Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
B
-2(6)
Document no
.
M1144951-004
Table 2
Guidance and manufacturer’s declaration – electromagnetic immunity
Guidance and manufacturer’s declaration – electromagnetic immunity
The S/5™ CAM/ CCCM is intended for use in the electromagnetic environment specified below. The customer or
the user of the S/5™ CAM/ CCCM should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Electrical fast
transients/bursts
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for
input/output lines
±2 kV for power
supply lines
±1 kV for input/output
lines
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common mode
±1 kV differential
mode
±2 kV common mode
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations on
power supply lines
IEC 61000-4-11
<5 %
U
T
(>95 % dip in
U
T
)
for 0.5 cycle
40 % U
T
(60 % dip in
U
T
)
for 5 cycles
70 %
U
T
(30 % dip in
U
T
)
for 25 cycles
<5 %
U
T
(>95 % dip in
U
T
)
for 5 sec
<5 %
U
T
(>95 % dip in
U
T
)
for 0.5 cycle
40 % U
T
(60 % dip in
U
T
)
for 5 cycles
70 %
U
T
(30 % dip in
U
T
)
for 25 cycles
<5 %
U
T
(>95 % dip in
U
T
)
for 5 sec
Mains power quality should be that of a
typical commercial or hospital
environment. If user of the S/5™ CAM/
CCCM requires continued operation
during power mains interruptions, it is
recommended that the S/5™ CAM/
CCCM be powered from an
uninterruptible power supply or a
battery.
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic field should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
NOTE:
U
T
is the a.c. mains voltage prior to application of the test level.
Содержание Datex-Ohmeda S/5 Series
Страница 4: ......
Страница 90: ......
Страница 100: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 8 Document no M1144953 001 For your notes ...
Страница 118: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 26 Document no M1144953 001 For your notes ...
Страница 124: ......
Страница 128: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors iv Document no M1144955 001 For your notes ...
Страница 172: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 44 Document no M1144955 001 4 10 More ...
Страница 212: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors A 2 2 Document no M1144955 001 For your notes ...
Страница 214: ......
Страница 218: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors iv Document no M1144956 003 ...
Страница 266: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 48 Document no M1144956 003 ...
Страница 272: ......
Страница 304: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors A 4 4 Document no M1144958 001 For your notes ...
Страница 306: ......
Страница 308: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors ii Document no M1144959 001 ...