162
Warning
Set Count or
Time of
Condition
Mode of Failure Priority Probable Cause Criterion for
Clearing Alarm
Corrective Action
Patient Circuit
Failed
10 breath
delay
Insufficient
resistance from
one way valve
High
One way valve
has been
removed from
patient circuit or
has deteriorated
(ripped or torn or
missing
membrane)
Substitute one
way valve with
new one
Replace patient
circuit
(disposable) or
replace and/or
service one way
valve (re-useable)
Low Heart
Rate
2 breaths
delay
The heart rate is
below the rate
limit
High Clinical
Condition Heart rate
exceeds the
lower alarm
limit for 1
breath
Check the patient
condition.
Low O
2
15 breath
delay
Oxygen
concentration is
below alarm
setting
High (LT) Alarm set too
high
No or low oxygen
flow to system
Calibration
required
High variability in
patient breathing
F
i
O
2
is higher
than alarm
setting for 2
breaths
Check oxygen
supply
Verify delivered
O
2
concentration
with an external
analyzer
Refer unit to
qualified service
technician for
calibration
Decrease alarm
settings
Change from
Adaptive Flow™
to fixed manual
flow
Low O
2
Pressure*
10 seconds
after the
detection of O
2
inlet pressure
below 10 PSI
External oxygen
supply pressure
is low or
disconnected
High (LT) Oxygen supply is
disconnected
Low oxygen
pressure
Nebulizer is
activated without
Oxygen supply.
N/A
Adequate O
2
pressure is
detected
Reconnect or
replace oxygen
supply source
Ensure proper
regulation of
external oxygen
pressure
* While “Low O2 Pressure” alarm condition is active the FiO2 field on the Main screen flashes red
and instead of the set FiO2, the measured FiO2 value is displayed.
Содержание iVent 201 1.4
Страница 1: ...GE Healthcare iVentTM201 User s Reference Manual Revision 13 ...
Страница 2: ...VersaMed a General Electric Company doing business as GE Healthcare ...
Страница 4: ......
Страница 14: ......
Страница 42: ......
Страница 68: ...56 Notes ...
Страница 70: ......
Страница 98: ...86 Notes ...
Страница 100: ......
Страница 140: ...128 Notes ...
Страница 142: ......
Страница 152: ......
Страница 154: ......
Страница 184: ......
Страница 201: ...7 Care Maintenance and Tests 189 Notes ...
Страница 202: ......
Страница 203: ...Appendix A Glossary 191 APPENDIX A GLOSSARY ...
Страница 204: ......
Страница 210: ...198 Notes ...
Страница 211: ...Appendix B Warranty 199 APPENDIX B WARRANTY ...
Страница 212: ......
Страница 215: ...Appendix C Operating Theory 203 APPENDIX C OPERATING THEORY ...
Страница 216: ......
Страница 219: ...Appendix C Operating Theory 207 ...
Страница 220: ......
Страница 221: ...Appendix C Operating Theory 209 Notes ...
Страница 222: ......
Страница 223: ...Appendix D Theory of Breath Delivery 211 APPENDIX D THEORY OF BREATH DELIVERY ...
Страница 224: ......
Страница 233: ...Appendix D Theory of Breath Delivery 221 Notes ...
Страница 234: ......
Страница 235: ...Appendix E Ventilation Modes 223 APPENDIX E VENTILATION MODES ...
Страница 236: ......
Страница 243: ...Appendix E Ventilation Modes 231 Notes ...
Страница 244: ......
Страница 246: ......
Страница 252: ...240 Notes ...
Страница 253: ...Appendix G Accessories 241 APPENDIX G ACCESSORIES ...
Страница 254: ......
Страница 276: ...264 5 Fasten the external battery using the strap belt Figure 199 Connecting the Optional Battery Bag ...
Страница 277: ...Appendix H Remote Alarm Connector 265 APPENDIX H REMOTE ALARM CONNECTOR ...
Страница 278: ......
Страница 280: ......
Страница 281: ...Appendix I Part Numbers and Accessories 269 APPENDIX I PART NUMBERS AND ACCESSORIES ...
Страница 282: ......
Страница 286: ...274 ...
Страница 289: ...Index 277 ...