PROTEUS XR/a
GE MEDICAL SYSTEMS
Operator Manual
REV 11
DIRECTION 2259724-100
14-8
TABLE 14-13
KVA LOAD CHARACTERISTICS 50KW
Phase
3 3 3 3 3 3
Nominal
line
Voltage(Vac)
380 400 420 440 460 480
Voltage range(Vac)
+/−
10
%
+/−
10
%
+/−
10
%
+/−
10
%
+/−
10
%
+/−
10
%
Momentary line
current(Amp)
110 105 100 95 92 88
Continuous line
current(Amp)
7 6.7 6.2 6 5.7 5.5
Power
demand(kVA) 70 70 70 70 70 70
Frequency
47 /53Hz and 57/63Hz
TABLE 14-14
MINIMUM WIRE SIZE (65 kW)
PARAMATER THREE PHASE GENERATOR - 65KW
Input Voltage 380VAC
400 VAC
415 VAC
440 VAC
460 VAC
480 VAC
Wire Size
Length
15 m (50 ft.)
16 mm
2
(#6 AWG) 16 mm
2
(#6 AWG) 16 mm
2
(#6 AWG) 16 mm
2
(#6 AWG) 16 mm
2
(#6 AWG) 16 mm
2
(#6
AWG)
30 m (100 ft.) 22 mm
2
(#5
AWG) 22 mm
2
(#5
AWG) 22 mm
2
(#5
AWG) 22 mm
2
(#5 AWG) 22 mm
2
(#5 AWG) 22 mm
2
(#5
AWG)
46 m (150 ft.) 30 mm
2
(#3
AWG) 30 mm
2
(#3
AWG) 30 mm
2
(#3
AWG) 30 mm
2
(#3 AWG) 30 mm
2
(#3 AWG) 30 mm
2
(#3
AWG)
60 m (200 ft.) 44 mm
2
(#1
AWG) 44 mm
2
(#1
AWG) 44 mm
2
(#1
AWG) 44 mm
2
(#1 AWG) 44 mm
2
(#1 AWG) 44 mm
2
(#1
AWG)
TABLE 14-15
KVA LOAD CHARACTERISTICS 65KW
Phase
3 3 3 3 3 3
Nominal
line
voltage
(Vac)
380 400 420 440 460 480
Voltage range (Vac)
+/−
10
%
+/−
10
%
+/−
10
%
+/−
10
%
+/−
10
%
+/−
10
%
Momentary line current
(Amp)
147 140 133 127 122 117
Continuous line current
(Amp)
7 6.7 6.2 6 5.7 5.5
Power
demand
(kVA) 97 97 97 97 97 97
Frequency
47 /53Hz and 57/63Hz
TABLE 14-16
MINIMUM WIRE SIZE (80 kW)
PARAMATER THREE PHASE GENERATOR - 80KW
Input Voltage 380VAC
400 VAC
415 VAC
440 VAC
460 VAC
480 VAC
Wire Size
Length
15 m (50 ft.)
22 mm
2
(#5
AWG) 22 mm
2
(#5
AWG) 22 mm
2
(#5
AWG) 22 mm
2
(#5 AWG) 22 mm
2
(#5 AWG) 22 mm
2
(#5 AWG)
30 m (100 ft.) 30 mm
2
(#3
AWG) 30 mm
2
(#3
AWG) 30 mm
2
(#3
AWG) 30 mm
2
(#3 AWG) 30 mm
2
(#3 AWG) 30 mm
2
(#3 AWG)
46 m (150 ft.) 43 mm
2
(#1
AWG) 43 mm
2
(#1
AWG) 43 mm
2
(#1
AWG) 43 mm
2
(#1 AWG) 43 mm
2
(#1 AWG) 43 mm
2
(#1 AWG)
60 m (200 ft.) 54 mm
2
(#0
AWG) 54 mm
2
(#0
AWG) 54 mm
2
(#0
AWG) 54 mm
2
(#0 AWG) 54 mm
2
(#0 AWG) 54 mm
2
(#0 AWG)
TABLE 14-17
KVA LOAD CHARACTERISTICS 80KW
Phase
3 3 3 3 3 3
Nominal
line
Voltage
(Vac)
380 400 420 440 460 480
Voltage range (Vac)
+/−
10
%
+/−
10
%
+/−
10
%
+/−
10
%
+/−
10
%
+/−
10
%
Momentary line current
(Amp)
190 180 170 163 156 150
Continuous line current
(Amp)
7 6.7 6.2 6 5.7 5.5
Power
demand
(kVA)
125 125 125 125 125 125
Frequency
47 /53Hz and 57/63Hz
Note:
Wire size from Distribution Transformer to Disconnect Switch. Voltage is
three-phase measured line to line, balanced to neutral.