44
CIRCUIT RESISTANCE
It is important to check the inspiratory and expiratory resistance specification of
the patient circuits used with the iVent
TM
201 ventilator to ensure they do not
exceed the following limits when adding attachments or other components or
subassemblies to the breathing circuit:
O.6 KPA (6cmH2O) at 60 L/min for adult patients.
O.5 KPA (5cmH2O) at 30 L/min for pediatric patients.
OTHER CONNECTIONS
HME
If required, a Heat and Moisture Exchanger may be placed between the flow
sensor and the patient connection.
CAUTION
A contaminated HME filter can impede the ventilator’s Patient Disconnect
alarm detection. Do not use any filter that appears to be discolored or filled
with water.
NOTE
Adding attachments, components or sub-assemblies to the
iVent
TM
201
breathing circuit may increase the pressure gradient.
MDI
To administer medication, install an adapter for MDI treatment between the flow
sensor and the patient connection.
USING THE IVENT
TM
201 WITH HEATED
HUMIDIFICATION (1.4 STEPPER BASED MODELS
ONLY)
The following procedures explain how to connect the iVent
TM
201 to a heated
humidifier.
To use the iVent
TM
201 with a heated humidifier:
1.
Disconnect the Inspiratory Limb from the ventilator outlet of the iVent201
and reconnect it to the humidifier outlet.
2.
Remove the one-way valve from the Patient Wye and reconnect the tube.
Содержание iVent 201 1.4
Страница 1: ...GE Healthcare iVentTM201 User s Reference Manual Revision 13 ...
Страница 2: ...VersaMed a General Electric Company doing business as GE Healthcare ...
Страница 4: ......
Страница 14: ......
Страница 42: ......
Страница 68: ...56 Notes ...
Страница 70: ......
Страница 98: ...86 Notes ...
Страница 100: ......
Страница 140: ...128 Notes ...
Страница 142: ......
Страница 152: ......
Страница 154: ......
Страница 184: ......
Страница 201: ...7 Care Maintenance and Tests 189 Notes ...
Страница 202: ......
Страница 203: ...Appendix A Glossary 191 APPENDIX A GLOSSARY ...
Страница 204: ......
Страница 210: ...198 Notes ...
Страница 211: ...Appendix B Warranty 199 APPENDIX B WARRANTY ...
Страница 212: ......
Страница 215: ...Appendix C Operating Theory 203 APPENDIX C OPERATING THEORY ...
Страница 216: ......
Страница 219: ...Appendix C Operating Theory 207 ...
Страница 220: ......
Страница 221: ...Appendix C Operating Theory 209 Notes ...
Страница 222: ......
Страница 223: ...Appendix D Theory of Breath Delivery 211 APPENDIX D THEORY OF BREATH DELIVERY ...
Страница 224: ......
Страница 233: ...Appendix D Theory of Breath Delivery 221 Notes ...
Страница 234: ......
Страница 235: ...Appendix E Ventilation Modes 223 APPENDIX E VENTILATION MODES ...
Страница 236: ......
Страница 243: ...Appendix E Ventilation Modes 231 Notes ...
Страница 244: ......
Страница 246: ......
Страница 252: ...240 Notes ...
Страница 253: ...Appendix G Accessories 241 APPENDIX G ACCESSORIES ...
Страница 254: ......
Страница 276: ...264 5 Fasten the external battery using the strap belt Figure 199 Connecting the Optional Battery Bag ...
Страница 277: ...Appendix H Remote Alarm Connector 265 APPENDIX H REMOTE ALARM CONNECTOR ...
Страница 278: ......
Страница 280: ......
Страница 281: ...Appendix I Part Numbers and Accessories 269 APPENDIX I PART NUMBERS AND ACCESSORIES ...
Страница 282: ......
Страница 286: ...274 ...
Страница 289: ...Index 277 ...