Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
26
Document no
.
M1144951-004
5.
Wait for approximately 80 seconds. After the start-up screen appears, enter the Service
View and make sure that the information regarding monitor software has been updated.
Memorize the serial number of new software.
6.
Remove the software card.
7.
Replace the original device plate for monitor software by the new one supplied with the
software card.
8.
Perform Factory Reset,
see “Performing Factory Reset” .
Make sure that the monitor
functions normally after the restart.
9.
Set the time and date.
10. Set the monitor's network communication according to the network software in use, if
necessary. Enter the service menu:
Monitor Setup -Install/Service (16-4-34) - Service
(26-23-8)- Frame - Network- Config
and set
•
Network software S-CNET99 -> DRI Level = 1999
•
Network software S-CNET01 -> DRI Level = 2001
•
Network software S-CNET02 -> DRI Level = 2001 or 2002 (WLAN)
•
Network software L-NET03 -> DRI Level = 2003
•
Network software L-NET05 -> DRI Level = 2005
NOTE: The DRI level 1999 is not selectable in all monitor software versions.
NOTE: If the DRI level is changed, the monitor will restart automatically.
11. Fill out all necessary documentation regarding the new monitor software.
NOTE: The license agreement of monitor software needs to be in accordance with the monitor
software serial number. Make sure you archive the license agreement in a secure location.
NOTE: The first start-up after software loading takes considerably longer.
3.3.8 Performing Factory Reset
NOTE: The factory reset is necessary after replacing monitor software.
NOTE: The factory reset will restore all customized defaults, including language selection, to
factory defaults.
1.
Press the
Monitor Setup
key.
2.
Select
Install/Service
and password (16-4-34).
3.
Select
Service
and password (26-23-8).
4.
Select
Set/ Test
and
Factory Reset
.
5.
The monitor will perform an automatic restart. After the restart is completed, restart the
monitor also manually.
3.3.9 Installing the Datex-Ohmeda Network and MemCard Upgrade, U-CMMEM
The following instructions apply to upgrading the monitor with both the MemCard and the
Network options. Detailed upgrade instructions are supplied with the corresponding upgrade
product.
NOTE: You can download the option software only onto one monitor. During the downloading
of option software, the serial number of the monitor is written on the software card, and if the
downloading for some reason would fail, the software can be downloaded again onto the
same monitor, but not onto any other monitor.
1.
Make sure that the monitor is switched to standby.
Содержание Datex-Ohmeda S/5 Series
Страница 4: ......
Страница 90: ......
Страница 100: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 8 Document no M1144953 001 For your notes ...
Страница 118: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 26 Document no M1144953 001 For your notes ...
Страница 124: ......
Страница 128: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors iv Document no M1144955 001 For your notes ...
Страница 172: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 44 Document no M1144955 001 4 10 More ...
Страница 212: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors A 2 2 Document no M1144955 001 For your notes ...
Страница 214: ......
Страница 218: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors iv Document no M1144956 003 ...
Страница 266: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 48 Document no M1144956 003 ...
Страница 272: ......
Страница 304: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors A 4 4 Document no M1144958 001 For your notes ...
Страница 306: ......
Страница 308: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors ii Document no M1144959 001 ...