Страница 86: ...are configured at 5V and 10ms respectively from the factory To change these settings refer to the Calibration chapter of your TRAM modules service manual ECG Signal Pin 7 Ground Pin 3 Probe Type x10 Time Division 0 2S Volts Division 0 5V Arterial BP Signal Pin 6 Ground Pin 5 Probe Type x10 Time Division 0 2S Volts Division 0 2V 023A 024A ...

Страница 87: ...ed ECG waveforms 4 Remove the jumper Marker Out Frequency Signal Pin 1 Ground Pin 4 Probe Type x10 Time Division 0 2S Volts Division 1V Marker Out Pulse Width Signal Pin 1 Ground Pin 4 Probe Type x10 Time Division 5mS Volts Division 1V 025A 026A Observe normal R waves before the jumper is installed Observe negative spikes in the R waves while the jumper is installed Note that the spikes are small ...

Страница 88: ...65 075C Maintenance and checkout Checkout procedure Completion Test additional modules using procedures found in the service manual that accompanied the module Disconnect all test equipment Return the monitor and acquisition device to service ...

Страница 89: ...g disinfecting and storing GE ECG cables and leadwires Electrical safety tests on page 4 8 1 ____ Power outlet test 2 ____ Ground earth integrity 3 ____ Ground earth wire leakage current tests 4 ____ Enclosure leakage current test 5 ____ Patient source leakage current test 6 ____ Patient sink leakage current test Checkout procedure on page 4 16 1 ____ Solar 8000M i Unity View display 2 ____ Touchs...

Страница 90: ...ons 3 ____ Respiration functions 4 ____ Invasive blood pressure functions 5 ____ Temperature functions 6 ____ Cardiac output functions 7 ____ SpO2 functions Masimo ____ SpO2 functions Nellcor OxiMax ____ SpO2 functions Nellcor OxiSmart Tram only 8 ____ NBP calibration check 9 ____ Defib sync and analog output test Patient Data Module ____ Defib sync and analog output test Tram Module ...

Страница 91: ...2026265 075C Solar 8000M i patient monitor 4 39 Maintenance and checkout Repair Log Repair Log Unit serial number Institution name Date Maintenance or repair Technician ...

Страница 92: ...4 40 Solar 8000M i patient monitor 2026265 075C Maintenance and checkout Repair Log ...

Страница 93: ...2026265 075C Solar 8000M i patient monitor 5 1 5 Troubleshooting ...

Страница 94: ...ation appears on the display Cold start A cold start is used only in extreme circumstances It erases the protected memory language unit name and bed number automatically discharges the patient and restores factory defaults Ethernet address internet address software feature level and CRG trend option remain unchanged 1 Hold down NBP Go Stop and Zero All 2 Press and release the TRIM KNOB control and...

Страница 95: ...ils later in this chapter Service menu boot code The Service Menu found in the Boot Code is used when downloading the Boot Code and main processor code See details later in this chapter Warm start boot code A warm start activates the software previously downloaded The following steps activates a warm start 1 Hold down NBP Go Stop and Zero All 2 Press and release the TRIM KNOB control 3 Keep holdin...

Страница 96: ...rvice menu when downloading new Boot Code or Main Code software to the patient monitor or when the patient monitor exhibits a serious failure Activate the Boot Code program as follows 1 Hold down NBP Go Stop and Zero All on the keypad or remote control 2 Press and release the TRIM KNOB control 3 Keep holding NBP Go Stop and Zero All until the Boot Code information appears on the display Set up fir...

Страница 97: ...used when downloading software from a laptop PC Serial Download Boot This option is used when downloading software from a laptop PC View Main Code Error Logs Allows access to the error logs generated by the Main Code Tools Menu This password protected menu contains service diagnostic tools Use password mei Video Test Screens Various color screens for testing the display Options Menu In this menu y...

Страница 98: ...ta Module or other discrete modules Refer to the module service manual for details See Chapter 6 Configuration for calibration details HARDWARE TEST This menu provides access to the hardware tests for the monitor Hardware Status The one of a kind patient monitor LAN address is listed here The current revision of the main processor and FPGA are also listed Video Test WARNING Loss of patient data Do...

Страница 99: ...ion for detailed procedures ADMIT MENU STANDARD This menu selection allows you to determine the function of the patient monitor The four variables include stationary or ambulatory telemetry patient monitoring with a monitor that always stays in one room STANDARD or a monitor that moves from room to room ROVER SOFTWARE LEVEL This menu selection displays the software feature level this monitor is us...

Страница 100: ...patient monitor SET LOCATION ID Enter the location identification for the MUSE system SET SITE NUMBER A site number may be necessary if your institution has multiple buildings A MUSE system can be used as a site number STORE DEFAULTS FOR NETWORK TRANSFER This menu selection saves the monitor defaults for transfer to another monitor with the same software patient monitor type and country code It st...

