Optima IGS 320, Optima IGS 330

Pre-Installation Manual

 

OPERATING DOCUMENTATION

 

 

 

 

5537562-1-1EN

Revision  3

Содержание Optima IGS 320

Страница 1: ...Optima IGS 320 Optima IGS 330 Pre Installation Manual OPERATING DOCUMENTATION 5537562 1 1EN Revision 3...

Страница 2: ...Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 2...

Страница 3: ...e du an osigurati prijevod Ne poku avajte servisirati opremu ako niste u potpunosti pro itali i razumjeli ovaj servisni priru nik Zanemarite li ovo upozorenje mo e do i do ozljede davatelja usluge ope...

Страница 4: ...atuur niet te onderhouden alvorens deze onderhoudshandleiding werd geraadpleegd en begrepen is Indien deze waarschuwing niet wordt opgevolgd zou het onderhoudspersoneel de operator of een pati nt gewo...

Страница 5: ...nt des blessures dues des dangers lectriques m caniques ou autres WARNUNG DE Diese Serviceanleitung existiert nur in englischer Sprache Falls ein fremder Kundendienst eine andere Sprache ben tigt ist...

Страница 6: ...dell apparecchiatura solo dopo aver consultato il pre sente manuale ed averne compreso il contenuto Il mancato rispetto della presente avvertenza potrebbe causare lesioni all addetto al la manutenzion...

Страница 7: ...podr czni kiem serwisowym i zrozumienia go Niezastosowanie si do tego ostrze enia mo e doprowadzi do obra e serwisan ta operatora lub pacjenta w wyniku pora enia pr dem elektrycznym zagro enia mechan...

Страница 8: ...pasnosti UPOZORNENIE SK Tento n vod na obsluhu je k dispoz cii len v angli tine Ak z kazn kov poskytovate slu ieb vy aduje in jazyk ako angli tinu poskytnutie prekladate sk ch slu ieb je zodpovednos o...

Страница 9: ...ebuje priro nik v drugem jeziku mora stranka za gotoviti prevod Ne posku ajte servisirati opreme e tega priro nika niste v celoti prebrali in razu meli e tega opozorila ne upo tevate se lahko zaradi e...

Страница 10: ...Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 This page left intentionally blank 10 Important Information...

Страница 11: ...2015 Initial release of 5537562 1 1EN 204 5537562 1 1EN rev 2 June 2015 Second release of 5537562 1 1EN 206 5537562 1 1EN rev 3 July 2015 Third release of 5537562 1 1EN 202 Optima IGS 320 Optima IGS 3...

Страница 12: ...Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 This page left intentionally blank 12 Revision History...

Страница 13: ...ng Information 27 3 1 2 Detail Of Shipping Information 28 3 1 2 1 Gantry On Shipping Dolly 28 3 1 2 2 Gantry Air Shipment 29 3 1 2 3 System Control Cabinet on pallet 30 3 1 2 4 Omega Shipment 31 3 1 2...

Страница 14: ...2 4 Description of the X Ray generator 47 1 2 5 Description of the X Ray head 47 1 2 6 Description of the Imaging System 48 1 2 7 Description of the image monitors suspension 50 1 2 7 1 LCD monitors...

Страница 15: ...imension Requirements 85 2 2 Room Layout Drawings 89 2 2 1 Patient Room Layout 89 2 2 1 1 Room Layout for Optima System examples of room layouts 89 2 2 1 2 Room Layout for Optima System with Ergo susp...

Страница 16: ...Through bolts without insert 125 3 3 3 2 3 2286398 On grade Floor anchor 126 3 3 3 2 4 2285052 Above grade Floor anchor with insert 126 3 3 3 2 5 2285053 Through bolts with insert 127 3 3 3 2 6 228563...

Страница 17: ...le RF Communications Equipment IEC 60601 1 2 151 2 4 Limitations Management 151 2 5 Use Limitation 152 2 6 Installations Requirements Environment Control 153 Chapter 4 ENVIRONMENTAL REQUIREMENTS 155 1...

Страница 18: ...171 1 3 1 For USA only 171 1 3 2 Third party monitors on suspension 171 2 System Cable Information 172 2 1 Physical Runs 172 2 1 1 Physical Run Synoptic 172 2 1 2 System Core Matrix 173 2 1 3 PHYSICA...

Страница 19: ...ction 193 2 IP Addressing Process 194 Chapter 7 APPENDIX 195 1 PIST0021 US Fluoro UPS and PDB connection 195 1 3 1 Connection of the Power Distribution Box 196 1 3 2 Connection of the Fluoro UPS Optio...

Страница 20: ...Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 This page left intentionally blank 20 Table of Contents...

Страница 21: ...e documents are intended to assist the Installation Specialist and the Site Planner in properly preparing a site for the installation of this system It provides pre installation data such as site prep...

