Display Unit for F-CM(C)1 (Rev. 05), F-CM(C)REC1 (Rev. 01)
31
Document no. M1144956-003
47. Disconnection recognition
Disconnect the Mon-Net cable from the Frame Unit. Check that the message ‘Network
down:’ appears in the message field within 30 seconds. When the message appears, the
service menu values are no longer updated.
Reconnect the Mon-Net cable and check that the monitor connects to the Datex-Ohmeda
Network again.
"
48. ID-plug recognition
Switch the monitor to standby. Disconnect the Identification plug from the Frame Unit.
Switch the monitor back on and check that the message ‘Check network connectors’
appears in the message field.
Reconnect the Identification plug and check that the monitor connects to the
Anaesthesia Network.
"
Wireless LAN option
49. Signal strength
Check that the green LED on the wireless antenna card blinks and the WLAN signal
strength symbol roams or stays fixed on the monitor screen.
"
50. Configure access point
Configure one access point appropriately. Bring the Compact Monitor to good WLAN
coverage area. Check that the field level indicator stops scrolling.
Go to the
WLAN Status page (... Service -> Frame -> Network -> WLAN - WLAN Status).
Check that the Connection is active, AP MAC address is recognized and the connection
speed is at least 2 Mbits/s.
"
51. Connection to the Network
Make sure that the monitor's wireless LAN antenna card is in line of sight with a wireless
LAN access point antenna. Check that the monitor connects to the Datex-Ohmeda
Network, i.e. the wireless LAN network symbol appears in the upper right-hand corner of
the screen.
"
General
52. Trend retaining
Check that the monitor is capable of storing the trend information and temporary
settings in a short (max. 15 minutes) standby.
Press the membrane key
Display Trends
or
Pt.Data & Trends
-->
Trends
.
Check that
there is trend information available for the monitored parameters.
Содержание Datex-Ohmeda S/5 Series
Страница 4: ......
Страница 90: ......
Страница 100: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 8 Document no M1144953 001 For your notes ...
Страница 118: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 26 Document no M1144953 001 For your notes ...
Страница 124: ......
Страница 128: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors iv Document no M1144955 001 For your notes ...
Страница 172: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 44 Document no M1144955 001 4 10 More ...
Страница 212: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors A 2 2 Document no M1144955 001 For your notes ...
Страница 214: ......
Страница 218: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors iv Document no M1144956 003 ...
Страница 266: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 48 Document no M1144956 003 ...
Страница 272: ......
Страница 304: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors A 4 4 Document no M1144958 001 For your notes ...
Страница 306: ......
Страница 308: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors ii Document no M1144959 001 ...