Страница 101: ...o another monitor See Chapter 6 Configuration for detailed procedures TIME AND DATE This menu selection allows changes to the time and date and may affect the time and date for the entire monitoring network Refer to Chapter 6 Configuration for detailed procedures WARNING Loss of patient history This menu should rarely be used because patient histories will be lost ...

Страница 102: ...irs are attempted These troubleshooting procedures combined with training provide the service technician with skills necessary to service and repair a patient monitor in the event of a malfunction Before starting any detailed troubleshooting complete a thorough visual inspection of the following All cable connections secure Devices properly powered Connected to a proper power source Correct Ethern...

Страница 103: ...ended If the measurements are significantly out of range have a qualified electrician repair the outlet The voltage measurements should be as follows 1 120 VAC 10 VAC between the line contact and neutral and between the line contact and ground 2 Less than 3 VAC between neutral and ground 240 VAC 50 60 Hz Use a digital voltmeter set to measure at least 300 VAC to check the voltages of the NEMA 6 20...

Страница 104: ...ram rac housing The LED Troubleshooting Table explains the normal condition of the various LEDs on the pcb assemblies Do not perform the procedures listed unless you are a trained field or biomedical engineer Refer to Disassembly on page 7 7 for circuit board functions To aid in disassembly of the unit refer to Disassembly on page 7 7 WARNING Patient data loss This procedure is intended for use by...

Страница 105: ...lure Replace the power supply 4 Loose connections to power supply Tighten connections 5 Excessive loading on external connector due to malfunctioning peripheral device Isolate and replace the malfunctioning device 6 Other Replace the power supply 7 Power supply replaced but display still doesn t work Replace processor pcb Boot Code starts but Main Code does not 1 Check the boot banner for hardware...

Страница 106: ...e system 2 Make sure the alarms are enabled Consult the operator s manual for correct alarm settings 3 Verify that the alarm volume is not turned off Look in ALARM CONTROL ALARM VOL 4 Test if the speaker is functioning using the LEARN THE MONITOR menu option Patient alarms are not sounding at the desired central station The alarm function remains disabled until the patient is correctly admitted in...

Страница 107: ...tor does not match the central station care unit name then the monitor identification changes The monitor identifier shows the care unit name followed by a vertical slash and then the bed number The solution is to modify the monitor care unit name to match the central station s care unit name The remote control does not activate the display This is caused by an interruption of power to the remote ...

Страница 108: ...t LEDs on processor PCB DS15 Green Link I hear all talking on Ethernet ON when receive path is OK from hub to the Solar 8000M i M Port DS16 Green Link I hear all talking on Ethernet ON when receive path is OK from hub to the Solar 8000M i M Port DS17 Green Link I hear all talking on Ethernet ON when receive path is OK from hub to the Solar 8000M i M Port DS18 Green Link I hear all talking on Ether...

Страница 109: ...2026265 075C Solar 8000M i patient monitor 5 17 Troubleshooting Troubleshooting procedure DS1 DS2 DS3 and DS4 DS5 and DS7 DS 6 DS 19 DS 18 DS 17 DS 16 DS 15 DS14 DS13 DS12 DS9 and DS8 050B ...

Страница 110: ...nnected or flickers LOW with no Tram rac connected DS3 Green Transmit Enable I am talking on TRAM NET Flickers ON with red LED1 ON steady while graphing DS4 Yellow Error Detect I detect an error on this PCB OFF DS6 Red Rac COMM Tram rac 4 processor PCB is OK Flashes opposite DS5 twice per second DS5 Red Rac DAS Data acquisition PCB is OK Flashes opposite DS6 twice per second 1 Because I hear when ...

Страница 111: ...revision window the software update has not been activated Power cycle the monitor and view the software revisions window again If the part numbers are still missing repeat the update procedure for each missing file Waveforms do not appear at the central station If communication is corrupted do the following 1 Check all cables for a good connection 2 Ensure that the central station software is cor...

Страница 112: ...ted Replace the processor pcb See chapter 7 Upper Level Assembly Disassembly Guidelines for the Processor Board WARNING The last attempt to load Main Code failed The checksum calculated for the SDRAM did not match what was expected If this problem occurs repeatedly your monitor may need to be serviced 1 Power cycle 2 If problems persists replace processor pcb WARNING THIS VERSION OF BOOT CODE IS N...

Страница 113: ... the real time clock If problem persists and error message displays replace processor pcb EC1 displayed to the left of the ECG parameter block Perform all of the maintenance procedures listed in the Manufacturer Recommendations section of Chapter 4 Maintenance Service the PDM Error Code The Patient Data Module PDM has encountered a software hardware or both a software and hardware error Remove the...

Страница 114: ...e g July 4 0407 3 Select REVIEW ERRORS The menu provides six error log choices three for viewing error logs and three for clearing the error logs It also provides two event log choices one for viewing and one for clearing Error logs 1 Select VIEW SCOPE ERRORS to view one error in the log of errors 2 The error log in a monitor holds 100 errors that can be accessed with the NEXT or PREVIOUS command ...

