PROTEUS XR/a
GE MEDICAL SYSTEMS
Operator Manual
REV 11
DIRECTION 2259724-100
9-12
Bucky position
OTS Console position
Bucky
Angulation
Collimator
override
switch on
OTS
Console
Image
Receptor
Mode
chosen on
System
Console
Rotation
Vertical
position
Tube
angle
Alignment
with Wall
Stand
Bucky
Longitudin
al position
Vertical
position
Source-to-
Image
Distance
(SID)
(preset by
service
engineers)
1.Bucky is
in vertical
position,
deviation
within
±10°
Ensure the
switch to be
on the
AUTO mode
Wall Stand
0° or
180°
Any
position
within
vertical
traveling
range (Top
edge of
Bucky in
vertical
position:
from 24.57”
(624mm) to
72.60”
(1844
mm)).
90°±10°,
-90°±10°
Position at
OTS
transverse
detent for
Wall Stand
centerline
Position at
OTS
longitudina
l detents
for Wall
Stand
Aligned
with
SG120
Bucky
using
Collimator
light field
SIDs
(Position
with OTS
longitudin
-al detents
for Wall
Stand
when
taking
AEC
exposure)
2. Bucky is
in
horizontal
position,
deviation
within 10°
Ensure the
switch to be
on the
AUTO mode
Wall Stand
0° or
180°
Position
by
High/Low
Auto
Exposure
position
detents on
SG120 Wall
Stand
0°±10°
Position at
OTS
transverse
detent for
Wall Stand
centerline
Aligned
with Wall
Stand
Bucky
using
Collimator
light field
Position at
OTS High
Auto
Exposure
position
detent
when
SG120
Bucky
position at
High Auto
Exposure
position,
OTS Low
Auto
Exposure
position
detent
when
SG120
Bucky
position at
Low Auto
Exposure
position
High/Low
Auto
Exposure
position
SID
(Position
with OTS
High/Low
Auto
Exposure
position
detent)
FOR
TRAINING
PURPOSES
ONLY!