PROTEUS XR/a
GE MEDICAL SYSTEMS
Operator Manual
REV 11
DIRECTION 2259724-100
9-6
3. Insert the cassette support bracket into the appropriate holes.
4. Insert the cassette between the cassette clamps, resting it on the cassette
support bracket, Reposition the clamps against the cassette and close the
cassette-clamp locking assembly.
5. Insert the cassette tray into the bucky tray slot.
TO AVOID THE CASSETTE-CLAMP LOCKING ASSEMBLY CAN BE
DAMAGED, ALWAYS CLOSE IT BEFORE INSERTING THE CASSETTE TRAY
INTO THE BUCKY TRAY SLOT.
USE CAUTION TO AVOID CASSETTE SENSING ARM DEFORMATION. ITS
SHAPE IS IMPORTANT TO OBTAIN ACCURATE FIELD PLACEMENT. IF THE
ARM IS DAMAGED, IT HAS TO BE REPLACED. DO NOT TRY TO REPAIR
THE ARM IN CASE IT BECOMES DEFORMED.
Note:
Normally, it is not needed to remove completely the cassette tray from the
bucky in order to load a cassette.
Cassettes may be inserted in the tray by pulling the tray until movement is
stopped by the catch on the lower rear of the tray.
Note:
Even if it is not needed, if it is wanted to remove totally the cassette tray
from the bucky, just pull the tray out until it is stopped by the catch (located
on the upper or lower side of the tray depending if the SG120 is left or right
tray nsertion configured), then press the catch to the bottom of the tray and
keep it pressed while you slide carefully the tray out of the bucky assembly.
Cassette Support Bracket
WARNING
WARNING
FOR
TRAINING
PURPOSES
ONLY!
NOTE:
Once
downloaded,
this
document
is
UNCONTROLLED,
and
therefore
may
not
be
the
latest
revision.
Always
confirm
revision
status
against
a
validated
source
(ie
CDL).