Страница 115: ...in the error log select CLEAR SCOPE EVENTS menu option Be aware that once the clear menu option is executed all selected events in memory are erased Useful error data Below is sample error log followed by a description of parameters found in the error log Definition of parameters Name Description Process Name The name of the software task that was operating when the event or problem occurred Error...

Страница 116: ...fies that a network error occurred Channel Number Identifies the network channel exhibiting the error Definition of severity levels Levels Description Continue The event or error was logged the task may have or may not have been finished but the system was able to go on Most log entries will have a severity of CONTINUE Fatal The event or error was such that the task is not able to go on Recovery w...

Страница 117: ...est should only be performed by qualified service personnel 70D Rear power switch was turned off 70E Time was changed from this monitor This helps determine how the system wide time may have been altered 70F Date was changed from this monitor This helps determine how the system wide date was altered 710 Incompatible software was detected If the main processor software finds that the software opera...

Страница 118: ...ble and other event data if applicable This data can be useful in interpreting the event log Event description Event ID Bucket data Other event data Alarm Graph On Off Change 800 0 Alarm graphing turned OFF 1 Alarm graphing turned ON Graph Location Change 801 1 Manual graph location changed 2 Alarm graph location changed 3 Print window location changed 4 12 lead print location changed Text describ...

Страница 119: ...Medical staff should be prepared to cover patients in need during these periods of lost monitoring functions Requirements PC to patient monitor PC to Unity Network MC network One patient monitor software CD ROM V5 or later NOTE The netUpdate tool is only available on a software CD ROM Microsoft Windows XP 2000 or NT operating system CD ROM drive Network card Category 5 crossover patch cable pn 415...

Страница 120: ...ry 5 crossover patch cable as follows Connect to the ETHERNET port of the patient monitor if no network is used Connect to the Unity Network MC network Ethernet tap if transferring logs from other devises on the network If you connect directly to the Unity Network MC network Ethernet tap verify that the patient monitor is also connected to the Unity Network MC network Ethernet tap 3 Power on the P...

Страница 121: ...k the IP address on the PC must match the institution s Unity Network MC network domain The network domain is defined by the first numbers used in the IP address and subnet mask If the patient monitor s address does NOT begin with 126 contact the IT department for an IP address If using a PC connected to a patient monitor and the IP address does not begin with 126 change the PC IP address to match...

Страница 122: ... toolbar Windows XP Start Control Panel Administrative Tools Services Windows 2000 Start Settings Control Panel Administrative Tools Services 2 In the list of available firewall networking services double click the service e g BlackICE to be disabled 3 In the specific service e g BlackICE properties window select Stop 4 Verify that the Service status reads Stopped 5 Select OK and close all windows...

Страница 123: ...iew Always run the netUpdate exe application from the CD ROM directory Do not run the application from any other directory 1 Make sure the patient monitor is in monitoring mode 2 Insert the CD ROM into the CD ROM drive 3 Double click My Computer 4 Create a folder for the log files in the root directory Filename can not have spaces e g C logs 5 Return to My Computer and double click the CD ROM driv...

Страница 124: ...ct Get All Logs and are connected to the Unity Network logs from every device type selected will transfer 9 Select the folder you created for log files then click OK 10 The log file displays and a message at the top of the netUpdate window displays a message that the log retrieval was successful ...

Страница 125: ...2026265 075C Solar 8000M i patient monitor 6 1 6 Configuration ...

Страница 126: ...onnected to the Unity Network write down the exact care unit name from the upper left hand corner of the central station write down the bed name for the new monitor know if the monitor will be used for either stationary or ambulatory telemetry monitoring or both know if the monitor will be moved from one Ethernet connection to another Select procedures Choose and program the procedures listed belo...

Страница 127: ...to select and change each character Up to seven characters may be entered 5 Select SET UNIT NAME and press the TRIM KNOB control to exit Set bed number The bed number identifies a particular patient bed Up to five characters are used to identify bed number This number displays at the top right of the screen Access SET BED NUMBER option starting from the MAIN menu 1 Select MORE MENUS MONITOR SETUP ...

Страница 128: ...emote control is DIDCA programmed for specific monitor types The error message WARNING REMOTE MISMATCHED WITH MONITORING MODE displays if the monitor and keypad remote control do not match Access PATIENT MONITOR TYPE option starting from the MAIN menu 1 Select MORE MENUS MONITOR SETUP SERVICE MODE 2 Enter password using the TRIM KNOB control to select the day and month from monitor screen with lea...

Страница 129: ...oose the 12 lead print location from the list Communication confirmation Confirm communication across the network 1 Admit and generate a waveform at the monitor with a simulator 2 Press Graph Go Stop and observe graph output at chosen locations Problems If the writer or printer does not graph Ensure the writer or printer is turned on Check all cables for a good connection Check programmed alarms a...

