6.
Open the four screws holding the DC/DC power board.
7.
Lift the DC/DC power board carefully from the Display Unit.
3.3.5 To remove the Synchronization Unit
1.
Disassemble the Display Unit.
2.
Detach carefully the Synchronization board foil cable by opening the lock at the
connector X9 and pulling the foil out.
3.
Open the two screws holding the whole unit in the Display Unit back cover.
NOTE: Do not open the screws holding the Synchronization board in the Synchronization Unit.
3.3.6 To remove the Net Unit
1.
Disassemble the Display Unit.
2.
Detach the NetID and Network cables from the CPB, connectors X8 and X20 respectively.
3.
Open the two screws holding the Net Unit in the Display Unit cover.
4.
Open the two screws holding the ID connector in the Display Unit back cover (from the
outer side of the cover).
3.3.7 To remove the loudspeaker and fan
1.
Disassemble the Display Unit.
2.
Detach the loudspeaker and fan cables from the CPB, connectors X17 and X16
respectively.
3.
Open the screws holding the loudspeaker and fan.
NOTE: When reassembling, be sure to connect the cables to correct connectors on the PCB.
3.3.8 Disassembling the Connector Board and Inverter Board
1.
Disassemble the Display Unit.
2.
Set the Front Unit carefully lying with the LCD screen downwards. Note! Be careful not to
damage the LCD screen or the ComWheel.
3.
Detach carefully the lower keyboard flat cable from the Connector Board.
4.
Detach carefully the side panel keyboard flat cable from the Connector Board.
5.
Detach the Connector Board – Inverter Board cable.
6.
Detach the ComWheel cable.
7.
Open the four screws holding the Connector board and remove it.
8.
Detach the backlight cables from the Inverter Board.
9.
Open the three screws holding the Inverter Board.
NOTE: The Inverter Board is sensitive to bending.
NOTE: When replacing the Inverter Board, fasten the inverter screws gently, using only as
much force as necessary to fasten the board firmly enough in place. The screws must not be
fastened so tightly that the board will be bent from around the screws. Also, do not bend or
twist the Inverter Board during the assembly.
3.3.9 To Disassemble the LCD Display Element, F-CM(C)(REC)1 rev. 00
NOTE: Wear a static control wrist strap and soft cotton gloves (dust free) when handling the
LCD display parts. Hold them by their corners or edges. Do not touch the connector pins.
Содержание Datex-Ohmeda S/5 Series
Страница 4: ......
Страница 90: ......
Страница 100: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 8 Document no M1144953 001 For your notes ...
Страница 118: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 26 Document no M1144953 001 For your notes ...
Страница 124: ......
Страница 128: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors iv Document no M1144955 001 For your notes ...
Страница 172: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 44 Document no M1144955 001 4 10 More ...
Страница 212: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors A 2 2 Document no M1144955 001 For your notes ...
Страница 214: ......
Страница 218: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors iv Document no M1144956 003 ...
Страница 266: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 48 Document no M1144956 003 ...
Страница 272: ......
Страница 304: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors A 4 4 Document no M1144958 001 For your notes ...
Страница 306: ......
Страница 308: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors ii Document no M1144959 001 ...