4.2.5 Functional check
There are two ways to check the function of the Device Interfacing Solution:
1.
Press the
Monitor Setup
key.
2.
Select
Interfacing
and open the
Status Page
menu. The status page shows you the
current communication status of the interfacing module (1 - 10).
NOTE: The status message ‘Connected
’
appears on the Status Page after you have connected
the external device to the interfacing module and turned it on. Note also that the monitor and
the interfacing module must be operational.
•
Check the LED indicators on the interfacing module (the green LED indicates physical
connections, the yellow LED software selections)
:
4.2.6 Selecting the parameter data source
Select the external device via
Monitor Setup
- Interfacing
menu:
•
Select the desired parameter, for example
Gases.
•
Select the desired source by name, for example
Aestiva
.
NOTE: The name of the device is visible on the list only if the device is correctly connected.
Detailed information about interfacing module related mountings, connections and settings is
included in the installation guides that are delivered with the interfacing modules.
GREEN
YELLOW
INDICATION
lit
dark
Physical connections between the monitor, interfacing
module and external device are in order and the device
has been selected in the menu.
dark
lit
There is something wrong with the physical connections
between the monitor, interfacing module and external
device. The external device has not been selected in the
menu.
lit
lit
Physical connections between the monitor, interfacing
module and external device are in order but the external
device has not been selected in the menu.
dark
dark
The interfacing module is not connected to the monitor.
Содержание Datex-Ohmeda S/5 Series
Страница 4: ......
Страница 90: ......
Страница 100: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 8 Document no M1144953 001 For your notes ...
Страница 118: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 26 Document no M1144953 001 For your notes ...
Страница 124: ......
Страница 128: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors iv Document no M1144955 001 For your notes ...
Страница 172: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 44 Document no M1144955 001 4 10 More ...
Страница 212: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors A 2 2 Document no M1144955 001 For your notes ...
Страница 214: ......
Страница 218: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors iv Document no M1144956 003 ...
Страница 266: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors 48 Document no M1144956 003 ...
Страница 272: ......
Страница 304: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors A 4 4 Document no M1144958 001 For your notes ...
Страница 306: ......
Страница 308: ...Datex Ohmeda S 5 Compact Anesthesia and Compact Critical Care Monitors ii Document no M1144959 001 ...