Страница 130: ...e STANDARD configuration only ROVER configures the monitor to move from room to room for stationary monitoring only COMBO configures the monitor to stay in one room for both stationary and ambulatory telemetry monitoring This monitor displays all Tram module data combined with ECG data for ambulatory patients ROVER COMBO configures the monitor to move from room to room for both stationary and ambu...

Страница 131: ...b sync voltage and pulse width The Solar 8000M i patient monitor controls the analog out signal used to trigger a defibrillator Refer to the defibrillator manufacturer s manual for the required pulse amplitude and duration Use the Boot Code SERVICE MENU to configure or change the MARKER OUT signal of the DEFIB SYNC connector on a Solar ECG module Activate the Boot Code 1 Hold down NBP Go Stop and ...

Страница 132: ...try Selection above except in the Configuration Menu select 5 Set Language and select the desired language WARNING Changing the language will discharge the monitor and erase any saved monitor defaults The monitor defaults will be set to the factory defaults Calibrate touchscreen Access the CALIBRATE TOUCHSCREEN option starting from the MAIN menu 1 Select MORE MENUS MONITOR SETUP SERVICE MODE 2 Ent...

Страница 133: ...k the time and date is automatically updated from the network time WARNING Loss of patient data history Changing the time or date settings may result in the loss of patient data history If one monitor s time or date is changed all monitors on the network listen and follow suit within 3 5 seconds Changing the time base of one monitor may cause some loss of patient data history for all the monitors ...

Страница 134: ...mote control 2 Press and release the TRIM KNOB control 3 Keep holding NBP Go Stop and Zero All until the Boot Code information appears on the display 4 Select Option Menu Change Software Level Enable options For information about enabling software options see Service menu boot code on page 5 3 Transfer monitor defaults The monitor defaults set on one monitor may be transferred to another monitor p...

Страница 135: ... set up the monitor defaults you want to store Refer to Monitor Setup in the Solar 8000i Operator s Manual 2 When defaults are set start from the MAIN menu and select MORE MENUS MONITOR SETUP SERVICE MODE 3 Enter password using the TRIM KNOB control to select the day and month from monitor screen with leading zeros e g July 4 0407 4 Select MONITOR SETTINGS 5 Select STORE DEFAULTS FOR NETWORK TRANS...

Страница 136: ...P SERVICE MODE 2 Enter password using the TRIM KNOB control to select the day and month from monitor screen with leading zeros e g July 4 0407 3 Select COPY UNIT DEFAULTS 4 Select SELECT UNIT TO COPY MONITOR DEFAULTS FROM 5 From the list of units on the Unity Network choose the server monitor 6 Select SELECT BED TO COPY MONITOR DEFAULTS FROM 7 Scroll through the list of beds within the selected un...

Страница 137: ...aved on the server monitor ERROR COPYING UNIT DEFAULTS UNIT TYPE MISMATCH The server and client monitors are set to different unit types i e ADULT NEO or OR ERROR COPYING UNIT DEFAULTS SOFTWARE VERSION MISMATCH The server and client monitors have either different revisions of software or different software versions i e Solar 7 8000 vs Solar 8000i ERROR COPYING UNIT DEFAULTS DATA CORRUPTION ERROR T...

Страница 138: ... address you must first record all other Ethernet addresses used on your network to avoid duplication Fax a password request to GE Customer Relationship Center at 877 633 8181 to obtain a password You must provide your product serial number and Ethernet address The Ethernet address displays in the Boot Code banner information Activate the Boot Code program as follows 1 Hold down NBP Go Stop and Ze...

Страница 139: ...the Unity Network to meet its intended use A private LAN is not the same as a private IP address WARNING Lost Data Duplication of an internet IP address on a network will cause lost data If you change the factory assigned internet address you must first record all other internet addresses used on your network to avoid duplication Valid and invalid IP addresses Class A Class B and Class C IP addres...

Страница 140: ...pairs Convert each hexadecimal pair separately to a decimal number All internet addresses must be in decimal 3 The first number of the internet address is always 126 decimal The low three hex pairs of the IP address are the low three Ethernet address hex pairs converted to decimal Enter internet address Use the monitor SERVICE MODE menu in Main Code 1 Access the SET INTERNET ADDRESS option startin...

Страница 141: ...2026265 075C Solar 8000M i patient monitor 7 1 7 Field replaceable units ...

Страница 142: ... serviceable If you require additional information or troubleshooting assistance contact Technical Support To order parts contact Service Parts at the address or telephone number listed on the How to Reach Us page found in the front of this manual For the latest parts information including substitutions obsolescence and compatibility please visit our Parts ID Portal website at http egems gehealthc...

Страница 143: ...mbly 801674 001 28 Processor FRU kit includes PCB assembly Solar 8000M rear label Solar 8000i rear label Hardware kit See Hardware kit on page 7 4 Note The processor pcb is dual video and has dual function TRAMNET e Port connection It is compatible with all previous hardware configurations of Solar 8000M i but requires software version 5 1 2026943 008 Wire Harness Processor to Power Supply 422812 ...

