164
Warning
Set Count or
Time of
Condition
Mode of Failure Priority Probable Cause Criterion for
Clearing Alarm
Corrective Action
qualified
technical
personnel
Needs Cal
(Calibration)
System needs
calibration
Mid Calibrated
values
irregularity
N/A Perform
full
calibration by
qualified
technician
Over
Temperature
Immediate
Internal
temperature of
ventilator has
exceeded 80°C
High Cooling
vents
of
the ventilator are
blocked
High external
heat
Malfunction of
internal cooling
fan
Temperature
sensor failure
N/A Perform
Planned
Maintenance
procedures
Verify cooling
vents are clear
and clean
Remove or
protect unit from
external heat
source
Remove unit for
inspection and
servicing by
qualified service
personnel
Patient
Disconnect
2 breaths
delay
Loss of patient
resistance
detection
Loss of exhaled
volume
High (LT) Patient has been
disconnected
Leak in patient
circuit or in
endotracheal
tube
Improper settings
Reconnection of
patient
detected
Reconnect patient
Correct leak
Correct setting
Sensor
Disconnect
0.5 – 1 second
delay
Blockage,
disconnection or
incorrect
connection of the
3 tubes --
ventilator enters
Open Loop mode
High Blockage of
tubes
Disconnection or
incorrect
connection of the
3 tubes
Sensor tubes
detected
(Note: the mode
must be
manually
restored if
alarm occurs 3
times in 1
minute)
Reconnect tubing
Restore normal
ventilation mode
Replace patient
circuit
Содержание iVent 201 1.4
Страница 1: ...GE Healthcare iVentTM201 User s Reference Manual Revision 13 ...
Страница 2: ...VersaMed a General Electric Company doing business as GE Healthcare ...
Страница 4: ......
Страница 14: ......
Страница 42: ......
Страница 68: ...56 Notes ...
Страница 70: ......
Страница 98: ...86 Notes ...
Страница 100: ......
Страница 140: ...128 Notes ...
Страница 142: ......
Страница 152: ......
Страница 154: ......
Страница 184: ......
Страница 201: ...7 Care Maintenance and Tests 189 Notes ...
Страница 202: ......
Страница 203: ...Appendix A Glossary 191 APPENDIX A GLOSSARY ...
Страница 204: ......
Страница 210: ...198 Notes ...
Страница 211: ...Appendix B Warranty 199 APPENDIX B WARRANTY ...
Страница 212: ......
Страница 215: ...Appendix C Operating Theory 203 APPENDIX C OPERATING THEORY ...
Страница 216: ......
Страница 219: ...Appendix C Operating Theory 207 ...
Страница 220: ......
Страница 221: ...Appendix C Operating Theory 209 Notes ...
Страница 222: ......
Страница 223: ...Appendix D Theory of Breath Delivery 211 APPENDIX D THEORY OF BREATH DELIVERY ...
Страница 224: ......
Страница 233: ...Appendix D Theory of Breath Delivery 221 Notes ...
Страница 234: ......
Страница 235: ...Appendix E Ventilation Modes 223 APPENDIX E VENTILATION MODES ...
Страница 236: ......
Страница 243: ...Appendix E Ventilation Modes 231 Notes ...
Страница 244: ......
Страница 246: ......
Страница 252: ...240 Notes ...
Страница 253: ...Appendix G Accessories 241 APPENDIX G ACCESSORIES ...
Страница 254: ......
Страница 276: ...264 5 Fasten the external battery using the strap belt Figure 199 Connecting the Optional Battery Bag ...
Страница 277: ...Appendix H Remote Alarm Connector 265 APPENDIX H REMOTE ALARM CONNECTOR ...
Страница 278: ......
Страница 280: ......
Страница 281: ...Appendix I Part Numbers and Accessories 269 APPENDIX I PART NUMBERS AND ACCESSORIES ...
Страница 282: ......
Страница 286: ...274 ...
Страница 289: ...Index 277 ...