PROTEUS XR/a
GE MEDICAL SYSTEMS
Operator Manual
REV 11
DIRECTION 2259724-100
5-4
5-1-3 Group 3 Error Message Screen
The Error message area of the display screen displays informational
messages to inform the user of system and subsystem operational
status. In this situation, all buttons will be inactive except the “OK”
button. The following messages will appear in the error message area.
Console Message: X-ray Room Door Open
Recommended Operator Action
The door to the x-ray room is not
closed. The system will pro-hibit an
x-ray exposure until the door is
closed.
Console Message: Receptor Selection error
Recommended
Operator
Action
This error will occur when the
selected receptor is not configured
on the Jedi generator configuration
menu.
Console Message:
Error 30
Tube Spit error
Recommended
Operator
Action: The Proteus XR/a system has
detected a tube spit error. Press the
reset button and try the exposure
again. If error occurs again note the
error and call service.
Console Message:
Error 40
Rotation error
Recommended
Operator
Action: The Proteus XR/a system has
detected a rotation error. Press the
reset button and try the exposure
again. If error occurs again note the
error and call service.
Console Message:
Error 50
Heat (filament) error
Recommended
Operator
Action: The Proteus XR/a system has
detected a heat (filament) error.
Press the reset button and try the
exposure again. If error occurs
again note the error and call
service.
Console Message:
Error 60
Exposure error
Recommended
Operator
Action: The Proteus XR/a system has
detected a exposure error. Press
the reset button and try the
exposure again. If error occurs
again note the error and call
service.
Console Message:
Error 70
Power Supply error
Recommended
Operator
Action: The Proteus XR/a system has
detected a power supply error.
Press the reset button and try the
exposure again. If error occurs
again note the error and call
service.
FOR
TRAINING
PURPOSES
ONLY!
NOTE:
Once
downloaded,
this
document
is
UNCONTROLLED,
and
therefore
may
not
be
the
latest
revision.
Always
confirm
revision
status
against
a
validated
source
(ie
CDL).