Страница 144: ...EW PH M3 X 8MM 5 4 SPACER Q3 NYLON 2 5 SCREW SEMS M4 7X10 1 6 SCREW SEMS PH M3 X 6MM 3 7 SCREW SEMS PH M3 5X8 6 9 SCR SL BDGH 6 32 X 3 16 NYLON 4 10 CLAMP CABLE 1 29 SCR MACH FLHD M3 X 6MM SS N8 25 GRAY 4 30 SCR DIN912 M3 X 8MM ZINC VIBRA 7 33 GASKET EMI SHIELD FOAM NA C2 E67 3 3 1 34 GASKET EMI SHIELD FOAM NA C2 E67 9 5 1 36 GASKET EMI SHIELD FOAM NA C2 E67 12 8 1 ...

Страница 145: ...2026265 075C Solar 8000M i patient monitor 7 5 Field replaceable units Hardware kit ...

Страница 146: ...ot Cable 420857 001 10 Foot Cable 420857 002 20 Foot Cable 420857 003 50 Foot Cable 420857 004 100 Foot Cable 420857 xxx Description Item number 6 Foot Serial Cable for touchscreen interface 2006733 001 10 Foot Serial Cable for touchscreen interface 2006733 002 15 Foot Serial Cable for touchscreen interface 2006733 003 25 Foot Serial Cable for touchscreen interface 2006733 004 Description Item num...

Страница 147: ...ation but can greatly reduce the potential for failure of any electronic assemblies being serviced Discharge any static charge you may have built up before handling semiconductors or assemblies containing semiconductors A grounded antistatic wristband 3M part number 2046 or equivalent or heel strap should be worn at all times while handling or repairing assemblies containing semiconductors Use pro...

Страница 148: ... at the rear power switch and disconnect the AC power cord and all communication cables WARNING Due to possible high voltage present use an insulated screwdriver at all times Refer to the exploded view of the unit in this chapter when disassembling Reassemble the unit in reverse order of disassembly unless otherwise stated WARNING Shock hazard High voltages exist in this unit Use insulated tools R...

Страница 149: ...n Boot Code Bed number set in Main Code Unit name set in Main Code Subnet masks 1 and 2 set in Boot Code Gateway addresses 1 and 2 set in Boot Code Internet IP addresses 1 and 2 set in Boot Code Product type Unity View or Solar 8000M i set in Boot Code Line frequency 50 60 Hz set in Boot Code Sync pulse height and width set in Boot Code Hi Res trends enabled disabled set in Boot Code Software feat...

Страница 150: ...processor pcb 1 Clean or remove the old label and apply the new label from the FRU kit 2 Perform a cold start See Cold start on page 5 2 Visually verify that the performing cold start message appears to be sure the a cold start was actually performed 3 Restore the above settings 4 If the patient monitor is not on the Unity Network be sure to set the time and date WARNING Patient data loss Do not c...

Страница 151: ...ROR THE INTERNAL BATTERY THAT MAINTAINS THE MONITOR S DEFAULTS HAS FAILED SERVICE MAY BE REQUIRED When this error message appears complete the steps listed above for the processor board Then desolder the battery and solder in a new lithium battery Speaker 1 Remove the cover 2 Using a metric hex or allen wrench with ball end unscrew two screws 3 Remove speaker from chassis ...

Страница 152: ...7 12 Solar 8000M i patient monitor 2026265 075C Field replaceable units Disassembly ...

Страница 153: ...2026265 075C Solar 8000M i patient monitor A 1 A Appendix A Technical specifications ...

Страница 154: ... TRAM NET Proprietary CSMA CD protocol at 921 Kbits sec Unity Ethernet IEEE 802 3 10 BASE T LAN Processing Main processor Motorola MPC860P PowerPC 66 MHz 32 bit Graphics processor Silicon MotionGraphics Accelerator with Silicon Image Transmitter providing images on VGA and DFP displays TRAM NET communication processor Within Motorola MPC860P PowerPC 66 MHz 32 bit LAN communication processor Within...

Страница 155: ...nd digital flat panel CRT displays Display resolution 1024 x 768 Sweep speed 25 mm sec typical user has ability to vary speeds of individual traces Aspect ratio 4 to 3 fixed not user adjustable Touch screen Utilizes ELO graphics and surface acoustic wave technology Frequency response Limited by input response of data acquisition device Number of traces 1 to 8 Number of seconds trace 6 5 seconds ty...

Страница 156: ...L 60601 1 CAN CSA C22 2 No 601 1 IEC EN 60601 1 2 IEC EN 60601 1 4 IEC EN 60601 2 27 IEC EN 60601 2 30 IEC EN 60601 2 34 IEC EN 60601 2 49 IEC EN 60601 2 51 EN 12470 4 ISO 9919 ANSI AAMI SP10 ANSI AAMI EC13 ANSI AAMI EC11 CE Marking Medical Device Directive 93 42 EEC JJG 760 2003 YY 91079 1999 Solar 8000M i with Tram module and discrete modules IEC EN UL 60601 1 CAN CSA C22 2 No 601 1 IEC EN 60601...

