14-6-3 Eclipse Proteus Collimator Specifications
TABLE 14-23
ECLIPSE PROTEUS COLLIAMTOR ELECTRICAL SPECIFICATIONS
Description
Specification
Dimensions (L x W x H) [mm]
330 x 210 x 180
Weight [kg]
10
±
0,5
Angle of rotation around central beam axis [°]
±
90
Maximum FOV [cm] @ SID = 100cm
50 x 50
Minimum FOV [cm] @ SID = 180cm
4.5 x 4.5
Power supply: DC input voltage [V]
24
±
10%
Max. DC input current [A]
2
AC input voltage [V]
37
±
10%
Max. AC input current [A]
Lamp type
7
Philips 12999: 12v, 50W
Temperature: Operation [°C]
5
≤
T
≤
40
Transport and storage [°C]
-40
≤
T
≤
70
Humidity: Operation [%]
10
≤
T
≤
90, non-condensing
Storage [%]
5
≤
T
≤
95, non-condensing
Pressure: Operation [hPa]
650
≤
T
≤
1013
Storage [hPa]
525
≤
T
≤
1013
Inherent filtration
1,0 mm Al at 70 kV
Leakage radiation
< 0,5 mGy/ h
Rectangular blade positioning accuracy
±2mm@SID=1m
Rectangular blade positioning repeatability
±1mm@SID=1m