9-4-7 Bucky Rotation
It is possible to rotate the SG120 bucky assembly SG120 from 0º to 180º. The
rotation movement is permitted means of the lock lever located on the back side
of the bucky assembly.
Procedure
1. Release the lock lever.
2. Rotate the bucky to position.
Rotation Movement
3. Put the lock lever back into locked position.
Note:
The bucky can be rotated CW (clockwise) or CCW (counter-clockwise).
Note:
To avoid degradation of image quality and loss of bucky functionality, it is
recommended not to perform exposures with the bucky in other position
than 0º or 180º.
The bucky is designed to operate only at 0º or 180º positions, but will
function in a range within 30º of these positions. The image quality may be
degraded, however.