6 Alarms
169
THE SENSOR FAILURE ALARM
When the iVent
TM
201 detects a
Sensor Disconnect
alarm or
Check Sensor
condition it enters
Open Loop
ventilation backup mode. During
Open Loop
,
ventilation is provided based on an average of the previous inhaled tidal volumes
(See Appendix E).
If the
Sensor Disconnect
alarm triggers the following message pop-up window
(Figure 149):
Figure 149: the Sensor Disconnect Pop-Up Window
Press the
Clear
button.
Check the connection of flow transducer tubing to the ventilator.
Then, check for excessive amounts of water or secretions, which can set off
the sensor alarm, in the tubing and flow transducer.
If no water or secretions are visible, wait approximately 10 breaths –
sometimes coughing can cause
Check Sensor
or
Sensor Disconnect
alarms.
If the ventilator does not automatically return to the previous mode and
setup, press the Restore option in the Backup Mode Activated message and
check the setting values: PEEP, Rise-Time (auto).
WARNING
If the Check Sensor alarm persists repeatedly, provide an alternative
source of ventilation. Then replace the patient circuit and run the O.V.T.
test.
If the “Patient Disconnect” alarm is available and switched off, the Sensor
Failure alarm and backup mode is not activated.
NOTE
If the
Sensor Disconnect
alarm is detected three (3) times within a minute,
the ventilator does not restore the previous mode automatically, but remains
in Open Loop instead. Press the
Restore
option in the
Backup Mode
Activated
pop-up window to restore the ventilator to the previous mode.
Содержание iVent 201 1.4
Страница 1: ...GE Healthcare iVentTM201 User s Reference Manual Revision 13 ...
Страница 2: ...VersaMed a General Electric Company doing business as GE Healthcare ...
Страница 4: ......
Страница 14: ......
Страница 42: ......
Страница 68: ...56 Notes ...
Страница 70: ......
Страница 98: ...86 Notes ...
Страница 100: ......
Страница 140: ...128 Notes ...
Страница 142: ......
Страница 152: ......
Страница 154: ......
Страница 184: ......
Страница 201: ...7 Care Maintenance and Tests 189 Notes ...
Страница 202: ......
Страница 203: ...Appendix A Glossary 191 APPENDIX A GLOSSARY ...
Страница 204: ......
Страница 210: ...198 Notes ...
Страница 211: ...Appendix B Warranty 199 APPENDIX B WARRANTY ...
Страница 212: ......
Страница 215: ...Appendix C Operating Theory 203 APPENDIX C OPERATING THEORY ...
Страница 216: ......
Страница 219: ...Appendix C Operating Theory 207 ...
Страница 220: ......
Страница 221: ...Appendix C Operating Theory 209 Notes ...
Страница 222: ......
Страница 223: ...Appendix D Theory of Breath Delivery 211 APPENDIX D THEORY OF BREATH DELIVERY ...
Страница 224: ......
Страница 233: ...Appendix D Theory of Breath Delivery 221 Notes ...
Страница 234: ......
Страница 235: ...Appendix E Ventilation Modes 223 APPENDIX E VENTILATION MODES ...
Страница 236: ......
Страница 243: ...Appendix E Ventilation Modes 231 Notes ...
Страница 244: ......
Страница 246: ......
Страница 252: ...240 Notes ...
Страница 253: ...Appendix G Accessories 241 APPENDIX G ACCESSORIES ...
Страница 254: ......
Страница 276: ...264 5 Fasten the external battery using the strap belt Figure 199 Connecting the Optional Battery Bag ...
Страница 277: ...Appendix H Remote Alarm Connector 265 APPENDIX H REMOTE ALARM CONNECTOR ...
Страница 278: ......
Страница 280: ......
Страница 281: ...Appendix I Part Numbers and Accessories 269 APPENDIX I PART NUMBERS AND ACCESSORIES ...
Страница 282: ......
Страница 286: ...274 ...
Страница 289: ...Index 277 ...