Страница 22: ...pre installation process The checklists summarize the required preparations and allow to verify the proper completion of the pre installation procedures You will find hereafter a chart of the informa...

Страница 23: ...re Installation Checklist Pre Installation Checklist Contract Changes Be sure to inform the customer that the cost of any alteration or modification not specified in the sales contract are liable to t...

Страница 24: ...gainst harmful ingress of water Ordinary equipment enclosed equipment without protection against ingress of wat er except Footswitch which is a IPX8 device protected against the effects of submersion...

Страница 25: ...arance requirements Deviations that meet installation requirements may be red lined if red lining is allowed by local code Seismic requirements are identified on construction drawings X 2 Delivery rou...

Страница 26: ...work in installation areas X 9 Mechanical supplier has been provided with a set of equipment installation draw ings for reference For California permit ted construction drawings or PMI speci fied inst...

Страница 27: ...9 3470 137 840 33 155 70 On pallet See Illustration 1 5 DL User parts 1040 41 860 33 9 680 26 8 220 100 On pallet X Ray tube housing 960 37 7 770 30 3 710 28 250 113 On pallet Chiller 4100 1200 47 2 5...

Страница 28: ...1 2 Detail Of Shipping Information 3 1 2 1 Gantry On Shipping Dolly Illustration 1 2 Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 28 3 Delivery Requirement...

Страница 29: ...3 1 2 2 Gantry Air Shipment Illustration 1 3 Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 Chapter 1 General Requirements 29...

Страница 30: ...3 1 2 3 System Control Cabinet on pallet Illustration 1 4 Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 30 3 Delivery Requirements...

Страница 31: ...3 1 2 4 Omega Shipment Illustration 1 5 Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 Chapter 1 General Requirements 31...

Страница 32: ...3 1 2 6 Other Elements Package NOTE All OEM parts are shipped inside there original boxes group as needed on pallets Illustration 1 7 Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5...

Страница 33: ...nch Spanner 46 176584P1 High Voltage Cable In stallation Laptop Computer MS DOS Windows Positioner Configura tion and Calibration Status Display Same as for LC Service Engineer s Tool Case Omega table...

Страница 34: ...ll wiring and hoses Chiller Autotransformer Coolix 4100 Phillips Flathead screwdriver Open chiller autotransformer Install wiring and hoses Optima IGS 320 Optima IGS 330 Pre Installation Manual Direct...

Страница 35: ...th The minimum door width needed to accommodate the System LC shipping dolly is 1 165 m 46 in with protective side rail 1 096 m 43 in with one protective side rail removed on site NOTE Door widths are...

Страница 36: ...1 8 3 4 2 Step Two Survey Data concerning the intended delivery route are recorded on the Route Survey in the following pages Record all loading capacities corridor widths door openings turning radii...

Страница 37: ...y hallway turn elevator obstruction etc Limitations loading capacity height width depth floor ing material radius etc Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revi...

Страница 38: ...ntered by an Installation Specialist on Site All cable and component selections must be sent at least 6 weeks before RSDD to your PMC The selection tool is available on the website http supportcentral...

Страница 39: ...siccant as well The lowest temperature e g 10 C 50 F and humidity is preferable If they are to be stored outside of their shipping box or in the inner shipping box without plastic wrapping they should...

Страница 40: ...nt the maximum shipment duration can be calculated depending on the duration of different temperature values application and using values given in the tables above 4 2 1 Fluoro UPS CE and UL The maxim...

Страница 41: ...Illustration 1 9 Gantry Omega Table Illustration 1 10 Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 Chapter 1 General Requirements 41...

Страница 42: ...Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 This page left intentionally blank 42 4 Product Storage and Handling Requirements...

Страница 43: ...IN AN ENVIRONMENT THAT MAY CONTAIN AIRBORNE PATHOGENS LIKE AN EXAM ROOM CONTROL ROOM THERE IS A RISK OF TRANSMISSION OF THESE AIRBORNE PATHOGENS FROM PATIENTS TO OTHER PATIENTS OR CLINICAL PERSONNEL N...

Страница 44: ...1 1 3 Control room VCIM DL Flat Panel DL Keyboard Monitors Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 44 1 System Components...

Страница 45: ...ding a X ray tube housing with a collimator and a Digital Detector TSSC Smart Box or Smart handle options for remote installation Illustration 2 1 L arm Pivot C arc Motorized elevator for the Digital...

Страница 46: ...country regulation i e USA and New Zealand the tube cover Spacer must be installed over the X ray tube cover 1 2 2 Description of the Patient table Optima IGS 320 Optima IGS 330 Systems can be supplie...

Страница 47: ...mounted on the Performix 160A X Ray tube housing 1 2 5 Description of the X Ray head The Optima System uses a Performix 160A X Ray tube housing X Ray tube collimator Ingrid HV tank oil water exchange...

