![GE Datex-Ohmeda S/5 Series Скачать руководство пользователя страница 248](http://html.mh-extra.com/html/ge/datex-ohmeda-s-5-series/datex-ohmeda-s-5-series_technical-reference-manual_128583248.webp)
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
30
Document no
.
M1144956-003
Wait until the message ‘Loading from Mem. Module’ disappears, then check that the
original trends are available again.
"
Network option
NOTE: The following steps are applicable only when the Network option is installed in the
monitor.
42. Connection to Network
Check that the Mon-Net cable connector is clean and intact, then connect it to the
network connectors on the Frame Unit.
Check that the monitor connects to the Datex-Ohmeda Network, i.e. the network symbol
appears under the clock in the upper right hand corner of the screen. Also a message
regarding the connected Central should appear in the message field of the screen.
NOTE: The battery symbol overrides the network symbol in Compact Monitor.
If the message regarding the Central does not appear, check the status of the network.
"
43. Ethernet address
Enter the network board service menu:
Monitor Setup
Install Service
(Password 16-4-34)
Service - Frame
(Password 26-23-8)
Network
Take down the monitor’s Ethernet address, which is shown beside the text Address.
44. Data counters
Check that the service menu counters for the received (‘In’) and transmitted (‘Out’) data
are updated frequently.
"
45. Data error counters
Check that the counters for data errors (CRC, Frame, Transm.) are stable.
NOTE: The counters may show values higher than 0. However, if any of the values is increasing
continuously, it indicates a problem.
"
46. Hardware error counters
Check that the counters for hardware errors (Intern., Missed, FIFO, Overrun) show all 0. If
any of the counters show a value higher than 0, replace the network board.
"
Содержание Datex-Ohmeda S/5 Series
Страница 4: ......
Страница 90: ......
Страница 100: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 8 Document no M1144953 001 For your notes ...
Страница 118: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 26 Document no M1144953 001 For your notes ...
Страница 124: ......
Страница 128: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors iv Document no M1144955 001 For your notes ...
Страница 172: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 44 Document no M1144955 001 4 10 More ...
Страница 212: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors A 2 2 Document no M1144955 001 For your notes ...
Страница 214: ......
Страница 218: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors iv Document no M1144956 003 ...
Страница 266: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 48 Document no M1144956 003 ...
Страница 272: ......
Страница 304: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors A 4 4 Document no M1144958 001 For your notes ...
Страница 306: ......
Страница 308: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors ii Document no M1144959 001 ...