Issue 1
2-8 Prepare the instrument
K0460 - [EN] English
2.8.1 Procedure to set the basic operations
:
Description
Sets the power off automatically after the
specified
Timeout
period. To save the battery
power, set this to
On
.
Status
: On or Off
Timeout
: 00:02:00 to 01:00:00 hours:minutes:seconds (hh:mm:ss)
Sets the backlight. Low values save battery
power; see
Section 2.5.5 (Operating time)
.
Timeout
: 00:02:00 to 02:00:00 hours:minutes:seconds (hh:mm:ss)
Intensity
: 20, 40, 60, 80, 100%
Sets the date and time. The calibration function
uses this to give calibration messages.
D
ate: day/month/year (dd/mm/yy) OR month/day/year (mm/dd/yy).
The format is factory configured.
Time
: 24 hour;
hours:minutes:seconds (hh:mm:ss)
Sets the language.
Selected
: English (Other languages to be released).
Sets the communications using the USB port.
Communications
:
Storage device, communications, active sync
.
TAP
Configuration
2
1
3
Settings
3
B
C
D
E
A
F
A
B
C
D
F
Содержание Druck DPI 620-IS
Страница 4: ...Page 4 of 98 K0460 Issue 1 ...
Страница 5: ...GE Measurement Control Systems Druck DPI 620 IS advanced modular calibrator user manual K0460 ...
Страница 12: ...Issue 1 viii About this manual K0460 EN English ...
Страница 21: ... EN English K0460 Instrument parts accessories and options 1 5 Issue 1 ...
Страница 22: ...Issue 1 1 6 Instrument parts accessories and options K0460 EN English ...
Страница 42: ...Issue 1 2 20 Prepare the instrument K0460 EN English ...
Страница 60: ...Issue 1 5 2 Instrument communications K0460 EN English ...
Страница 64: ...Issue 1 6 4 Datalog operation K0460 EN English ...
Страница 78: ...Issue 1 9 2 Maintenance procedures K0449 EN English ...
Страница 94: ...Issue 1 10 16 Calibration procedures K0460 EN English ...
Страница 96: ...Issue 1 11 2 General specification K0460 EN English ...
Страница 97: ...Customer service Visit our web site www gesensinginspection com ...
Страница 98: ......