PROTEUS XR/a
GE MEDICAL SYSTEMS
Operator Manual
REV 11
DIRECTION 2259724-100
13-2
Altitude and Atmospheric Pressure
See Table 13-2.
TABLE 13-2
ALTITUDE AND ATMOSPHERIC PRESSURE
PRODUCT OR
ALTITUDE
ATMOSPHERIC PRESSURE
COMPONENT
IN-USE STORAGE IN-USE STORAGE
MIN. MAX. MIN. MAX. MIN. MAX. MIN. MAX.
Elevating table
-30.5 m
(-100 ft.)
2440 m
(8005 ft.)
-30.5 m
(-100 ft.)
3048 m
(10000 ft.)
650 hPa 1060 hPa 500 hPa 1060 hPa
XT OTS
-30.5 m
(-100 ft.)
2440 m
(8005 ft.)
-30.5 m
(-100 ft.)
3048 m
(10000 ft.)
650 hPa 1060 hPa 500 hPa 1060 hPa
Generator -30.5
m
(-100 ft.)
2440 m
(8005 ft.)
-30.5 m
(-100 ft.)
3048 m
(10000 ft.)
650 hPa 1060 hPa 500 hPa 1060 hPa
PDU -30.5
m
(-100 ft.)
2440 m
(8005 ft.)
-30.5 m
(-100 ft.)
3048 m
(10000 ft.)
650 hPa 1060 hPa 500 hPa 1060 hPa
System Console
-30.5 m
(-100 ft.)
2440 m
(8005 ft.)
-30.5 m
(-100 ft.)
3048 m
(10000 ft.)
650 hPa 1060 hPa 500 hPa 1060 hPa
Tomography
Option
-30.5 m
(-100 ft.)
2440 m
(8005 ft.)
-30.5 m
(-100 ft.)
3048 m
(10000 ft.)
650 hPa 1060 hPa 500 hPa 1060 hPa
X-ray Tube (Rad
14)
-30.5 m
(-100 ft.)
2440 m
(8005 ft.)
-30.5 m
(-100 ft.)
3048 m
(10000 ft.)
650 hPa 1060 hPa 500 hPa 1060 hPa
X-ray Tube (MX
100)
-30.5 m
(-100 ft.)
2440 m
(8005 ft.)
-30.5 m
(-100 ft.)
6096 m
(20000 ft.)
650 hPa 1060 hPa 500 hPa 1060 hPa
SG120 Wallstand
-30.5 m
(-100 ft.)
2440 m
(8005 ft.)
-30.5 m
(-100 ft.)
3000 m
(9842 ft.)
650 hPa 1060 hPa 500 hPa 1060 hPa
13-2 Equipment Heat output (Dissipation)
See Table 13-3.
TABLE 13-3
HEAT OUTPUTS (BY COMPONENT)
PRODUCT OR
MAXIMUM HEAT OUTPUT (BTU/hr)
MAXIMUM HEAT OUTPUT (WATTS)
COMPONENT IN-USE
IN-USE
Table Assembly
1500
500
Generator 2036
597
PDU 500 145
Console
180
60
X-ray Tube (Rad 14)
2288
671
X-ray Tube (MX 100)
3600KHU/hr
740
FOR
TRAINING
PURPOSES
ONLY!
NOTE:
Once
downloaded,
this
document
is
UNCONTROLLED,
and
therefore
may
not
be
the
latest
revision.
Always
confirm
revision
status
against
a
validated
source
(ie
CDL).