Страница 157: ...ed lead Gain 1V 1 Ohm Range 0 4 Ohm to 10 Ohms Frequency response 0 05 Hz to 2 2 Hz Blood pressure Tram Outputs Arterial BP BP 2 BP 3 BP 4 signal ground Gain 10 mV mmHg 4 Frequency response DC to 50 Hz Range 25 mmHg to 300 mmHg 0 mmHg 0 0V 0 025V Pulse oximetry Outputs Value and waveform Value gain 10 mV saturation Value range 0 to 1 V Waveform gain Fixed Waveform range 0 to 2 5 V Frequency respon...

Страница 158: ...ure 10 C to 35 C 50 F to 95 F Operating humidity 40 to 95 noncondensing Storage temperature 10 C to 50 C 14 F to 122 F Storage humidity 0 to 95 noncondensing Physical specifications Tram rac 2 housing Height 11 9 cm 4 7 in Width 14 5 cm 5 7 in Depth 31 2 cm 12 3 in Weight 2 3 kg 5 lb Tram rac 4A housing Height 22 9 cm 9 0 in Width 14 5 cm 5 7 in Depth 31 8 cm 12 5 in without power supply 37 1 cm 1...

Страница 159: ...o S Heart rate range 30 to 300 beats per minute Input impedance Common mode 10 M at 50 60 Hz Differential 2 5 M from dc to 60 Hz Common mode rejection 90 dB minimum at 60 Hz Gain accuracy 5 diagnostic mode Linearity deviation 5 Noise 30 μV referred to input Accuracy of the input signal reproduction Overall System Error is tested using the method described in AAMI EC11 3 2 7 1 Overall System Error ...

Страница 160: ...iming based on single or multiple lead analysis Arrhythmia calls Full lethal only or no arrhythmia PVC alarm limit range 1 to 100 PVCs minute ST Segment analysis Measurement description ST segment deviation is measured and displayed for all acquired leads ST display Lead label ST deviation current complex superimposed over a reference complex J point indicator and 15 minute mini trends are shown f...

Страница 161: ...nels 2 Probe type Series 400 or 700 determined by input cable Temperature range 0 C to 45 C 32 F to 113 F Parameters displayed T1 T2 Alarms User selectable upper and lower limits for T1 T2 Accuracy 0 1 C 25 C to 45 C independent of sensor Display resolution 0 1 C Invasive blood pressure Tram 451 451M 451N 200SL 400SL 600SL 250SL 450SL 650SL and 300 only Number of channels Tram 451 451M 451N Tram 2...

Страница 162: ...A 10 Solar 8000M i patient monitor 2026265 075C Appendix A Technical specifications Tram modules and Solar parameter functionality Range 25 mmHg to 300 mmHg ...

Страница 163: ... or 0 to 40 Hz 3dB user selectable Display scale selections 0 30 0 40 0 60 0 100 0 160 0 200 0 300 mmHg Analog output 1 V 100 mmHg Alarms User selectable upper and lower limits for systolic diastolic and mean pressures range 99 to 350 mmHg Noninvasive blood pressure Tram 415 451M 451N 851 851M 851N 200SL 400SL 800A 800SL 250SL 450SL 850A 850SL and 300 only Measurement technique Oscillometric Displ...

Страница 164: ... reference prior to each cuff inflation Tubing length 12 feet adult 8 feet neonatal Automatic cuff deflation Cycle time exceeding 2 minutes 85 seconds neonatal power off or cuff pressure exceeds 290 mmHg 6 mmHg for adult 250 mmHg 6 mmHg for pediatric or 145 mmHg 5 mmHg for neonatal Cuff sizes Disposable Large adult adult small adult pediatric child and neonatal Reusable Adult thigh large adult adu...

Страница 165: ...N support Nellcor Oxismart XL probes Parameters monitored Arterial oxygen saturation SpO2 and peripheral pulse rate PPR Probe types Tram 451 851 GE Tram 451M 851M Masimo SET Tram 451N 851N Nellcor Oxismart XL Tram 451 and 851 range GE SpO2 PPR 0 to 100 30 to 300 beats per minute Tram 451M and 851M range Masimo SpO2 PPR 30 to 100 25 to 240 beats per minute Tram 451N and 851N range Nellcor SpO2 PPR ...

Страница 166: ...njectate temperature range 0 C to 30 C 32 F to 86 F Waveform display frequency response 0 to 10 Hz 3 dB Output parameters Cardiac output blood temperature injectate temperature real time cardiac output waveform Cardiac output review Accept reject individual measurements and store average Catheter sizes 5 6 7 7 5 8 French Injectate volume selection 3 5 10 cc Environmental specifications all modules...

