Service Menu
21
Document no. M1144955-001
Service Data
Sensor ADULT/PEDIATRIC
according to the selected measurement mode (sensor).
Insp/Exp Flow Gains
shown apply to the selected sensor.
Aw Press Zero
and
Flow Zero
are the result of zeroing in the user service menu. They can be adjusted in the
factory calibration menu, but not permanently stored.
Aw Press Gain
is directly the value used in sw. It can be adjusted, but not permanently stored.
Exp Flow
and
Insp Flow Gains
are scaled as (User gain)/(Factory gain).
Exp
and
Insp Flow Gains
can be
adjusted also in the user service menu.
Common Offset
is the compensation factor for pressure difference reading of the difference sensor when
applying an equal pressure on both sides of the sensor.
Valves
can be changed between
MEASUREMENT
and
ZEROING
.
Zeroing
automatic zeroing either
ENABLED
or
DISABLED
(only factory service menu).
Aw Pressure
shows the real time value of airway pressure.
Condition
shows in which reference conditions the results are. With calibration pump or spirometry tester, the
results are always in ATP. If breathing is detected (EtCO
2
>
1.0%), the results are according to the
Flow & Vol
Setup
selection.
Flow
shows the flow measurement value.
TVol Exp, TVol Insp, MVol Exp
and
MVol Insp
are shown to ease calibration. The numbers are the same as on
the main display. The former pump calibration procedure has been dropped out. We claim that calibration is
not needed in routine clinical use, so a separate Flow calibration menu is not needed. Calibration can be done
with the pump or spirometry tester. The results must be taken from the screen and gains adjusted accordingly.
VCO
2
and
VO
2
Gain
show a value near 1000, the correct gains have been measured in the factory. Not used in
MRI Monitor.
Y deadspace
is the geometric volume in ml between the Y-piece and the D-lite. Default is 5 ml for a standard Y-
piece (as delivered with Siemens 900C ventilators). Used for VCO
2
and VO
2
delay time corrections.
Set
N
2
injection
to 1 during laboratory tests with a Spirometry tester and injection of N
2
gas (for scientific
validations use only). At power on of the module, the value is always zero.
Ambient pressure
value measured by the module.
Содержание Datex-Ohmeda S/5 Series
Страница 4: ......
Страница 90: ......
Страница 100: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 8 Document no M1144953 001 For your notes ...
Страница 118: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 26 Document no M1144953 001 For your notes ...
Страница 124: ......
Страница 128: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors iv Document no M1144955 001 For your notes ...
Страница 172: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 44 Document no M1144955 001 4 10 More ...
Страница 212: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors A 2 2 Document no M1144955 001 For your notes ...
Страница 214: ......
Страница 218: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors iv Document no M1144956 003 ...
Страница 266: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 48 Document no M1144956 003 ...
Страница 272: ......
Страница 304: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors A 4 4 Document no M1144958 001 For your notes ...
Страница 306: ......
Страница 308: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors ii Document no M1144959 001 ...