Страница 48: ...e main components are the RTAC the DL the Firewall the HUB the KVM the 4 KVA PDU the positioner control and the Jedi generator Fluoro UPS interface box near System Control Cabinet Fluoro UPS option VC...

Страница 49: ...Illustration 2 5 System Control Cabinet Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 Chapter 2 Equipment Requirements 49...

Страница 50: ...be installed by strictly following the GEHC installation instructions The manufacturer specifically disclaims any and all liability arising out of or relating to the use or performance of the monitor...

Страница 51: ...1EN contained in the Service manual 1 2 8 Description of the Ergo monitors suspensions 1 2 8 1 Suspension System Configuration The monitor suspensions are mainly supplied with any vascular system it...

Страница 52: ...LCD suspension are pre installed with internal cable harness to run up to 3 monitors The suspension is designed to operate with any vascular system of GE Healthcare Illustration 2 9 Ergo LCD monitor l...

Страница 53: ...recommended as 3 8 m 12 6 After leaving cable track system cables go to technical room through the conduit on the ceiling ECG monitor VGA and power cable go to the EKG system through another conduit F...

Страница 54: ...ion and 1 video switcher Check Optima Installation Service Manual for switcher installation 1 2 11 CENTRICITY CA1000 option Refer to Centricity Cardiology CA 1000 V2 0 Preinstallation Guide in the OEM...

Страница 55: ...mount version without extension cable or Acist CVI pedestal version or Acist CVI table mount version 1 2 13 Fluoro UPS option The system can be protected with an optional Fluoro UPS There are two type...

Страница 56: ...Illustration 2 12 Fluoro UPS Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 56 1 System Components...

Страница 57: ...UL Model UPS Option Operating Manual SG UL Series 20 kVA CE Model UPS Option Operating Manual SitePro 20 kVA series 8 Illustration 2 13 UPS IF box Optima IGS 320 Optima IGS 330 Pre Installation Manual...

Страница 58: ...tration 2 21 Table Head Extender Illustration 2 22 Gantry and Omega IV Compact Patient Table Relative Positions Illustration 2 23 Gantry and Omega V Long Patient Relative Positions Illustration 2 24 S...

Страница 59: ...tion 2 14 Gantry Dimensions Side View Illustration 2 15 Gantry Dimensions Top view Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 Chapter 2 Equipment Requirem...

Страница 60: ...Illustration 2 16 Gantry Dimensions Front view Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 60 1 System Components...

Страница 61: ...Illustration 2 17 Omega IV Compact Patient Table Dimensions Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 Chapter 2 Equipment Requirements 61...

Страница 62: ...Illustration 2 18 Omega IV Compact Patient Table Interference Regions Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 62 1 System Components...

Страница 63: ...Illustration 2 19 Omega V Long Patient Table Dimensions Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 Chapter 2 Equipment Requirements 63...

Страница 64: ...Illustration 2 20 Omega V Long Patient Table Interference Regions Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 64 1 System Components...

Страница 65: ...Illustration 2 21 Patient Table side clearence CPR access Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 Chapter 2 Equipment Requirements 65...

Страница 66: ...Illustration 2 22 Table Head Extender Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 66 1 System Components...

Страница 67: ...ation 2 23 Gantry and Omega IV Compact Patient Table Relative Positions side view Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 Chapter 2 Equipment Requireme...

Страница 68: ...Illustration 2 24 Gantry and Omega V Long Patient Table Relative Positions side view Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 68 1 System Components...

Страница 69: ...Illustration 2 25 System Control Cabinet Dimensions Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 Chapter 2 Equipment Requirements 69...

Страница 70: ...Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 70 1 System Components...

Страница 71: ...Illustration 2 26 Dimension Diagram for X Ray Tube Chiller 4100 Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 Chapter 2 Equipment Requirements 71...

Страница 72: ...ow for maximum floor space Illustration 2 27 Floor Space Diagram for X Ray Tube Chiller 4100 NOTICE The chiller cannot be located more than 3 m 10 ft below the detector Optima IGS 320 Optima IGS 330 P...

Страница 73: ...Illustration 2 28 Detector Chiller Thermo Con Dimensions Orientation Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 Chapter 2 Equipment Requirements 73...

Страница 74: ...e clear width of the service area to insure electrical safety shall be at least 0 9 m In cases where 2 cabinets are installed face to face both sides of the access way the clear width shall be at leas...

Страница 75: ...essible for maintenance operation In front of the cabinet the clear width of the service area to insure electrical safety shall be at least 0 9 m In cases where 2 cabinets are installed face to face b...

Страница 76: ...Illustration 2 31 UPS IF Box Optional Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 76 1 System Components...

Страница 77: ...Illustration 2 32 EMI Filter Box CE Only Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 Chapter 2 Equipment Requirements 77...

Страница 78: ...Illustration 2 33 Gas box outlets Omega IV Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 78 1 System Components...

