Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product
Page 5-11
3.
To reselect continuous operation, press the “Pulse” key
again. The pulse rates will no longer
display on the status bar.
5.2.5.
Low Dose Imaging
Lose dose imaging is to reduce the exposure dose to the patient.
To produce a Low Dose image
1.
Press the
“
Low Dose
”
key
on the C-Arm control panel to enable low dose imaging.
2.
Press the related X-Ray switch on the foot/handswitch to make low dose fluoro exposures.
5.2.6.
Imaging Mode List
The following list summarizes the modes available on the system. Additionally it shows the X-Ray
techniques available with each mode, the status bar and C-Arm control panel display for each mode, and a
brief list of examples for each mode. The examples are not exhaustive and other uses may be as valid as
the examples shown. The user must be aware of the techniques associated with the various modes and
monitor dose rates and doses for each patient.
Auto: X-ray techniques (kVp/mA) are adjusted automatically by system to optimize the image.
Manual: X-ray techniques (kVp/mA) are adjusted manually by user to get the desired image.
The user can select Auto or Manual mode listed in the table below as she/he wants.
Mode
X-Ray
Techniques
Display
Imaging Task
Normal Fluoro
Auto/ Manual
kVp:40 to 110
mA: 0.1 to 4
General Fluoro Imaging
High Level Fluoro
Auto/ Manual
kVp:40 to 110
mA: 0.2 to 12
High quality but high-dose
Imaging
Low Dose
Auto/ Manual
kVp:40 to 110
mA: 0.1 to 2
Continuous fluoro imaging
with a lower dose
Low Dose High Level
Auto/ Manual
kVp:40 to 110
mA: 0.1 to 6
+
High quality with a lower
dose
Содержание Brivo OEC 715
Страница 9: ...Chapter1 Introduction and Safety...
Страница 34: ...Chapter2 System Overview...
Страница 36: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 3 2 2 C Arm Components...
Страница 38: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 5 2 3 Workstation Components...
Страница 40: ...Chapter3 Mechanical Positioning...
Страница 52: ...Chapter4 Start Up and Operating Control...
Страница 77: ...Chapter5 Usual Operations...
Страница 108: ...Chapter6 SpecialApplications...
Страница 139: ...Chapter7 Radiographic Film...
Страница 142: ...Chapter8 System Setup...
Страница 149: ...System Setup Page 8 8 Click any mouse button to return to normal screen...
Страница 150: ...Chapter9 Maintenance...
Страница 160: ...Chapter10 Display Messages...
Страница 162: ...Chapter11 Labels and Symbols...
Страница 167: ...Labels and Symbols Page 11 6 18 1 System rating plate indicates manufacture information and input power requirements...
Страница 168: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 11 7 19 1 X ray Generator rating plate Total Filtration label...
Страница 177: ...Labels and Symbols Page 11 16 42 12 15 29 40 15 13 10 18 39 31 27 24 23 37 38...
Страница 179: ...Chapter12 Technical Reference...
Страница 183: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 12 5 Workstation Weight 170 10kg...
Страница 194: ...Technical Reference Page 12 16 Lateral configuration 1 5m LOHMANN Lateral configuration 1m LOHMANN...
Страница 206: ...Technical Reference Page 12 28 12 11 System Block...
Страница 209: ......
Страница 210: ...www gehealthcare com Imagination at work...