[EN] English - K0460
Calibration procedures 10-5
Issue 1
10.6 Procedures
(CH1/CH2): DC
mV/Volts (measure)
1.
Connect the applicable calibration equipment in
Table 10-1
;
for example: see
Section 3.2.3 (Example procedure:
Measure DC voltage)
.
2.
Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3.
Use the calibration menu (
Section 10.3
) to do a three-point
calibration (-FS, Zero and +FS) for the applicable set of
ranges:
4.
To make sure the calibration is correct, select the applicable
Millivolts or Voltage (measure) function; see
Section 3.2.3
.
5.
Then apply the input values that are applicable to the
calibration:
•
mV: -2000, -1000, -200, -100, 0 (short circuit).
Then mV: 0, 100, 200, 1000, 2000.
OR
•
Volts (V): -30, -21, -20, -10, -5, 0 (short circuit).
Then volts (V): 0, 5, 10, 20, 21, 30.
6.
Make sure the error is in the specified limits (
Table 10-4
or
Table 10-5
).
mV (measure) ranges
Volts (measure) ranges
•
200 mV
•
20 V
•
2000 mV
•
30 V
Table 10-4:
Millivolts (measure) error limits
Applied
mV
Calibrator
uncertainty (mV)
Permitted
DPI 620-IS error (mV)
±2000
0.051
0.14
±1000
0.040
0.1
±200
0.0051
0.017
±100
0.0040
0.0125
0 (Short circuit)
0.0036
0.008
Содержание Druck DPI 620-IS
Страница 4: ...Page 4 of 98 K0460 Issue 1 ...
Страница 5: ...GE Measurement Control Systems Druck DPI 620 IS advanced modular calibrator user manual K0460 ...
Страница 12: ...Issue 1 viii About this manual K0460 EN English ...
Страница 21: ... EN English K0460 Instrument parts accessories and options 1 5 Issue 1 ...
Страница 22: ...Issue 1 1 6 Instrument parts accessories and options K0460 EN English ...
Страница 42: ...Issue 1 2 20 Prepare the instrument K0460 EN English ...
Страница 60: ...Issue 1 5 2 Instrument communications K0460 EN English ...
Страница 64: ...Issue 1 6 4 Datalog operation K0460 EN English ...
Страница 78: ...Issue 1 9 2 Maintenance procedures K0449 EN English ...
Страница 94: ...Issue 1 10 16 Calibration procedures K0460 EN English ...
Страница 96: ...Issue 1 11 2 General specification K0460 EN English ...
Страница 97: ...Customer service Visit our web site www gesensinginspection com ...
Страница 98: ......