Страница 79: ...Illustration 2 34 Gas box outlets Omega V Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 Chapter 2 Equipment Requirements 79...

Страница 80: ...sions NOTE Unscrew the four feet on the bottom Use these four holes to fix the VCR The mounting holes accommodate M3 x 10 mm screws The holes are 0 4 inches 10 mm deep Optima IGS 320 Optima IGS 330 Pr...

Страница 81: ...Illustration 2 37 Ergo LCD suspension dimension Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 Chapter 2 Equipment Requirements 81...

Страница 82: ...This document Includes tabletop TSSC W CF 5447733 5458650 This document Optional Second MAC For Omega ta ble only TABLE PANNING DE VICE 5184673 2 5184673 2 NA 5m 1 4 3 System Compatibilities Cross Ref...

Страница 83: ...9 m Inboard Monitor Bridge or 7 ft 9 in 2 4 m Inboard Monitor Bridge GD60 4 GE HAR 5420084 2378538 2 2378540 2 46 182884G2 46 193992G2 2393190 1 1EN B2057AE B2057AG PRECABLED ERGO SUSPENSION IN ROOM C...

Страница 84: ...OCUMENT NUMBER CATALOG NUMBER SMART HANDLE SP 5439667 5439667 This document SMART BOX SP 5450589 5450589 This document TSSC W CF 5447733 5458650 This document Note 1 w CF means with Contour Filter and...

Страница 85: ...HOGENS FROM PATIENTS TO OTHER PATIENTS OR CLINICAL PERSONNEL NOSOCOMIAL DISEASES TO REDUCE THIS RISK THE ELECTRONIC CABINETS MUST BE INSTALLED IN A ROOM SEPARATED FROM EXAM ROOM CONTROL ROOM I E TECHN...

Страница 86: ...imum exam room lengths see different table configu rations Table 2 4 Table 2 5 and Illustration 2 38 14 ft 5 in 4 40 m 9 ft 0 in 2 74 m Note 1 With a 3 monitor frame suspension take care with the ceil...

Страница 87: ...with Head Extender see note above Cardio 1395 mm 54 9 in 5470 215 3 5470 215 3 Table 2 5 Omega V Table Configuration Minimum Room length in mm inches without Head Extender with Head Extender Angio 12...

Страница 88: ...Illustration 2 38 Minimum Room Length Dimension Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 88 2 Room Layouts...

Страница 89: ...RISK OF TRANSMISSION OF THESE AIRBORNE PATHOGENS FROM PATIENTS TO OTHER PATIENTS OR CLINICAL PERSONNEL NOSOCOMIAL DISEASES TO REDUCE THIS RISK THE ELECTRONIC CABINETS MUST BE INSTALLED IN A ROOM SEPA...

Страница 90: ...rails parallel to patient table to side of table WARNING ILLUSTRATION 2 39 IS MANDATORY WITH HOSPITAL HAVING AIR FLOW INSTALLATION Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 55375...

Страница 91: ...ustration 2 40 Monitor suspensions rails parallel to patient table over table Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 Chapter 2 Equipment Requirements...

Страница 92: ...Illustration 2 41 Monitor suspensions rails perpendicular to patient table Gantry side Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 92 2 Room Layouts...

Страница 93: ...tration 2 42 Monitor suspensions rails perpendicular to patient table table side Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 Chapter 2 Equipment Requiremen...

Страница 94: ...out for Optima System with Ergo suspension Illustration 2 43 System with Ergo suspension room layout drawing Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 94...

Страница 95: ...or Optima System example with Mobile system 9800 Illustration 2 44 Room Layout for Optima System Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 Chapter 2 Equi...

Страница 96: ...FFD Monitors put on frame with monitoring screen 5 Injector on Pedestal NOTE The LAN outlet RJ45 connector must be located less than 1 meter 3 feet from the System Control Cabinet 3 Optima IGS 320 Opt...

Страница 97: ...2 2 2 Control Room Layout Illustration 2 45 Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 Chapter 2 Equipment Requirements 97...

Страница 98: ...low see Dimension Drawings and Room Layout Considerations NOTE The LAN outlet RJ45 connector must be located less than 1 meter 3 feet from the System Control Cabinet Illustration 2 46 Optima IGS 320 O...

Страница 99: ...sufficient space around the patient table for the unimpeded conduct of CPR Cardiac Pulmonary Resuscitation With the table in this position the table must be capable of rotating 45 Clinicians at the p...

Страница 100: ...0 2 in above the floor For the Optima System the equipment are Table C arm Monitors Injector Rad Shield Patient Interface Module s Table Side User Interfaces TSUI Smart Box Table Side System Control T...

Страница 101: ...orehead and chin in place Additional adjustment maybe made to the position of the head cradle by loosening and tightening the long handled knob if desired Clear Vu Arm Support Mattress and mattress sl...

Страница 102: ...vice area dimensions shown above are minimum requirements Service areas must comply with local regulations if more stringent Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1...

