[EN] English - K0460
Calibration procedures 10-3
Issue 1
When the calibration is complete, set the next calibration date;
see
Section 2.10.1 (Advanced: Calibration options)
.
10.4 Procedures
(CH1/CH2): Current
(measure)
1.
Connect the applicable calibration equipment specified in
Table 10-1
; for example: see
Section 3.2.2 (Example
procedure: Measure or source current)
.
2.
Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3.
Use the calibration menu (
Section 10.3
) to do a three-point
calibration (-FS, Zero and +FS) for each range: 20 mA and
55 mA
.
4.
To make sure the calibration is correct, select the applicable
Current (measure) function; see
Section 3.2.2
and apply
these values:
•
mA: -55, -25, -10, -5, 0 (open circuit)
Then mA: 0, 5, 10, 20, 25, 55.
5.
Make sure the error is in the specified limits (
Table 10-2
).
10.5 Procedures
(CH1/CH2): Current
(source)
1.
Connect the applicable calibration equipment in
Table 10-1
;
for example:
•
CH1/CH2 (24 mA range): see
Section 3.2.2 (Example
procedure: Measure or source current)
, example
B
.
•
CH2 (-24 mA range only): see
Section 3.2.6 (Example
procedure: Measure or source current (24V loop power))
,
example
B
.
2.
Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
(FS = full-scale)
Table 10-2:
Current (measure) error limits
Applied
mA
Calibrator
uncertainty (mA)
Permitted
DPI 620-IS error (mA)
±55
0.003
0.0055
±25
0.0025
0.0040
±20
0.00063
0.0022
±10
0.00036
0.0016
±5
0.00025
0.0013
0 (open circuit)
0.0002
0.0010
Содержание Druck DPI 620-IS
Страница 4: ...Page 4 of 98 K0460 Issue 1 ...
Страница 5: ...GE Measurement Control Systems Druck DPI 620 IS advanced modular calibrator user manual K0460 ...
Страница 12: ...Issue 1 viii About this manual K0460 EN English ...
Страница 21: ... EN English K0460 Instrument parts accessories and options 1 5 Issue 1 ...
Страница 22: ...Issue 1 1 6 Instrument parts accessories and options K0460 EN English ...
Страница 42: ...Issue 1 2 20 Prepare the instrument K0460 EN English ...
Страница 60: ...Issue 1 5 2 Instrument communications K0460 EN English ...
Страница 64: ...Issue 1 6 4 Datalog operation K0460 EN English ...
Страница 78: ...Issue 1 9 2 Maintenance procedures K0449 EN English ...
Страница 94: ...Issue 1 10 16 Calibration procedures K0460 EN English ...
Страница 96: ...Issue 1 11 2 General specification K0460 EN English ...
Страница 97: ...Customer service Visit our web site www gesensinginspection com ...
Страница 98: ......