Service Menu
41
Document no. M1144955-001
4.8 NMT
Service menu for Neuromuscular Transmission
Module, E-NMT or M-NMT.
Start Curr. test
is a test where the module itself
checks the difference between a given current and
the measured current. A 3 k
Ω
resistance should be
connected between the stimulus electrodes before
starting the test. All the currents checked will be
displayed on the service data screen. After the
current value, the status of the test is shown. If the
test is not passed, send the module back to the
factory for calibration.
T1%/ref
T1% / ref
gives a selection for changing
settings related to the NMT MechanoSensor. A setting
of
3
should always be used for normal monitoring.
The settings
1
and
2
are for research purposes only.
Record Data
prints out the displayed service data
and circuit board information (id., serial number, and
software id.) on the recorder.
Service Data
Cable
shows the type of cable being used.
Cable check value
shows the bit amount. To check the value for each cable, see the following chart:
- EMG -100...+100
- Cable OFF >1950
- Piezo 900...1100
- Regional block 600....800
Module keys
checks the function of the module keys. A blue background appears at the back of the text
when a key is pressed for more than one second.
Stimulus mode
shows the selected stimulus. Stimulus mode can be changed using the NMT Setup menu.
Measurement
indicates ON/OFF.
T1%, T2%
shows the measured response. A value of 1000 corresponds to 100%.
Absolute T1, T2
shows the voltage measured from the A/D converter.
Noise
indicates the interference just before the measurement. A typical value is <100.
Offset
is an average of the noise measurement. A typical value is 510.
Curr set
is the selected current, a value of 700 corresponds to 70 mA.
Curr meas
is the measured current, a value of 700 corresponds to 70 mA.
Pulses
indicates pulses the module has produced.
Piezo probe T1% and Ref. search
shows information related to the MechanoSensor settings.
Содержание Datex-Ohmeda S/5 Series
Страница 4: ......
Страница 90: ......
Страница 100: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 8 Document no M1144953 001 For your notes ...
Страница 118: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 26 Document no M1144953 001 For your notes ...
Страница 124: ......
Страница 128: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors iv Document no M1144955 001 For your notes ...
Страница 172: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 44 Document no M1144955 001 4 10 More ...
Страница 212: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors A 2 2 Document no M1144955 001 For your notes ...
Страница 214: ......
Страница 218: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors iv Document no M1144956 003 ...
Страница 266: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 48 Document no M1144956 003 ...
Страница 272: ......
Страница 304: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors A 4 4 Document no M1144958 001 For your notes ...
Страница 306: ......
Страница 308: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors ii Document no M1144959 001 ...