Страница 103: ...E Minimum distance between Digital Detector and UPS is 3 m 118 inches Fluoro UPS option NOTICE The distance of the Detector chiller cannot exceed 3 m 1 floor in height below the detector Optima IGS 32...

Страница 104: ...ineering report The floor level cannot exceed a general levelness of 5 mm 0 2 in for any 2 meters 79 in NOTICE The floor slabs on which the equipment is to be installed must have a levelness of 1 mm 0...

Страница 105: ...d installation method for the LC Positioner or the Omega tables is through bolting The through bolting method can be used in all seismic zones If through bolting cannot be used use provided floor anch...

Страница 106: ...79 4 mm Make sure to obtain data on compression strength of the concrete before using floor anchors 3 2 2 Pan Type Floor Construction Requirement For Pan type floor construction steel channels must be...

Страница 107: ...preferred location in concrete floor In the examination room the LC Positioner is not placed on a computer floor but directly put on concrete floor the location of the cable access needs to be careful...

Страница 108: ...Illustration 2 53 Hole location in concrete floor Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 108 3 Room Structural Requirements...

Страница 109: ...pull out effort on each fixation bolts 3 2 4 Water Pipe Requirements An LC System uses a Performix 160 A X Ray tube with a recirculating chiller Two water hoses are supplied to allow water circulation...

Страница 110: ...tries depending on local regulations it may be forbidden to run electrical cables and water pipes in the same conduit In this case two separate conduits are required Optima IGS 320 Optima IGS 330 Pre...

Страница 111: ...uit location with On Grade anchor kits Illustration 2 56 Cable Curvature with Above Grade anchor kits Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 Chapter 2...

Страница 112: ...n grade cabling because of the minimum curvature constraint of 150 mm 6 the cable will have to come from the side between LC Positioner and patient table Optima IGS 320 Optima IGS 330 Pre Installation...

Страница 113: ...2 5x16 9 2410 kg m2 492 3 lb ft2 Same as LC Position er See IMPORTANT NOTICE below System Control Cabinet See Illustration System Control Cabi net Dimensions in Dimension Draw ings 1206x717 47 5x28 2...

Страница 114: ...given Exact dimensions depend on chiller manufacturer type NOTE 4 Recommended supplier 3 3 2 Positioner and Table Floor Mounting The distances between the LC Positioner and the Tables are critical fo...

Страница 115: ...Illustration 2 58 Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 Chapter 2 Equipment Requirements 115...

Страница 116: ...Illustration 2 59 Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 116 3 Room Structural Requirements...

Страница 117: ...Illustration 2 60 Metallic Conduit Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 Chapter 2 Equipment Requirements 117...

Страница 118: ...Illustration 2 61 Water Electric Separator Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 118 3 Room Structural Requirements...

Страница 119: ...ils on Floor plate or Base plates refer toIllustration 2 63 and illustration Holes location in concrete floor in Floor Requirements Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537...

Страница 120: ...and 1992 daN per bolt if 8 used 1120 daN per bolt if 10 used and 2000 daN per bolt if 8 used 4432 daN per bolt with 4 bolts 272 daN per bolt if 24 used and 2008 daN per bolt if 12 used Number of hole...

Страница 121: ...Illustration 2 63 Gantry and table mounting holes Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 Chapter 2 Equipment Requirements 121...

Страница 122: ...If there is no integrated base plate plan 2359194 delivered with S18101SG please refer to drawings 2359194 Angio and Cardio configuration Illustration 2 65 Please order locally the integrated base pla...

Страница 123: ...s select kit 2285054 Refer to Section 3 3 3 4 Angio Cardio Neuro templates 2285056 Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 Chapter 2 Equipment Requirem...

Страница 124: ...ation 1 pc Doc 5421635 1 1EN Pre Installation Kit Install Proce dure 1 pc NOTE The 12 LC base plate mounting screws are delivered with the system s LC Gantry These screws are cap screws 5166774 Screw...

Страница 125: ...2 13 Item Name Part Description Quan Notes Bolt Hex 2296892 Through Bolt M20 500 400 12 bolts for LC Positioner and 4 bolts for Table 16 pc Through Bolt mounting method hardware Washer Flat 99142204...

Страница 126: ...nted metal core 12 mm 1 m 3 3 3 2 4 2285052 Above grade Floor anchor with insert Table 2 15 Item Name Part Description Quan Notes Floor An chor 46 302265P1 Floor Anchor Bolt 5 8 10 x 6 in drilling 12...

Страница 127: ...tioner grout dam Dowel 2290937 Wood Dowel 24 mm diameter 12 pc Water Elec tric Separa tor 2268647 Manufactured part introduced in Positioner key hole to fit 1 electri cal conduit of 6 1 pc Electrical...

