Functional check
63
Document no. M1144951-004
2.
Check that the quality of the recordings is acceptable.
"
5.3.20 Network connection
1.
Check that the Mon-Net cable connector and the Identification plug are clean and intact,
then connect them to the frame unit.
Check that the monitor connects to the network, i.e. the network symbol appears on the
upper right-hand corner of the screen. Also a message regarding the connected Central
should appear in the message field on the screen.
"
5.3.21 Wireless Network Option
1.
Check that the green LED on the wireless antenna card blinks and the WLAN signal
strenght symbol scrolls between zero and full or stays fixed on the monitor screen.
2.
Check that the wireless LAN network symbol appears on the upper right-hand corner of
the screen when the monitor connects to the Datex-Ohmeda Network.
NOTE: If the monitor does not connect to the Datex-Ohmeda Network, check the WLAN
configuration on the monitor and on the network.
"
5.3.22 Interface Module, E-INT/ M-INT
1.
Make sure that the monitor receives all necessary parameter data from the connected
devices. Check the screen configuration and the related interfacing settings, if necessary.
Monitor Setup
– Interfacing
"
5.3.23 Interface module for PSM, E-INTPSM
1.
Make sure that the monitor receives all necessary parameter data from the connected
devices. Check the screen configuration and the related interfacing settings, if necessary.
Monitor Setup
– Interfacing
"
5.3.24 Device Interfacing Solution, N-DISxxx
1.
Make sure that the monitor receives all necessary parameter data from the connected
devices. Check the screen configuration and the related interfacing settings, if necessary.
Monitor Setup
- Interfacing - Status Page
"
Содержание Datex-Ohmeda S/5 Series
Страница 4: ......
Страница 90: ......
Страница 100: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 8 Document no M1144953 001 For your notes ...
Страница 118: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 26 Document no M1144953 001 For your notes ...
Страница 124: ......
Страница 128: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors iv Document no M1144955 001 For your notes ...
Страница 172: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 44 Document no M1144955 001 4 10 More ...
Страница 212: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors A 2 2 Document no M1144955 001 For your notes ...
Страница 214: ......
Страница 218: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors iv Document no M1144956 003 ...
Страница 266: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 48 Document no M1144956 003 ...
Страница 272: ......
Страница 304: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors A 4 4 Document no M1144958 001 For your notes ...
Страница 306: ......
Страница 308: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors ii Document no M1144959 001 ...