PROTEUS XR/a
GE MEDICAL SYSTEMS
Operator Manual
REV 11
DIRECTION 2259724-100
5-8
AEC Density Compensations:
The system console has five stations for density correction. Normal
density is automatically selected when AEC is on. The five stations of
density corrections are: +2, +1, 0, -1, -2.
See the table below for density change specifications.
Scale Density
Factor Density
correction
t
olerance
2
59%more than A
A * 1.26 * 1.26
+/-10% * (A * 1.26* 1.26)
1
26%more than A
A * 1.26
+/-10% * (A * 1.26)
0
A 1
--
-1
20% less than A
A / 1.26
+/-10% * (A/1.26)
-2
37% less than A
A / 1.26 /1.26
+/-10% * (A/1.26/1.26)
FOR
TRAINING
PURPOSES
ONLY!
NOTE:
Once
downloaded,
this
document
is
UNCONTROLLED,
and
therefore
may
not
be
the
latest
revision.
Always
confirm
revision
status
against
a
validated
source
(ie
CDL).