Страница 128: ...e floor plate is installed and is for China installation only 3 3 3 4 2285054 Templates 3 3 3 4 1 2285056 A C N Templates Table 2 19 Item Name Part Description Quan Notes A C N Tem plates 2185979 LC P...

Страница 129: ...Illustration 2 65 Base plate plan 2359194 Angio and Cardio configurations Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 Chapter 2 Equipment Requirements 129...

Страница 130: ...rized on the same table accessory rail as the injector Never install injector and radiation protection on the same table accessory rail Typical installation on the front table accessory rail is Smart...

Страница 131: ...Do not assume that any support structure is level within specified tolerances particularly after removing suspensions from an existing room For Ergo suspension Holes for diameter 12 mm 1 2 bolt are ma...

Страница 132: ...ilable For GEMS Euro Special metric bolts are not supplied Local Field Service supplies them according to the additional structure mounted on site Unistrut or Halfen parts Table 2 21 European parts fo...

Страница 133: ...G SUSPENSION Table 2 22 Stationary rail in different length Rail length mm ft A C D INBOARD RAILS 4 724 15 6 7 660 4 4 623 51 B0186JA 5 639 18 6 8 660 4 5 283 254 51 B0222JA 5 791 19 8 660 4 5 283 406...

Страница 134: ...port for Monitor Cables A cable support cable drape is provided with a LC System The cable support kit contains 1 B2054 EK Drape with 3 M Bridge on suspensions for X Ray tubes and monitors contains 8...

Страница 135: ...tem Control Cabinet the PDB Cabinet must be securely fastened to the wall to prevent them from tipping Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 Chapter...

Страница 136: ...option US version only the Fluoro UPS supplier will deliver anchoring brackets see Illustration 2 68 The bolts will be provided by the customer Consider local seismic codes when planning cabinet mount...

Страница 137: ...tion for system components System Control Cabinet Illustration 2 69 Fluoro UPS UL Illustration 2 70 Fluoro UPS CE Illustration 2 71 PDB UL Illustration 2 72 PDB CE Illustration 2 73 Tube Chiller Cooli...

Страница 138: ...Illustration 2 69 System Control Cabinet Illustration 2 70 Fluoro UPS UL Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 138 4 Mounting Data Including Seismic...

Страница 139: ...Illustration 2 71 Fluoro UPS CE Illustration 2 72 PDB UL Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 Chapter 2 Equipment Requirements 139...

Страница 140: ...Illustration 2 73 PDB CE Illustration 2 74 Tube Chiller Coolix 4100 Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 140 4 Mounting Data Including Seismic...

Страница 141: ...Illustration 2 75 Detector Conditioner Illustration 2 76 Omega IV Table Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 Chapter 2 Equipment Requirements 141...

Страница 142: ...Illustration 2 77 Omega V Long Table Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 142 4 Mounting Data Including Seismic...

Страница 143: ...Illustration 2 78 LC positioner Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 Chapter 2 Equipment Requirements 143...

Страница 144: ...Illustration 2 79 19 Desk Mounted Monitor Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 144 4 Mounting Data Including Seismic...

Страница 145: ...Illustration 2 80 Mavig Overhead Flat Panel Suspension Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 Chapter 2 Equipment Requirements 145...

Страница 146: ...Illustration 2 81 Ergo Overhead Flat Panel Suspension Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 146 4 Mounting Data Including Seismic...

Страница 147: ...tions may be required The General Electric Company does not make recommendations regarding radiation protection It is the purchasers responsibility to consult a radiation physicist for advise on radia...

Страница 148: ...its The system uses Radio Frequency energy only for its internal function There fore its Radio Frequency emissions are very low and are not likely to cause any interference in nearby electronic equipm...

Страница 149: ...ins power quality is that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply in put lines IEC 61000 4 11 0 Un for 5 sec 0 Un for 5...

Страница 150: ...arge Permanently In stalled Medical Equip ment E1 3 V m See statement be low for Large Per manently Installed Medical Equipment NOTE Field strengths from fixed transmitters such as base stations for c...

Страница 151: ...d 3 5 E1 P d 7 E1 P Rated Power of Transmitter watts Distance meters Distance meters Distance meters 10 mW 0 11 0 11 0 22 100 mW 0 37 0 37 0 74 1 1 1 1 1 2 3 10 3 7 3 7 7 4 100 12 12 23 For transmitte...

Страница 152: ...ase of an Optima system the Gantry digital detector shall be apart 1 meter from the generator cabinet and 1 meter apart from the analog CRT monitors These distances specifications shall minimize the l...

Страница 153: ...n panel line NOTE In order to avoid interferences the same AC power distribution panel should supply all components accessories the system subsystems as the Advantage Workstation The separated AC powe...

Страница 154: ...d outgoing power lines refer to Electrical Connections NOTE The maximum distance between the EMI filter and the PDB is 3 m NOTE The Fluoro UPS CE option requires an EMI filter box placed upstream the...

