[EN] English - K0460
Calibration procedures 10-7
Issue 1
10.8 Procedures
(CH1): Frequency
(measure/source)
It is only necessary to do one frequency calibration; use the
measure function or the source function.
Frequency calibration (measure function):
With the
necessary calibration equipment, use this procedure:
1.
Connect the applicable calibration equipment in
Table 10-1
;
for example: see
Section 3.2.8 (Example procedure:
Measure or source frequency signals)
, example
A
.
2.
Let the equipment get to a stable temperature (minimum: 5
minutes since the last power on).
3.
Set-up the equipment with these conditions:
4.
Use the calibration menu (
Section 10.3
) to do a one-point
calibration then do the calibration checks (
step 5
).
Frequency calibration (source function):
With the
necessary calibration equipment, use this procedure:
1.
Connect the applicable calibration equipment in
Table 10-1
;
for example: see
Section 3.2.8 (Example procedure:
Measure or source frequency signals)
, example
B
.
2.
Let the equipment get to a stable temperature (minimum: 5
minutes since the last power on).
Table 10-7:
Voltage (source) error limits
Source
V
Calibrator
uncertainty (V)
Permitted
DPI 620-IS error (V)
0
0.000004
0.00042
3
0.000010
0.0006
6
0.000018
0.00078
9
0.000027
0.00096
12
0.000036
0.0011
Signal generator:
Output = 10V, unipolar,
square wave
Frequency = 990 Hz
DPI 620-IS:
Input units = Hz
Input trigger level = 5 V
Содержание Druck DPI 620-IS
Страница 4: ...Page 4 of 98 K0460 Issue 1 ...
Страница 5: ...GE Measurement Control Systems Druck DPI 620 IS advanced modular calibrator user manual K0460 ...
Страница 12: ...Issue 1 viii About this manual K0460 EN English ...
Страница 21: ... EN English K0460 Instrument parts accessories and options 1 5 Issue 1 ...
Страница 22: ...Issue 1 1 6 Instrument parts accessories and options K0460 EN English ...
Страница 42: ...Issue 1 2 20 Prepare the instrument K0460 EN English ...
Страница 60: ...Issue 1 5 2 Instrument communications K0460 EN English ...
Страница 64: ...Issue 1 6 4 Datalog operation K0460 EN English ...
Страница 78: ...Issue 1 9 2 Maintenance procedures K0449 EN English ...
Страница 94: ...Issue 1 10 16 Calibration procedures K0460 EN English ...
Страница 96: ...Issue 1 11 2 General specification K0460 EN English ...
Страница 97: ...Customer service Visit our web site www gesensinginspection com ...
Страница 98: ......