46
lowest point of the Inspiratory Limb is highly recommended.
Follow the humidifier manufacturer’s instructions for the operation of
the humidifier.
If a heated wire circuit is required, use the VersaMed inlet equipment
(Patient Wye with one-way valve and exhalation valve. Replace the
VersaMed inspiratory tubing with the heated wire tubing.
WARNING
To prevent water and/or secretions from entering the sensor
tubing, always keep the patient sensor tilted upward.
PNEUMATIC NEBULIZER
If required, a pneumatic nebulizer may be connected to the machine.
NOTE
100% oxygen is used as a driver gas for the nebulizer. Thus an external
high-pressure oxygen supply must be connected to the machine to
enable the nebulizer operation.
To activate the nebulizer you need to select the High option under
Oxygen Supply in the advance menu (See page 95)
To connect a nebulizer device:
1.
Connect the Nebulizer tube to the Nebulizer outlet on the front panel. The
Nebulizer icon is still displayed in gray at the lower bar of the display,
showing that the nebulizer is connected but not activated (Figure 17).
Figure 17: nebulizer Icon
2.
Install the Nebulizer with the required medication to the patient circuit after
the Patient Wye and the Flow Sensor.
3.
Connect the other side of the Nebulizer tube to the nebulizer.
4.
To activate the Nebulizer see page 102.
PULSE OXIMETER
The pulse oximeter and accessories are indicated for continuous non-invasive
monitoring of oxygen saturation (SpO2) and pulse rate (measured by the SpO2
sensor). The pulse oximeter and accessories are intended for use with adult and
Содержание iVent 201 1.4
Страница 1: ...GE Healthcare iVentTM201 User s Reference Manual Revision 13 ...
Страница 2: ...VersaMed a General Electric Company doing business as GE Healthcare ...
Страница 4: ......
Страница 14: ......
Страница 42: ......
Страница 68: ...56 Notes ...
Страница 70: ......
Страница 98: ...86 Notes ...
Страница 100: ......
Страница 140: ...128 Notes ...
Страница 142: ......
Страница 152: ......
Страница 154: ......
Страница 184: ......
Страница 201: ...7 Care Maintenance and Tests 189 Notes ...
Страница 202: ......
Страница 203: ...Appendix A Glossary 191 APPENDIX A GLOSSARY ...
Страница 204: ......
Страница 210: ...198 Notes ...
Страница 211: ...Appendix B Warranty 199 APPENDIX B WARRANTY ...
Страница 212: ......
Страница 215: ...Appendix C Operating Theory 203 APPENDIX C OPERATING THEORY ...
Страница 216: ......
Страница 219: ...Appendix C Operating Theory 207 ...
Страница 220: ......
Страница 221: ...Appendix C Operating Theory 209 Notes ...
Страница 222: ......
Страница 223: ...Appendix D Theory of Breath Delivery 211 APPENDIX D THEORY OF BREATH DELIVERY ...
Страница 224: ......
Страница 233: ...Appendix D Theory of Breath Delivery 221 Notes ...
Страница 234: ......
Страница 235: ...Appendix E Ventilation Modes 223 APPENDIX E VENTILATION MODES ...
Страница 236: ......
Страница 243: ...Appendix E Ventilation Modes 231 Notes ...
Страница 244: ......
Страница 246: ......
Страница 252: ...240 Notes ...
Страница 253: ...Appendix G Accessories 241 APPENDIX G ACCESSORIES ...
Страница 254: ......
Страница 276: ...264 5 Fasten the external battery using the strap belt Figure 199 Connecting the Optional Battery Bag ...
Страница 277: ...Appendix H Remote Alarm Connector 265 APPENDIX H REMOTE ALARM CONNECTOR ...
Страница 278: ......
Страница 280: ......
Страница 281: ...Appendix I Part Numbers and Accessories 269 APPENDIX I PART NUMBERS AND ACCESSORIES ...
Страница 282: ......
Страница 286: ...274 ...
Страница 289: ...Index 277 ...