Service Menu
11
Document no. M1144955-001
1.3.2 Communication
Interface status
indicates the status of the data link
between the CPU and MEM sections. If the MemCard
option is installed, the status should always be on
ACTIVE. If the status blinks between ACTIVE and
CLOSED, a communications error has occurred.
Message types
indicates the type of data packets
that have been sent (
Tx
) and received (
Rx
) since last
monitor start. Data types are listed on the lines below
Message types
text.
Data types:
Record K
indicates the communication
between the Monitor and Record
Keeper.
File Operation
indicates the operations of Patient data.
Service
indicates the MEM section operations.
Modes
indicates the User Mode operations.
Module
status
indicates the number of sent/received data packets that relate to the MEM section status.
Packets
total
indicates the total amount of data packets that have been sent/received since the last
monitor start.
Bytes total
indicates the total amount of data bytes that have been sent/received since the last monitor
start.
The last four lines indicate transmission errors:
Timeouts
indicates the number of timeouts that have occurred in MEM section data transmission since
the last monitor start.
Chksum
err
indicates the number of checksum errors in data packets from MEM section since the last
monitor start.
Length
err
indicates the number of data packets with erroneous length from the MEM section since the
last monitor start.
Duplicated
indicates the number of duplicate data packets from the MEM section since the last monitor
start.
Содержание Datex-Ohmeda S/5 Series
Страница 4: ......
Страница 90: ......
Страница 100: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 8 Document no M1144953 001 For your notes ...
Страница 118: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 26 Document no M1144953 001 For your notes ...
Страница 124: ......
Страница 128: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors iv Document no M1144955 001 For your notes ...
Страница 172: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 44 Document no M1144955 001 4 10 More ...
Страница 212: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors A 2 2 Document no M1144955 001 For your notes ...
Страница 214: ......
Страница 218: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors iv Document no M1144956 003 ...
Страница 266: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 48 Document no M1144956 003 ...
Страница 272: ......
Страница 304: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors A 4 4 Document no M1144958 001 For your notes ...
Страница 306: ......
Страница 308: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors ii Document no M1144959 001 ...