Issue 1
3-8 Electrical operations
K0460 - [EN] English
3.
If necessary change the
Settings
and continue with the
measure operation.
Values to set
:
TC Type (Set the applicable TC).
CJ compensation (Mode: Automatic/Manual).
Automatic
uses the internal cold junction. Use
Manual
mode for an
external cold junction.
CJ Value. For
Manual
mode, set an applicable value. The
value is not used in
Automatic
mode.
Example B
1.
Set the applicable software options; see
Section 3.2.1
(Procedure overview)
.
2.
Complete the electrical connections.
3.
If necessary, change the
Source Settings
and continue with
source operation.
Values to set
:
TC Type (Set the applicable TC).
CJ compensation (Mode: Automatic/Manual).
Automatic
mode
uses the internal cold junction.
Manual
mode uses an
external cold junction.
CJ Value. For
Manual
mode, set an applicable value. The
value is not used in
Automatic
mode.
4.
Automation
: Set the applicable output value; see
Section 2.9.2
.
3.2.9 Example procedure: Switch test
CH1, P1, P2 functions use the CH2 switch connections; CH2
functions use the CH1 switch connections.
Switch operation
When setting the
Switch Test
utility on one channel, the
software automatically sets-up the other channel for the switch
connections.
Note:
If there is a measure or source function on the switch
connection channel, it is automatically disabled with the screen
message “... Function Disabled”.
An attempt to set a measure or source function on the switch
connection channel, causes a screen message “... Function not
set”.
Example
This example shows a thermocouple switch:
Содержание Druck DPI 620-IS
Страница 4: ...Page 4 of 98 K0460 Issue 1 ...
Страница 5: ...GE Measurement Control Systems Druck DPI 620 IS advanced modular calibrator user manual K0460 ...
Страница 12: ...Issue 1 viii About this manual K0460 EN English ...
Страница 21: ... EN English K0460 Instrument parts accessories and options 1 5 Issue 1 ...
Страница 22: ...Issue 1 1 6 Instrument parts accessories and options K0460 EN English ...
Страница 42: ...Issue 1 2 20 Prepare the instrument K0460 EN English ...
Страница 60: ...Issue 1 5 2 Instrument communications K0460 EN English ...
Страница 64: ...Issue 1 6 4 Datalog operation K0460 EN English ...
Страница 78: ...Issue 1 9 2 Maintenance procedures K0449 EN English ...
Страница 94: ...Issue 1 10 16 Calibration procedures K0460 EN English ...
Страница 96: ...Issue 1 11 2 General specification K0460 EN English ...
Страница 97: ...Customer service Visit our web site www gesensinginspection com ...
Страница 98: ......