36
be substituted and programmed according to the VersaMed Service
Manual and only with the battery specified in those notes
Full Recharge Procedure
CAUTION
“Low Battery” or “Empty Battery” alarm appears, the internal battery must
y recharged, as described below
Continued usage of the battery after the “Empty Battery” alarm appears may
disable the battery’s charging capability and/or lead to battery failure. Ventilation
parameters may not be met in this condition.
NOTE
The full recharge procedure first empties the battery of all its charge.
Charging it to capacity then allows for proper calibration of the battery
gauge.
To perform a full recharge procedure:
1.
Plug the AC power cord into the ventilator. Verify that the amber Charge LED
is lit.
2.
Allow the unit to charge for at least 10 hours.
NOTE
If the Extended battery is in use, charge the unit for at least 24 hours
3.
Switch on the ventilator. When the Weight Selection window appears, select
the 70kg patient weight setting.
4.
Set the pressure alarm to 60 (cmH20).
5.
Connect the ventilator with a patient circuit to the Rp20 resistor and test
lung.
6.
Press Start on the ventilator (See page 59) for directions on starting the
iVent
TM
201.)
7.
Adjust the tidal volume (Vt) so that a PIP (peak inspiratory pressure) of 40 is
attained on each breath – for example, a value of 850.
8.
Disconnect the power cord. The AC Disconnect alarm sounds and a pop-up
window appears. Press the red Silence button on the keypad to
remove the pop-up window and silence the alarm.
9.
Allow the ventilator to run off the battery continuously. When the Empty
Battery alarm sounds, clear it by pressing the red Silence button on the
keypad.
10.
Place the unit in Standby mode (Page 61), then switch it off. Connect the AC
power cord.
11.
Let the unit charge for at least 10 hours.
Содержание iVent 201 1.4
Страница 1: ...GE Healthcare iVentTM201 User s Reference Manual Revision 13 ...
Страница 2: ...VersaMed a General Electric Company doing business as GE Healthcare ...
Страница 4: ......
Страница 14: ......
Страница 42: ......
Страница 68: ...56 Notes ...
Страница 70: ......
Страница 98: ...86 Notes ...
Страница 100: ......
Страница 140: ...128 Notes ...
Страница 142: ......
Страница 152: ......
Страница 154: ......
Страница 184: ......
Страница 201: ...7 Care Maintenance and Tests 189 Notes ...
Страница 202: ......
Страница 203: ...Appendix A Glossary 191 APPENDIX A GLOSSARY ...
Страница 204: ......
Страница 210: ...198 Notes ...
Страница 211: ...Appendix B Warranty 199 APPENDIX B WARRANTY ...
Страница 212: ......
Страница 215: ...Appendix C Operating Theory 203 APPENDIX C OPERATING THEORY ...
Страница 216: ......
Страница 219: ...Appendix C Operating Theory 207 ...
Страница 220: ......
Страница 221: ...Appendix C Operating Theory 209 Notes ...
Страница 222: ......
Страница 223: ...Appendix D Theory of Breath Delivery 211 APPENDIX D THEORY OF BREATH DELIVERY ...
Страница 224: ......
Страница 233: ...Appendix D Theory of Breath Delivery 221 Notes ...
Страница 234: ......
Страница 235: ...Appendix E Ventilation Modes 223 APPENDIX E VENTILATION MODES ...
Страница 236: ......
Страница 243: ...Appendix E Ventilation Modes 231 Notes ...
Страница 244: ......
Страница 246: ......
Страница 252: ...240 Notes ...
Страница 253: ...Appendix G Accessories 241 APPENDIX G ACCESSORIES ...
Страница 254: ......
Страница 276: ...264 5 Fasten the external battery using the strap belt Figure 199 Connecting the Optional Battery Bag ...
Страница 277: ...Appendix H Remote Alarm Connector 265 APPENDIX H REMOTE ALARM CONNECTOR ...
Страница 278: ......
Страница 280: ......
Страница 281: ...Appendix I Part Numbers and Accessories 269 APPENDIX I PART NUMBERS AND ACCESSORIES ...
Страница 282: ......
Страница 286: ...274 ...
Страница 289: ...Index 277 ...