Страница 167: ...i patient monitor A 15 Appendix A Technical specifications Dual temperature module 400 and 700 series Dual temperature module 400 and 700 series Refer to the BP CO Temperature Modules Service Manual for Technical Specifications ...

Страница 168: ...lectable Barometric pressure compensation Manual as set in monitor Technology Sensor Mainstream non dispersive infrared NDIR absorption dual wavelength ratiometric true single beam optics Calibration Simple one step calibration less than 20 seconds no calibration gases required Alarms Type As set in monitor CO2 High inspired CO2 high low expired CO2 Respiratory rate Adjustable high and low No brea...

Страница 169: ...erature 15 C to 35 C 59 F to 95 F Relative humidity 10 to 95 noncondensing Storage conditions Temperature 0 C to 50 C 32 F to 122 F Relative humidity 0 to 95 noncondensing Physical specifications Module Height 4 0 cm 1 6 in Depth 28 6 cm 11 25 in Width 11 4 cm 4 5 in Weight 0 50 kg 1 11 lb Sensor Height 4 0 cm 1 60 in Depth 2 2 cm 87 in Width 3 7 cm 1 45 in Weight 28 grams 1 oz ...

Страница 170: ...90 response to oxygen saturation step function in 5 sec Oxygen saturation computed each second from 5 second running average of light intensity measurement SvO2 accuracy 2 oxygen saturation 1 standard deviation over 40 to 100 range Stability Drift 2 oxygen saturation over 24 hours Data acquisition rate 244 sets of samples sec Alarms Type Audible and visual Environmental specifications Operating co...

Страница 171: ...s 0 to 1 0 to 200 1 1000s 0 to 5000 dyne sec cm 5 0 to 10 000 dyne sec cm 5 m2 0 to 200 g m m2 0 to 25 kg m m2 40 to 250 beats min Power requirements Maximum Power Consumption non isolated 16 5 volts 610 mA 5 0 volts 135 mA Environmental specifications Operating Conditions Refer to the monitor specifications Storage Conditions Temperature Relative Humidity 40 C to 70 C 40 F to 160 F 0 to 95 non co...

Страница 172: ...sps Transmission of EEG waveforms may be enabled or disabled with a software command Filter options for filtered EEG waveforms sent to host monitor 256 samples second only Normal mode High pass 0 25 1 0 2 0 Hz Low pass none 30 50 70 Hz Notch none 50 60 50 60 Hz Power requirements Maximum Power Consumption non isolated 16 5V 300mA 5V 2mA Environmental specifications Operating Conditions Refer to th...

Страница 173: ...A Technical specifications Patient Data Module Patient Data Module Refer to the Patient Data Module service manual for technical specifications Solar SpO2 Masimo SET module Refer to the Solar SpO2 Masimo SET Module service manual for technical specifications ...

Страница 174: ...s and patient monitoring Display functionality must be verified for proper operation with the Solar 8000M i patient monitor in the clinical environment prior to placing into service Purchaser s responsibility The display purchaser is responsible for meeting Solar 8000M i display specifications GE does not make recommendations regarding specific display models other than those it offers for sale Fo...

Страница 175: ...0M i system with a computer grade display GE is responsible for meeting the specification As a result GE will only install non medical grade displays with appropriate isolation transformers See the following information Isolation transformers Powervar has designed an isolation transformer specifically for this application Powervar headquarters listed below will process orders and drop ship to any ...

Страница 176: ...ms NOTE Multiscan displays indicate the sync rate and the refresh rate as ranges e g 20 KHz to 80 KHz rather than a single value To meet the required specifications ensure that the ranges indicated on your display include a 48 4 KHz horizontal sync rate and a 60 Hz vertical refresh rate Monitor Display Connections Captive cable or interconnect cable processor unit to display 15 pin 3 row Dsub male...

Страница 177: ...Brightness Contrast Environmental Operating Temperature 10 to 40 degrees C Humidity 10 to 95 non condensing Bandwidth 60 MHz minimum Dot Pitch 0 28 mm maximum Misconvergence 0 3 mm maximum Non linearity Geometric Distortion 2 maximum Luminance 40 foot lamberts at screen center with 20 white screen CRT Anti glare coating and transmissiveness of approximately 57 Mounts If the monitor display will be...

Страница 178: ... 1024 x 768 NOTE Multiscan displays indicate the sync rate and the refresh rate as ranges e g 20 KHz to 80 KHz rather than a single value To meet the required specifications ensure that the ranges indicated on your display include a 48 4 KHz horizontal sync rate and a 60 Hz vertical refresh rate Monitor Display Connections Captive cable or interconnect cable processor unit to display 20 pin mini D...

Страница 179: ... 70 vertical minimum Input Voltage Range 90 135 VAC 60 Hz or 180 270 VAC 50 Hz Controls On Off Brightness Environmental Operating Temperature 10 to 40 degrees C Humidity 10 to 95 non condensing Bandwidth 65 MHz minimum Contrast Ratio 300 1 typical Brightness 150 nits minimum Protective Filter Anti glare hard coat Mounts If the monitor display will be mounted it must accommodate GCX or equivalent m...

