PROTEUS XR/a
GE MEDICAL SYSTEMS
Operator Manual
REV 11
DIRECTION 2259724-100
13-1
CHAPTER 13 PHYSICAL REQUIREMENTS OF ROOM
13-1 Environmental Requirements/Limitations
13-1-1 Room Climate
Relative humidity and temperature
See Table 13-1. To obtain relative humidity and temperature requirements for components
TABLE 13-1
CLIMATE REQUIREMENT (BY COMPONENT) - RELATIVE HUMIDITY AND TEMPERATURE
PRODUCT OR
RELATIVE HUMIDITY
TEMPERATURE
COMPONENT
IN-USE STORAGE
IN-USE STORAGE
MIN. MAX. MIN. MAX.
MIN. MAX. MIN. MAX.
Elevating table
10%
75%
5%
95%
+10
°
C
40
°
C
-20
°
C
+70
°
C
OTS
10%
75%
5%
95%
+10
°
C
40
°
C
-20
°
C
+70
°
C
Generator 10%
75%
5%
95%
+10
°
C
40
°
C
-20
°
C
+70
°
C
System Console
10%
75%
5%
95%
+10
°
C
40
°
C
-20
°
C
+70
°
C
Tomolink Option
10%
75%
5%
95%
+10
°
C
40
°
C
-20
°
C
+70
°
C
X-ray Tube (Rad 14)
10%
75%
5%
95%
+10
°
C
40
°
C
-20
°
C
+70
°
C
X-ray Tube (MX 100)
10%
80%
5%
95%
0
°
C
40
°
C
-40
°
C
+80
°
C
SG120 Wallstand
20%
85%
10%
95%
+10
°
C
40
°
C
-20
°
C
+70
°
C
FOR
TRAINING
PURPOSES
ONLY!
NOTE:
Once
downloaded,
this
document
is
UNCONTROLLED,
and
therefore
may
not
be
the
latest
revision.
Always
confirm
revision
status
against
a
validated
source
(ie
CDL).