Страница 155: ...N CON DENSING TEMPERATURE IN USE IN USE See Note 2a RECOMMENDED See Note 2b MIN MAX MIN MAX MIN MAX Examination room 30 70 15 C 59 F 32 C 90 F Design for Patient Operator Com fort Technical room See N...

Страница 156: ...is critical to install the cabinets where there is no risk of flood from the air conditioner NOTICE Ensure the air outlet positions in the exam room is not in area closed to monitor suspension carria...

Страница 157: ...F PRODUCT OR COMPONENT ALTITUDE meters ATMOSPHERIC PRESSURE kPa IN USE IN USE MIN MAX MIN MAX Examination room 0 2000 79 4 106 Technical room See Notice below 0 2000 79 4 106 Control room 0 2000 N A N...

Страница 158: ...ditioner 0 21 709 0 21 709 0 21 709 0 21 709 Main disconnect panel PDB 0 4 1534 0 45 1534 0 45 1534 0 45 1534 Total for core system 4 6 15762 7 0 24020 8 9 30324 11 7 39881 Room Options 3 5 Stand by M...

Страница 159: ...cooling should enter from underneath the UPS through the appropriate aperture on the raised floor If the UPS runs in a dusty environment we recommend strongly to install filters on the air inlet of t...

Страница 160: ...HE PDB TO THE UPS AND CONFIGURING THE PDB IN BYPASS MODE GEHC IS RESPONSIBLE FOR POWERING ON THE SYSTEM WITH THE UPS IN BYPASS MODE GEDE IS RESPONSIBLE FOR UPS COMMISSIONING CAUTION The Fluoro UPS can...

Страница 161: ...A at 1 meter for Digital Detector Chiller Thermo Con Less than 50 dB A at 1 meter for a DL LCD monitor Less than 60 dB A at 1 meter for the Fluoro UPS less than 55 dB A at 20 degrees Celsius measured...

Страница 162: ...Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 This page left intentionally blank 162 4 Acoustic Specifications...

Страница 163: ...the other Power supply and ground cables provided by the customer shall be compliant with local regulations e g UL NFPA 70 CSA IEC CCC 1 1 1 Core system Table 5 1 Nominal voltage Frequency Power cons...

Страница 164: ...IOR TO EACH INSTALLATION ENSURE THAT THE ECG POWER CABLE IS CONNECTED TO A LINE THAT IS PROTECTED AGAINST SHORT CIRCUIT HAZARDS ACCORDING TO LOCAL REGULATIONS AND THAT THE ECG MONITOR IS POWERED ACCOR...

Страница 165: ...hase neutral line and protective Earth and are provided by GE 1 2 1 Schematic Illustration 5 1 Electrical and grounding schematic NOTE For the complete system connection schematic refer to the MIS Map...

Страница 166: ...PDB System Control Cabinet 3 phases Max length 12m Min diame ter AWG10 5 34 mm 4 PDB System Control Cabinet 3 phases Max length 12 m Diameter in conformity with the Max Line Impedance below For the X...

Страница 167: ...pedance table Only with CE Fluoro UPS Neutral Max length 3 m Diameter in conformity with the Max Line Impedance table Ground AWG2 35 mm and not smaller than the neutral con ductor 9 PDB Fluoro UPS 3 p...

Страница 168: ...terminals 2 3 from the Positioner Bulkhead of the System Control Cabinet shall be connected to PDB TB1 terminals 24 25 cable provided by the installer maximum length 24 m minimum diameter AWG16 The la...

Страница 169: ...t is 120V 16A max the cables diameter shall be minimum AWG12 The impedance of the ground cable between the outlet and the PDB shall be 0 4 ohms maximum This wall outlet shall comply with the following...

Страница 170: ...xcessive leakage currents The outlet shall be marked with the safety sign Illustration 5 2 such that it is visible in normal use with the maximum allowed continuous output in amperes or volt amperes a...

Страница 171: ...be line to I sense 1 3 2 Third party monitors on suspension WARNING THIRD PARTY MONITORS INSTALLED ON THE SUSPENSION MUST NOT BE POWERED BY THE SYSTEM DANGEROUS VOLTAGE MAY BE PRESENT AT THE INPUT OF...

Страница 172: ...ysical Run Synoptic Illustration 5 3 Mini Maxi Interconnection Length NOTE A 24m Emergency UPS stop cable 27561A is available in FRU Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 553...

Страница 173: ...ustration 5 4 Example of cable group schematic Cable length data is as follows Cable Length the total cable length connector to connector example above is 18 24 meters X Y used length for connection w...

Страница 174: ...Illustration 5 5 Cable Group 1 From Technical Area To Exam Room Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 174 2 System Cable Information...

Страница 175: ...Illustration 5 6 Cable Group 2 From Technical Area to Control Room Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 Chapter 5 Electrical Requirements 175...

