PROTEUS XR/a
GE MEDICAL SYSTEMS
Operator Manual
REV 11
DIRECTION 2259724-100
14-2
14- 2 Table Specifications
See Table 14-1 for Table Specifications.
See Table 14-2 for Table Bucky Specifications.
TABLE 14-1
TABLE SPECIFICATIONS
Parameter Elevating
Tables
Table Top Length
2174.5 mm
Table Top Width
685.5 mm
Height
500
±
10mm ~ 800
±
10mm Elevating
Table Top Longitudinal Travel
+/-390
±
10mm
Table Top Transverse Travel
+/-116
±
10mm
Table Top Operating Force
<49N
Table Top Detent Force
>150N
Vertical Travel Speed (50Hz)
42
±
5mm/s
Vertical Travel Speed (60Hz)
52
±
5mm/s
Travel Stopping Distance
<10mm
Down motion prohibit Force
20
±
2kg
Table top Load (Max.)
220 kg
Table Top Filtration
<1.0 mm Aluminium equivalency at 100 kVp
Table Top to Film Distance (For Bucky)
<=76mm
Table Weight
250 kg
Power Supply
220+/-10%VAC, 50Hz/60Hz, 5Amp
TABLE 14-2
TABLE BUCKY SPECIFICATIONS
Parameter Specification
Longitudinal Travel in Table
590
±
10mm
Grid Size
438mm
×
479.5mm
Grid
36line/cm; 12:1 ratio; 100 cm focus
Cassette Size Sensing
18 cm (7 in.) minimum to 43 cm (17 in.) maximum