Issue 1
2-16 Prepare the instrument
K0460 - [EN] English
Scaling:
2.9.2 Set the Automation options (source
)
Note:
Section 2.7 (Display operation)
shows how to set and
change the values on the display.
Nudge
Span Check
Set a scale for
Point 1 and Point 2.
Set On or Off
Set a label name.
Previous: Back one display.
Home: Back to start.
1
2
3
1
TAP
Description
Process
: Nudge;
Values to set:
Step Size.
Use these buttons to increase or decrease the output
value. Increments = Step size.
Use this button (set-point) to set the output value.
1
TAP
Description
Process
: Span Check;
Values to set:
Low, High, Dwell
Use these buttons to change the value manually from
High
to
Low
.
Use the start and stop buttons to change the value
automatically from
High
to
Low
.
Dwell
sets the period between each change. The cycle
repeats automatically.
Содержание Druck DPI 620-IS
Страница 4: ...Page 4 of 98 K0460 Issue 1 ...
Страница 5: ...GE Measurement Control Systems Druck DPI 620 IS advanced modular calibrator user manual K0460 ...
Страница 12: ...Issue 1 viii About this manual K0460 EN English ...
Страница 21: ... EN English K0460 Instrument parts accessories and options 1 5 Issue 1 ...
Страница 22: ...Issue 1 1 6 Instrument parts accessories and options K0460 EN English ...
Страница 42: ...Issue 1 2 20 Prepare the instrument K0460 EN English ...
Страница 60: ...Issue 1 5 2 Instrument communications K0460 EN English ...
Страница 64: ...Issue 1 6 4 Datalog operation K0460 EN English ...
Страница 78: ...Issue 1 9 2 Maintenance procedures K0449 EN English ...
Страница 94: ...Issue 1 10 16 Calibration procedures K0460 EN English ...
Страница 96: ...Issue 1 11 2 General specification K0460 EN English ...
Страница 97: ...Customer service Visit our web site www gesensinginspection com ...
Страница 98: ......