Страница 176: ...ration 5 7 Cable Group 3 From Technical Area to Technical Area System Control Cabinet Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 176 2 System Cable Inform...

Страница 177: ...cabling two length options are available 24 meter or 36 meter 78 ft 9 in or 118 ft 1 in Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 Chapter 5 Electrical Re...

Страница 178: ...Illustration 5 8 Cable Group 4 4 with LCD Mavig Suspension Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 178 2 System Cable Information...

Страница 179: ...Illustration 5 9 Ergo Suspension System Cable Group Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 Chapter 5 Electrical Requirements 179...

Страница 180: ...L Style Voltage rating V Max Voltage carried V Cable di ameter mm Connector type Bigger Plug size mm Group n 1 From System Control Cabinet to Gantry Table 27663A 5506727 x 2789 600 13 8 DB 11 pin 34 4...

Страница 181: ...Amplimite 100 pin 84 6 27674A 5506574 x 2464 300 10 6 DB 25 pin 56 6 27722A 5404123 300 15 AMP 25 pin 55 29172A 5506917 x 2464 300 12 7 Group no 2 From System Control Cabinet to Control Booth 27676A...

Страница 182: ...85191 9 30V VGA 28048 2385192 9 30V VGA 28049 2385191 x 30V VGA 28050 2385192 x 30V VGA 28051 2385191 x 28052 2385192 x 28053 2385191 x 28054 2385192 x 28066 5314764 7 1354 30V BNC 14 4 28067 5314765...

Страница 183: ...2 2 2789 30 DB 9pin 28173 2385213 2 2789 30 DB 9pin 28186 2385197 300 Power 28187 2385198 300 Power 28189 2385200 G Y 28190 2385201 G Y 28194 2385213 3 2560 30 6 6 28195 2390545 2789 300 DB 9pin 28196...

Страница 184: ...354 BNC 29183 5389527 2 2941 DB9 28186 5389524 2 28189 5389525 2 1283 28195 5389526 2 2517 DB9 50001 5406224 2 2464 Fluoro UPS Option Cable Set 28256A 5139650 28258A 5139792 28262A 5191563 28257A 5305...

Страница 185: ...a IGS 320 Optima IGS 330 MIS Charts General Guidelines Optima System introduce a new system interconnect with a star distribution for all cables from the technical area The cables are shipped on spool...

Страница 186: ...nded for cables running overhead between rooms especially when a diagonal run provides the shortest cable path Separate conduits must be used for Hospital and Fluoro UPS power wires These wires must b...

Страница 187: ...bracket Figure 1 Connectors mounted above mounting bracket Figure 2 Illustration 5 12 MEDRAD Avanta floor mounting methods For further MEDRAD Avanta floor mounting see the Installation guide MEDRAD Av...

Страница 188: ...Illustration 5 13 Cable Channeling Layout Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 188 2 System Cable Information...

Страница 189: ...he physio gases box If no conduit available between rear of table and Control Room no Remote TSUI no MacLab need to define proper cable routing or create new conduit as per PIM requirements If there i...

Страница 190: ...product components The minimum required average luminance Em shall be of 500Lx and minimum color rendering factor Ra of 80 as per IEC EN 12464 1 Light and lighting Lighting of work places Indoor work...

Страница 191: ...examination room has a window with an aperture outside of the controlled light area day light other a curtain has to maintain the light intensity under a limit fixed to 150 lux Optima IGS 320 Optima...

Страница 192: ...Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3 This page left intentionally blank 192 3 Lighting Specifications...

Страница 193: ...ailable in the Broadband Connectivity PIM available through your local GEHC sales and service representative InSite requires an Internet Address connecting it to the System This address must be availa...

Страница 194: ...sses for DL ITU Mac lab if equipped And also get the IP addresses for SCP Destination s for sending image Worklist MPPS broker Structured Dose reporting server Optima IGS 320 Optima IGS 330 Pre Instal...

Страница 195: ...installed Power cables are as follows 1 From PDB to UPS rectifier input 4x6 AWG recommended 4x3 AWG max 3ph PE 2 Bypass 5x8 AWG recommended 5x3 AWG max 3ph N PE 3 From UPS output to PDB 5x8 AWG recom...

Страница 196: ...sequence must be corrected only at the line side of the main circuit breaker CB1 1 Connect the ON OFF Remote control and Emergency Power Off devices Illustration 7 1 2 If the Fluoro UPS option is inst...

Страница 197: ...Connection of the Fluoro UPS Option Cables 1 Connect the Fluoro UPS power cables a Remove the protection panel Optima IGS 320 Optima IGS 330 Pre Installation Manual Direction 5537562 1 1EN Revision 3...

Страница 198: ...ration 7 4 d The power cable shall enter the UPS by the bottom right hand side entry i Remove RHS UPS cover remove two top screws make two bottom screws loose ii Remove bottom cover plate Optima IGS 3...