98
disabled, such as Low O2, High O2, and O2 sensor failed. After selecting the Low
option, a
Warning
pop-up window is displayed (Figure 86):
Figure 86: Low Option Warning
The alarm settings and readings are disabled, and the FiO2 field displays - -
If the O2 sensor fails, no alarm is displayed.
None Option
The
None
option is used when no O2 Supply is connected to the iVent
TM
201.
When this option is selected all the O2 related alarms are disabled. The alarm
settings and readings are also disabled, and the FiO2 field displays - -.
If the O2 sensor fails, no alarm is displayed.
ADAPTIVE PEAK FLOW
The rate for
Adaptive Peak Flow
can be adjusted with the
Advanced Settings
menu. (For information about
Adaptive Peak Flow
see Appendix D.)
To change the Adaptive Peak Flow setting:
1.
From the
Main
screen select
Menu – Advanced Settings
.
2.
In the
Advance Settings
screen select
Adaptive Peak Flow
. The current
status is shown in the right column of the screen (Figure 87).
Содержание iVent 201 1.4
Страница 1: ...GE Healthcare iVentTM201 User s Reference Manual Revision 13 ...
Страница 2: ...VersaMed a General Electric Company doing business as GE Healthcare ...
Страница 4: ......
Страница 14: ......
Страница 42: ......
Страница 68: ...56 Notes ...
Страница 70: ......
Страница 98: ...86 Notes ...
Страница 100: ......
Страница 140: ...128 Notes ...
Страница 142: ......
Страница 152: ......
Страница 154: ......
Страница 184: ......
Страница 201: ...7 Care Maintenance and Tests 189 Notes ...
Страница 202: ......
Страница 203: ...Appendix A Glossary 191 APPENDIX A GLOSSARY ...
Страница 204: ......
Страница 210: ...198 Notes ...
Страница 211: ...Appendix B Warranty 199 APPENDIX B WARRANTY ...
Страница 212: ......
Страница 215: ...Appendix C Operating Theory 203 APPENDIX C OPERATING THEORY ...
Страница 216: ......
Страница 219: ...Appendix C Operating Theory 207 ...
Страница 220: ......
Страница 221: ...Appendix C Operating Theory 209 Notes ...
Страница 222: ......
Страница 223: ...Appendix D Theory of Breath Delivery 211 APPENDIX D THEORY OF BREATH DELIVERY ...
Страница 224: ......
Страница 233: ...Appendix D Theory of Breath Delivery 221 Notes ...
Страница 234: ......
Страница 235: ...Appendix E Ventilation Modes 223 APPENDIX E VENTILATION MODES ...
Страница 236: ......
Страница 243: ...Appendix E Ventilation Modes 231 Notes ...
Страница 244: ......
Страница 246: ......
Страница 252: ...240 Notes ...
Страница 253: ...Appendix G Accessories 241 APPENDIX G ACCESSORIES ...
Страница 254: ......
Страница 276: ...264 5 Fasten the external battery using the strap belt Figure 199 Connecting the Optional Battery Bag ...
Страница 277: ...Appendix H Remote Alarm Connector 265 APPENDIX H REMOTE ALARM CONNECTOR ...
Страница 278: ......
Страница 280: ......
Страница 281: ...Appendix I Part Numbers and Accessories 269 APPENDIX I PART NUMBERS AND ACCESSORIES ...
Страница 282: ......
Страница 286: ...274 ...
Страница 289: ...Index 277 ...