PROTEUS XR/a
GE MEDICAL SYSTEMS
Operator Manual
REV 11
DIRECTION 2259724-100
14-14
14- 7 Wall Stand (GPCP No.: 2260354)
Specifications
See Table 14-24, 14-25
TABLE 14-24
WALL STAND SPECIFICATIONS
Parameter
Specification
Weight
100 kg (220 lbs.)
Vertical Travel (Film Centerline)
1310 mm (52-1/3 in.)
Vertical Travel (Centerline Minimum to Maximum)
460 mm to 1770 mm
Grid Holding Force
>196N
Wall stand panel Filtration
0.6mm Aluminium equivalency at 100 kVp
Bucky Loading
Left hand or Right hand
Cassette Load
Right Hand (Left hand load field configurable)
Power Supply
31VDC+/-15%, 1.5Amp
Object to film distance
55mm
Panel to column distance (with/without knee
spacer)
500mm/167mm
TABLE 14-25
WALL STAND BUCKY SPECIFICATIONS
Parameter Specification
Grid Size
438mm
×
479.5mm
Grid
43line/cm; 10:1 ratio; 150 cm focus
Cassette Size Sensing
18 cm (7 in.) minimum to 43 cm (17 in.) maximum