Filter:
Set the
Band
and the
Time Constant
for the low pass filter:
Definitions
Flow:
(Square root function)
Set the band (%FS). See
Definitions
.
Set On or Off.
Set the time constant. See
Definitions
.
Previous: Back one display.
Home: Back to start.
1
2
3
TAP
Band:
The filter compares each new value with the
previous value. If the new value is outside the
band, it is not filtered.
Time
Constant:
This sets the cut-off frequency for the filter.
Higher value (in seconds) = more filtering.
Set On or Off.
Previous: Back one display.
Home: Back to start.
1
2
3
TAP
Содержание Druck DPI 620-IS
Страница 4: ...Page 4 of 98 K0460 Issue 1 ...
Страница 5: ...GE Measurement Control Systems Druck DPI 620 IS advanced modular calibrator user manual K0460 ...
Страница 12: ...Issue 1 viii About this manual K0460 EN English ...
Страница 21: ... EN English K0460 Instrument parts accessories and options 1 5 Issue 1 ...
Страница 22: ...Issue 1 1 6 Instrument parts accessories and options K0460 EN English ...
Страница 42: ...Issue 1 2 20 Prepare the instrument K0460 EN English ...
Страница 60: ...Issue 1 5 2 Instrument communications K0460 EN English ...
Страница 64: ...Issue 1 6 4 Datalog operation K0460 EN English ...
Страница 78: ...Issue 1 9 2 Maintenance procedures K0449 EN English ...
Страница 94: ...Issue 1 10 16 Calibration procedures K0460 EN English ...
Страница 96: ...Issue 1 11 2 General specification K0460 EN English ...
Страница 97: ...Customer service Visit our web site www gesensinginspection com ...
Страница 98: ......