[EN] English - K0460
HART
®
device operations 8-7
Issue 1
Preferences
•
Selecting the
icon allows the changing of the poll address,
short tag or long tag of the transmitter. The DPI 620-IS is set by
default to use Poll Address 0 (zero) only. This can be changed
by selecting the appropriate search radio button or by entering
the tag name in the search field.
8.3 HART
®
Connections
Examine the power supply connections for the transmitter before
connecting the HART
®
transmitter to the DPI 620-IS calibrator.
8.3.1 Power supply from
the calibrator
In this example, the DPI 620-IS calibrator supplies the loop power
and a 250
Ω
HART
®
resistor.
1.
Set the functions for channel 2 and HART
®
:
Measure current on
channel 2 (CH2) with
24V loop power
Range: ±55 mA
HART
®
Function is set
250
Ω
resistor is set
e
1
a
b
d
c
Set HART
Section 2.8.4
Set Current (24V); Measure
Section 2.8.4
Home display
TAP
Please note that the current
measured on Channel 2 for a
HART device should normally be
between 4 and 20 mA.
Содержание Druck DPI 620-IS
Страница 4: ...Page 4 of 98 K0460 Issue 1 ...
Страница 5: ...GE Measurement Control Systems Druck DPI 620 IS advanced modular calibrator user manual K0460 ...
Страница 12: ...Issue 1 viii About this manual K0460 EN English ...
Страница 21: ... EN English K0460 Instrument parts accessories and options 1 5 Issue 1 ...
Страница 22: ...Issue 1 1 6 Instrument parts accessories and options K0460 EN English ...
Страница 42: ...Issue 1 2 20 Prepare the instrument K0460 EN English ...
Страница 60: ...Issue 1 5 2 Instrument communications K0460 EN English ...
Страница 64: ...Issue 1 6 4 Datalog operation K0460 EN English ...
Страница 78: ...Issue 1 9 2 Maintenance procedures K0449 EN English ...
Страница 94: ...Issue 1 10 16 Calibration procedures K0460 EN English ...
Страница 96: ...Issue 1 11 2 General specification K0460 EN English ...
Страница 97: ...Customer service Visit our web site www gesensinginspection com ...
Страница 98: ......