6 Alarms
167
ALARMS TEST
Alarms tests check the following alarms:
Tube Disconnect
Patient Disconnect
Sensor Disconnect (Open Loop)
AC Disconnect
Low O2 Pressure
Low Minute Volume
High Pressure
High O2
Low O2
Apnea
High PEEP
Auto Start
WARNING
Do not perform these tests while the patient is connected to the
machine.
To perform an alarm test:
1.
Connect the ventilator to AC power and oxygen supply according to the
iVent
TM
201 specification (See page 32).
2.
Power up the ventilator. It automatically runs a self-test.
3.
Select the default set for patient weight 70kg. Set FiO2 to 40% (if default set is
configured to a different value).
4.
Run O.V.T. (See page 60).
5.
Attach a 2L test lung with Rp20 resistance to the patient circuit through an
HME filter and start ventilation.
6.
Tube Disconnect
alarm test
. Wait for at least 30 seconds after the
ventilation starts, and disconnect the patient circuit from the ventilator
outlet port. Verify that the
Tube Disconnect
alarm is activated during the
first inhale after disconnection. Reconnect the patient circuit to the
ventilator and verify that the
Tube Disconnect
alarm disappears
automatically.
7.
Patient Disconnect alarm test
. Disconnect the test lung from the patient
circuit. Verify that the Patient Disconnect alarm is activated. In case a
remote alarm cable is connected, verify that alarm at remote station is also
activated.
Reconnect the test lung to the patient circuit, and verify that the
Patient
Disconnect
alarm disappears automatically.
Содержание iVent 201 1.4
Страница 1: ...GE Healthcare iVentTM201 User s Reference Manual Revision 13 ...
Страница 2: ...VersaMed a General Electric Company doing business as GE Healthcare ...
Страница 4: ......
Страница 14: ......
Страница 42: ......
Страница 68: ...56 Notes ...
Страница 70: ......
Страница 98: ...86 Notes ...
Страница 100: ......
Страница 140: ...128 Notes ...
Страница 142: ......
Страница 152: ......
Страница 154: ......
Страница 184: ......
Страница 201: ...7 Care Maintenance and Tests 189 Notes ...
Страница 202: ......
Страница 203: ...Appendix A Glossary 191 APPENDIX A GLOSSARY ...
Страница 204: ......
Страница 210: ...198 Notes ...
Страница 211: ...Appendix B Warranty 199 APPENDIX B WARRANTY ...
Страница 212: ......
Страница 215: ...Appendix C Operating Theory 203 APPENDIX C OPERATING THEORY ...
Страница 216: ......
Страница 219: ...Appendix C Operating Theory 207 ...
Страница 220: ......
Страница 221: ...Appendix C Operating Theory 209 Notes ...
Страница 222: ......
Страница 223: ...Appendix D Theory of Breath Delivery 211 APPENDIX D THEORY OF BREATH DELIVERY ...
Страница 224: ......
Страница 233: ...Appendix D Theory of Breath Delivery 221 Notes ...
Страница 234: ......
Страница 235: ...Appendix E Ventilation Modes 223 APPENDIX E VENTILATION MODES ...
Страница 236: ......
Страница 243: ...Appendix E Ventilation Modes 231 Notes ...
Страница 244: ......
Страница 246: ......
Страница 252: ...240 Notes ...
Страница 253: ...Appendix G Accessories 241 APPENDIX G ACCESSORIES ...
Страница 254: ......
Страница 276: ...264 5 Fasten the external battery using the strap belt Figure 199 Connecting the Optional Battery Bag ...
Страница 277: ...Appendix H Remote Alarm Connector 265 APPENDIX H REMOTE ALARM CONNECTOR ...
Страница 278: ......
Страница 280: ......
Страница 281: ...Appendix I Part Numbers and Accessories 269 APPENDIX I PART NUMBERS AND ACCESSORIES ...
Страница 282: ......
Страница 286: ...274 ...
Страница 289: ...Index 277 ...