2 Setting up
35
provided by external batteries may change. To prevent switch-offs from
the external battery due to normal fluctuations in DC voltages, the
iVent
TM
201 is programmed to average the external DC voltage reading
over time. Therefore, if the user disconnects the external battery, the
ventilator interprets this as a dying battery and switches off the external
DC source for power. To re-enable input from the external battery
source, the user must leave the external battery disconnected for 40
seconds before reconnecting the external battery
INTERNAL BATTERY
The ventilator switches to an internal battery when it detects a loss of external
power. When fully charged, the internal battery can supply approximately 1 to 2
hours of power, under typical conditions and settings.
NOTE
The extended internal battery can supply approximately 3 to 4 hours of
power under typical conditions and settings
CAUTION
The iVent
TM
201 should only be used with a properly functioning battery.
The battery is automatically charged when attached to an external power source,
whether the ventilator is operating, in Standby mode, or switched off.
Batteries should undergo a full recharge procedure:
Prior to initial use
After prolonged storage
Every 3 months during normal use
If after 8 hours charging, the battery indicator fails to indicate a full charge
Indicator
Colors
Description
Remarks
Green
Battery is fully charged
Gray/Green
Battery partly discharged
Indicator moves
downward in 10%
increments
Gray / Red
Low battery
At 10%, the indicator
changes to red
Gray
Empty battery
Below 7%, the indicator
changes to gray
CAUTION
The battery for the iVent
TM
201 is a critical component. It should be
replaced only with batteries supplied by VersaMed. If the user has
been qualified by VersaMed to service the
iVent
TM
201
, the battery may
Содержание iVent 201 1.4
Страница 1: ...GE Healthcare iVentTM201 User s Reference Manual Revision 13 ...
Страница 2: ...VersaMed a General Electric Company doing business as GE Healthcare ...
Страница 4: ......
Страница 14: ......
Страница 42: ......
Страница 68: ...56 Notes ...
Страница 70: ......
Страница 98: ...86 Notes ...
Страница 100: ......
Страница 140: ...128 Notes ...
Страница 142: ......
Страница 152: ......
Страница 154: ......
Страница 184: ......
Страница 201: ...7 Care Maintenance and Tests 189 Notes ...
Страница 202: ......
Страница 203: ...Appendix A Glossary 191 APPENDIX A GLOSSARY ...
Страница 204: ......
Страница 210: ...198 Notes ...
Страница 211: ...Appendix B Warranty 199 APPENDIX B WARRANTY ...
Страница 212: ......
Страница 215: ...Appendix C Operating Theory 203 APPENDIX C OPERATING THEORY ...
Страница 216: ......
Страница 219: ...Appendix C Operating Theory 207 ...
Страница 220: ......
Страница 221: ...Appendix C Operating Theory 209 Notes ...
Страница 222: ......
Страница 223: ...Appendix D Theory of Breath Delivery 211 APPENDIX D THEORY OF BREATH DELIVERY ...
Страница 224: ......
Страница 233: ...Appendix D Theory of Breath Delivery 221 Notes ...
Страница 234: ......
Страница 235: ...Appendix E Ventilation Modes 223 APPENDIX E VENTILATION MODES ...
Страница 236: ......
Страница 243: ...Appendix E Ventilation Modes 231 Notes ...
Страница 244: ......
Страница 246: ......
Страница 252: ...240 Notes ...
Страница 253: ...Appendix G Accessories 241 APPENDIX G ACCESSORIES ...
Страница 254: ......
Страница 276: ...264 5 Fasten the external battery using the strap belt Figure 199 Connecting the Optional Battery Bag ...
Страница 277: ...Appendix H Remote Alarm Connector 265 APPENDIX H REMOTE ALARM CONNECTOR ...
Страница 278: ......
Страница 280: ......
Страница 281: ...Appendix I Part Numbers and Accessories 269 APPENDIX I PART NUMBERS AND ACCESSORIES ...
Страница 282: ......
Страница 286: ...274 ...
Страница 289: ...Index 277 ...