184
2.
If the ventilator has failed the VVT, verify that the flow sensor and
exhalation valve control tubes are adequately mated to the front panel
luer fittings, perform the calibration procedure (See Calibration, page
176) and repeat the test. If VVT failure persists, contact your authorized
service facility.
NOTE
In the case that VVT fails, on each power up the ventilator posts an alarm pop-
up window that indicates a VVT must be done before it is put into use.
CONFIGURATION SCREEN
Use the Configuration Screen to adjust:
Default startup weight
Default start screen
Default FiO2 setting
To access the Configuration screen:
1.
From the
Main
screen select
Menu - Maintenance
.
2.
A red
Caution
pop-up window appears. Select
Yes
.
3.
The
Maintenance
menu appears. Select
Configuration
Screen.
WARNING
Several functions on the Configuration Screen are designed only as
service indicators and are not accessible. Do not attempt to change
any settings on the iVent
TM
201 unless you are certain of the options
you intend to alter. Make sure settings are in accordance with
standard protocol within the hospital or patient facility.
CHOOSING A DIFFERENT STARTUP SCREEN
Use the Default start screen option if you would like the iVent
TM
201 to power up
with a different screen.
To set the startup screen:
1.
From the
Configuration
screen dial the Control Knob to select
Default
Start
screen, then press the knob to confirm.
2.
A menu pop-up window appears and shows
Main, Monitoring, and
Home Care
screens. Select the required screen, and press the knob to
confirm.
NOTE
Not all screens are operational in all iVent
TM
201 models.
Содержание iVent 201 1.4
Страница 1: ...GE Healthcare iVentTM201 User s Reference Manual Revision 13 ...
Страница 2: ...VersaMed a General Electric Company doing business as GE Healthcare ...
Страница 4: ......
Страница 14: ......
Страница 42: ......
Страница 68: ...56 Notes ...
Страница 70: ......
Страница 98: ...86 Notes ...
Страница 100: ......
Страница 140: ...128 Notes ...
Страница 142: ......
Страница 152: ......
Страница 154: ......
Страница 184: ......
Страница 201: ...7 Care Maintenance and Tests 189 Notes ...
Страница 202: ......
Страница 203: ...Appendix A Glossary 191 APPENDIX A GLOSSARY ...
Страница 204: ......
Страница 210: ...198 Notes ...
Страница 211: ...Appendix B Warranty 199 APPENDIX B WARRANTY ...
Страница 212: ......
Страница 215: ...Appendix C Operating Theory 203 APPENDIX C OPERATING THEORY ...
Страница 216: ......
Страница 219: ...Appendix C Operating Theory 207 ...
Страница 220: ......
Страница 221: ...Appendix C Operating Theory 209 Notes ...
Страница 222: ......
Страница 223: ...Appendix D Theory of Breath Delivery 211 APPENDIX D THEORY OF BREATH DELIVERY ...
Страница 224: ......
Страница 233: ...Appendix D Theory of Breath Delivery 221 Notes ...
Страница 234: ......
Страница 235: ...Appendix E Ventilation Modes 223 APPENDIX E VENTILATION MODES ...
Страница 236: ......
Страница 243: ...Appendix E Ventilation Modes 231 Notes ...
Страница 244: ......
Страница 246: ......
Страница 252: ...240 Notes ...
Страница 253: ...Appendix G Accessories 241 APPENDIX G ACCESSORIES ...
Страница 254: ......
Страница 276: ...264 5 Fasten the external battery using the strap belt Figure 199 Connecting the Optional Battery Bag ...
Страница 277: ...Appendix H Remote Alarm Connector 265 APPENDIX H REMOTE ALARM CONNECTOR ...
Страница 278: ......
Страница 280: ......
Страница 281: ...Appendix I Part Numbers and Accessories 269 APPENDIX I PART NUMBERS AND ACCESSORIES ...
Страница 282: ......
Страница 286: ...274 ...
Страница 289: ...Index 277 ...