Service Menu
51
Document no. M1144955-001
Pneuma error
Module pneumatics error status. See
Table 6 Module pneumatics error status
for a detailed
description of the message.
Table 6
Module pneumatics error status
Bit
Error label
Description
What to do
0
pumping error
Stepper motor is not able to
generate negative pressure at the
pump pressure sensor during
zeroing.
Check the function of the stepper
motor.
1
eq. to amb. press failed Before filling the catheter, the
pressure in the module tubing is
equilibrated to the ambient pressure
by switching the pump valve on and
off. If ambient pressure is not
reached, this error is shown.
Calibrate the pressure sensors. Check
the pump valve function.
2
tubing/zero block occl Pressure at the pump pressure
sensor gets too low, when air is
aspirated though the zeroing valve.
Check tubing/valves for occlusion.
8
press sens values
differ
Pump pressure and catheter
pressure readings differ more than
50 mbars.
Calibrate the pressure sensors.
9
pneuma fatal error
This error is shown, if the pneumatic
error occurred is considered fatal.
Check for other pneumatics errors.
10
cath press too low
Catheter pressure reading is below
reasonable pressure values.
Make sure that you are not applying
too low pressure to the catheter
connector. Calibrate the pressure
sensors.
11
cath press too high
Catheter pressure reading is above
reasonable pressure values.
Make sure that you are not applying
too high pressure to the catheter
connector. Calibrate the pressure
sensors.
12
pump press too low
Pump pressure reading is below
reasonable pressure values.
Make sure that you are not applying
too low pressure to the catheter
connector. Calibrate the pressure
sensors.
13
pump press too high
Pump pressure reading is above
reasonable values.
Make sure that you are not applying
too high pressure to the catheter
connector. Calibrate the pressure
sensors.
Содержание Datex-Ohmeda S/5 Series
Страница 4: ......
Страница 90: ......
Страница 100: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 8 Document no M1144953 001 For your notes ...
Страница 118: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 26 Document no M1144953 001 For your notes ...
Страница 124: ......
Страница 128: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors iv Document no M1144955 001 For your notes ...
Страница 172: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 44 Document no M1144955 001 4 10 More ...
Страница 212: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors A 2 2 Document no M1144955 001 For your notes ...
Страница 214: ......
Страница 218: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors iv Document no M1144956 003 ...
Страница 266: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 48 Document no M1144956 003 ...
Страница 272: ......
Страница 304: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors A 4 4 Document no M1144958 001 For your notes ...
Страница 306: ......
Страница 308: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors ii Document no M1144959 001 ...