Страница 180: ...A 28 Solar 8000M i patient monitor 2026265 075C Appendix A Technical specifications Solar 8000M i display ...

Страница 181: ...2026265 075C Solar 8000M i patient monitor 8000i Patient Monitor B 1 B Appendix B Electromagnetic compatibility ...

Страница 182: ...rify normal operation in the configuration in which it is being used Guidance and manufacturer s declaration electromagnetic emissions The Solar 8000M i is intended for use in the electromagnetic environment specified below It is the responsibility of the customer or user to assure that the Solar 8000M i is used in such an environment Emissions test Compliance Electromagnetic environment guidance ...

Страница 183: ...that of a typical commercial or hospital environment 1 kV for input output lines 1 kV for input output lines Surge EN 61000 4 5 1 kV differential mode 1 kV differential mode Mains power should be that of a typical commercial or hospital environment 2 kV common mode 2 kV common mode 5 Ut 95 dip in Ut for 0 5 cycles 5 Ut 95 dip in Ut for 0 5 cycles Voltage dips short interruptions and voltage variat...

Страница 184: ...urveya should be less than the compliance level in each frequency rangeb Interference may occur in the vicinity of equipment marked with the following symbol Note 1 At 80 MHz and 800 MHz the higher frequency range applies Note 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by reflection from structures objects and people aField strengths from fixed trans...

Страница 185: ...utput power of the communications equipment For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equitation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE These guidelines may not apply in a...

Страница 186: ...G Neonatal Cable 3 6 m 12 ft Multi Link 3 Ld ECG Cable w Grabber 3 6 m 12 ft Multi Link 5 Ld ECG ESU 3 6 m 12 ft Multi Link 3 5 Ld ECG Cable 3 6 m 12 ft Multi Link 3 Ld Neonatal ECG Cable 3 6 m 12 ft Multi Link 3 Ld ECG Cable w Grabber 3 6 m 12 ft Multi Link 6 Ld ECG Cable 3 6 m 12 ft Multi Link 12SL ECG Cable 3 6 m 12 ft Multi Link 3 5 6 Adapter N A Multi Link to 300 Series 3 Ld Trunk Cable 3 m 1...

Страница 187: ... Ld Set w clips 1 5 m 60 in 300 Series 3 Ld Set w snaps 75 cm 30 in 300 Series 5 Ld Set w clips 125 cm 49 in 300 Series 5 Ld Set w clips 1 5 m 60 in 300 Series 5 Ld Set w snaps 125 cm 49 in 300 Series replacement clip lead 125 cm 49 in 300 Series replacement clip lead 1 5 m 60 in 300 Series replacement clip lead 75 cm 30 in 300 Series replacement snap lead 125 cm 49 in Invasive BP cables and trans...

Страница 188: ...20 cm 8 in Pulse oximetry cables and sensors Nellcor RECAL Interface Cable 3 6 m 12 ft Masimo SET Interface Cable 3 6 m 12 ft Masimo SET Interface Cable 3 6 m 12 ft Nellcor RCAL Interface Cable 3 6 m 12 ft Nellcor OxiMax Interface Cable 3 m 10 ft Masimo Finger Sensor Pediatric N A Masimo Finger Sensor Adult N A Masimo Sensor Finger Probe N A ICG cable ICG Patient Cable 4 8 m 15 ft BIS EEG cables B...

Страница 189: ...mperature Module N A Dual BP Module N A Tram rac Series N A Unity Network ID N A Respiratory Mechanics Module N A PRN 50 Printer N A TRAM NET Cable 7 6 m 25 ft RJ45 series Category 5 cable N A Remote Control N A Key Pad N A Serial Cable 25 ft Power Cord 2 4 m 8 ft Digital Video Cable 1 8 m 6 ft Touch Cable RS232 1 8 m 6 ft Analog Video Cable 1 8 m 6 ft 18 inch Medical Grade Flat Panel Display N A ...

Страница 190: ... compatibility Electromagnetic compatibility EMC 15 inch Medical Grade iPanel Computer N A Keyboard USB N A Mouse USB optical N A PS2 Y Adapter Cable N A Kit HHP 3800PDF USB Scanner with USB Cable N A 3800PDF Scanner Cable USB Type A 10 ft Category Description Maximum length ...

Страница 191: ......

Страница 192: ...3 5 D 79111 Freiburg Germany Tel 49 761 45 43 0 Fax 49 761 45 43 233 Asia Headquarters GE Medical Systems Information Technologies Asia GE China Co Ltd 24th Floor Shanghai MAXDO Center 8 Xing Yi Road Hong Qiao Development Zone Shanghai 200336 P R China Tel 86 21 5257 4650 Fax 86 21 5208 2008 GE Medical Systems Information Technologies a General Electric Company going to market as GE Healthcare www...