170
PATIENT DISCONNECT ALARM
The
Patient Disconnect
alarm is activated if low inspiratory pressure or the low
exhaled tidal volume is detected during two consecutive breaths. After alarm
activation the unit continues ventilation at the previously set parameters with the
following settings:
1.
The flow trigger sensitivity switches to OFF to eliminate false triggering.
2.
The ventilator stops oxygen control and keeps the oxygen blender in the
position it was in prior to alarm activation.
The ventilator automatically restores the previous settings when one full breath is
detected.
NOTE
During ventilation under high leak conditions (e.g., non-inflated cuff of the
endotracheal tube), a Patient Disconnect alarm is activated if the exhaled
tidal volume criteria is not met. In this case, ventilating the patient in Bi-Level
mode (See Appendix D) is recommended. To disable the Patient Disconnect
alarm see page 156.
An irregular ventilation condition such as patient coughing can cause short
periods of Patient Disconnect alarm.
PATIENT CIRCUIT FAILED ALARM
If the one-way valve at the Wye sensor becomes blocked or is otherwise
inoperable, a Patient Circuit Failed pop-up window appears:
Figure 150: The One-Way Exhalation Valve Pop-up Window
The patient circuit should be replaced at once.
Содержание iVent 201 1.4
Страница 1: ...GE Healthcare iVentTM201 User s Reference Manual Revision 13 ...
Страница 2: ...VersaMed a General Electric Company doing business as GE Healthcare ...
Страница 4: ......
Страница 14: ......
Страница 42: ......
Страница 68: ...56 Notes ...
Страница 70: ......
Страница 98: ...86 Notes ...
Страница 100: ......
Страница 140: ...128 Notes ...
Страница 142: ......
Страница 152: ......
Страница 154: ......
Страница 184: ......
Страница 201: ...7 Care Maintenance and Tests 189 Notes ...
Страница 202: ......
Страница 203: ...Appendix A Glossary 191 APPENDIX A GLOSSARY ...
Страница 204: ......
Страница 210: ...198 Notes ...
Страница 211: ...Appendix B Warranty 199 APPENDIX B WARRANTY ...
Страница 212: ......
Страница 215: ...Appendix C Operating Theory 203 APPENDIX C OPERATING THEORY ...
Страница 216: ......
Страница 219: ...Appendix C Operating Theory 207 ...
Страница 220: ......
Страница 221: ...Appendix C Operating Theory 209 Notes ...
Страница 222: ......
Страница 223: ...Appendix D Theory of Breath Delivery 211 APPENDIX D THEORY OF BREATH DELIVERY ...
Страница 224: ......
Страница 233: ...Appendix D Theory of Breath Delivery 221 Notes ...
Страница 234: ......
Страница 235: ...Appendix E Ventilation Modes 223 APPENDIX E VENTILATION MODES ...
Страница 236: ......
Страница 243: ...Appendix E Ventilation Modes 231 Notes ...
Страница 244: ......
Страница 246: ......
Страница 252: ...240 Notes ...
Страница 253: ...Appendix G Accessories 241 APPENDIX G ACCESSORIES ...
Страница 254: ......
Страница 276: ...264 5 Fasten the external battery using the strap belt Figure 199 Connecting the Optional Battery Bag ...
Страница 277: ...Appendix H Remote Alarm Connector 265 APPENDIX H REMOTE ALARM CONNECTOR ...
Страница 278: ......
Страница 280: ......
Страница 281: ...Appendix I Part Numbers and Accessories 269 APPENDIX I PART NUMBERS AND ACCESSORIES ...
Страница 282: ......
Страница 286: ...274 ...
Страница 289: ...Index 277 ...