5 Adaptive Bi-Level
133
4.
Accept or adjust the parameters on the pop-up window. The factory default
settings are displayed below (Table 6):
Table 6: Adaptive Bi-Level Default Settings
Respiratory Rate
= 8 bpm
IPAP
= 10 cmH
2
O
Easy Exhale™
= On
EPAP =
5
cmH
2
O
Rise Time
= Auto
I-Time
= 1.5 sec
Peak Flow Termination (Esens)
= 40%
Triggering
= -2 cmH
2
O for pressure and 2 L/min for flow
triggering.
5.
If facemask ventilation is required, apply a facemask to the patient. Connect
the ventilator breathing circuit to the facemask.
6.
Adjust the ventilator parameters for optimal ventilator patient synchrony and
adequate ventilation (See page 134).
7.
Select suitable patient alarms by proceeding to the Main Menu and selecting
the Alarm Settings option. Enter the appropriate alarms according to facility
protocol.
NOTE
Adaptive Bi-Level provides a backup respiratory rate if spontaneous patient
breathing efforts are not detected.
In cases where a severe leak is present, the Patient Disconnect alarm can be
disabled via the
Alarm Settings
screen (See page 144)
Содержание iVent 201 1.4
Страница 1: ...GE Healthcare iVentTM201 User s Reference Manual Revision 13 ...
Страница 2: ...VersaMed a General Electric Company doing business as GE Healthcare ...
Страница 4: ......
Страница 14: ......
Страница 42: ......
Страница 68: ...56 Notes ...
Страница 70: ......
Страница 98: ...86 Notes ...
Страница 100: ......
Страница 140: ...128 Notes ...
Страница 142: ......
Страница 152: ......
Страница 154: ......
Страница 184: ......
Страница 201: ...7 Care Maintenance and Tests 189 Notes ...
Страница 202: ......
Страница 203: ...Appendix A Glossary 191 APPENDIX A GLOSSARY ...
Страница 204: ......
Страница 210: ...198 Notes ...
Страница 211: ...Appendix B Warranty 199 APPENDIX B WARRANTY ...
Страница 212: ......
Страница 215: ...Appendix C Operating Theory 203 APPENDIX C OPERATING THEORY ...
Страница 216: ......
Страница 219: ...Appendix C Operating Theory 207 ...
Страница 220: ......
Страница 221: ...Appendix C Operating Theory 209 Notes ...
Страница 222: ......
Страница 223: ...Appendix D Theory of Breath Delivery 211 APPENDIX D THEORY OF BREATH DELIVERY ...
Страница 224: ......
Страница 233: ...Appendix D Theory of Breath Delivery 221 Notes ...
Страница 234: ......
Страница 235: ...Appendix E Ventilation Modes 223 APPENDIX E VENTILATION MODES ...
Страница 236: ......
Страница 243: ...Appendix E Ventilation Modes 231 Notes ...
Страница 244: ......
Страница 246: ......
Страница 252: ...240 Notes ...
Страница 253: ...Appendix G Accessories 241 APPENDIX G ACCESSORIES ...
Страница 254: ......
Страница 276: ...264 5 Fasten the external battery using the strap belt Figure 199 Connecting the Optional Battery Bag ...
Страница 277: ...Appendix H Remote Alarm Connector 265 APPENDIX H REMOTE ALARM CONNECTOR ...
Страница 278: ......
Страница 280: ......
Страница 281: ...Appendix I Part Numbers and Accessories 269 APPENDIX I PART NUMBERS AND ACCESSORIES ...
Страница 282: ......
Страница 286: ...274 ...
Страница 289: ...Index 277 ...